Bone conduction hearing sensitivity in normal-hearing subjects: Transcutaneous stimulation at BAHA vs BCI position

被引:34
作者
Reinfeldt, Sabine [1 ]
Hakansson, Bo [1 ]
Taghavi, Hamidreza [1 ]
Eeg-Olofsson, Mans [2 ]
机构
[1] Chalmers Univ Technol, Dept Signals & Syst, SE-41296 Gothenburg, Sweden
[2] Univ Gothenburg, Sahlgrenska Acad, Dept Otorhinolaryngol Head & Neck Surg, Sahlgrenska Univ Hosp, Gothenburg, Sweden
关键词
Bone conduction; hearing sensitivity; transcutaneous stimulation; bone conduction implant; bone-anchored hearing aid; threshold shift; ear-canal sound pressure; objective measure; normal-hearing subjects; SKIN REACTIONS; SOUND; TRANSMISSION; IMPLANT; COMPLICATIONS; THRESHOLDS; DEVICE; DESIGN; SYSTEM;
D O I
10.3109/14992027.2014.880813
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: Bone conduction (BC) stimulation closer to the cochlea has previously been shown to give higher cochlear promontory acceleration measured by laser Doppler vibrometry (LDV). This study is investigating whether stimulation closer to the cochlea also gives improved hearing sensitivity. Furthermore, the study compares shifts in hearing sensitivity (BC thresholds) and ear-canal sound pressure (ECSP). Design: BC hearing thresholds and ECSP have been measured for stimulation at two positions: the existing bone-anchored hearing aid (BAHA) position, and a new bone conduction implant (BCI) position that is located closer to the cochlea. Study sample: The measurements were made on 20 normal-hearing subjects. Results: Depending on frequency, the ipsilateral hearing threshold was 3-14 dB better, and the ipsilateral ECSP was 2-12 dB higher for the BCI than for the BAHA position, with no significant differences between threshold and ECSP shifts at group level for most frequencies, and individually only for some subjects. Conclusions: It was found that both the objective ECSP and the subjective hearing threshold measurements gave similar improvement as previous LDV measurements for stimulation closer to the cochlea. One exception was that the LDV measurements did not show the improved sensitivity for frequencies below 500 Hz found here.
引用
收藏
页码:360 / 369
页数:10
相关论文
共 31 条
[1]  
[Anonymous], 1994, ISO3893
[2]  
[Anonymous], 2003, ISO226
[3]  
[Anonymous], 2017, ISO 389-1: Acoustics-Reference Zero for the Calibration of Audiometric Equipment
[4]  
[Anonymous], OSS 2013 M NEWC UK
[5]   Revision BAHA surgery [J].
Battista, Robert A. ;
Littlefield, Philip D. .
OTOLARYNGOLOGIC CLINICS OF NORTH AMERICA, 2006, 39 (04) :801-+
[6]   Assessment of More Than 1,000 Implanted Percutaneous Bone Conduction Devices: Skin Reactions and Implant Survival [J].
Dun, Catharina A. J. ;
Faber, Hubert T. ;
de Wolf, Maarten J. F. ;
Mylanus, Emmanuel A. M. ;
Cremers, Cor W. R. J. ;
Hol, Myrthe K. S. .
OTOLOGY & NEUROTOLOGY, 2012, 33 (02) :192-198
[7]   The Bone Conduction Implant-First Implantation, Surgical and Audiologic Aspects [J].
Eeg-Olofsson, Mans ;
Hakansson, Bo ;
Reinfeldt, Sabine ;
Taghavi, Hamidreza ;
Lund, Henrik ;
Jansson, Karl-Johan Freden ;
Hakansson, Emil ;
Stalfors, Joacim .
OTOLOGY & NEUROTOLOGY, 2014, 35 (04) :679-685
[8]   Transmission of bone conducted sound - Correlation between hearing perception and cochlear vibration [J].
Eeg-Olofsson, Mans ;
Stenfelt, Stefan ;
Taghavi, Hamidreza ;
Reinfeldt, Sabine ;
Hakansson, Bo ;
Tengstrand, Tomas ;
Finizia, Caterina .
HEARING RESEARCH, 2013, 306 :11-20
[9]   Implications for Contralateral Bone-Conducted Transmission as Measured by Cochlear Vibrations [J].
Eeg-Olofsson, Mans ;
Stenfelt, Stefan ;
Granstrom, Gosta .
OTOLOGY & NEUROTOLOGY, 2011, 32 (02) :192-198
[10]   Transmission of bone-conducted sound in the human skull measured by cochlear vibrations [J].
Eeg-Olofsson, Mans ;
Stenfelt, Stefan ;
Tjellstrom, Anders ;
Granstrom, Gosta .
INTERNATIONAL JOURNAL OF AUDIOLOGY, 2008, 47 (12) :761-769