Performance prediction of cryogenically cooled silicon crystal monochromator

被引:0
作者
Zhang, L [1 ]
Lee, WK [1 ]
Wulff, M [1 ]
Eybert, L [1 ]
机构
[1] European Synchrotron Radiat Facil, F-38043 Grenoble, France
来源
SYNCHROTRON RADIATION INSTRUMENTATION | 2004年 / 705卷
关键词
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
To predict the performance of the cryogenically cooled silicon crystal, intensive studies have been carried out to sort out the influences of various parameters, such as heat load power and power distribution, cooling coefficient, and beam size. The thermal slope error of the crystal is calculated by finite element modeling. Quadratic law was applied to calculate the rocking-curve width. Heat load tests were also performed with a channel-cut silicon monochromator on beamline ID09 at the European Synchrotron Radiation Facility (ESRF). The silicon crystal is indirectly cooled from the sides by liquid nitrogen. Measured rocking-curve widths are compared with those calculated by finite element modeling. When we include the broadening from the intrinsic rocking-curve width and mounting strain, the calculated rocking-curve width versus heat load is in excellent agreement with experiment.
引用
收藏
页码:623 / 626
页数:4
相关论文
共 9 条
[1]   Performance limits of indirectly cryogenically cooled silicon monochromators - experimental results at the APS [J].
Lee, WK ;
Fezzaa, K ;
Fernandez, P ;
Tajiri, G ;
Mills, D .
JOURNAL OF SYNCHROTRON RADIATION, 2001, 8 (01) :22-25
[2]   Performance of synchrotron X-ray monochromators under heat load - Part 3: Comparison between theory and experiment [J].
Hoszowska, J ;
Mocella, V ;
Zhang, L ;
Mighore, JS ;
Freund, AK ;
Ferrero, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 467 :631-634
[3]   CRYOGENIC COOLING OF MONOCHROMATORS [J].
MAROT, G ;
ROSSAT, M ;
FREUND, A ;
JOKSCH, S ;
KAWATA, H ;
ZHANG, L ;
ZIEGLER, E ;
BERMAN, L ;
CHAPMAN, D ;
HASTINGS, JB ;
IAROCCI, M .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01) :477-480
[4]   A new approach to the solution of the Takagi-Taupin equations for X-ray optics: application to a thermally deformed crystal monochromator [J].
Mocella, V ;
Lee, WK ;
Tajiri, G ;
Mills, D ;
Ferrero, C ;
Epelboin, Y .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 2003, 36 :129-136
[5]   Performances of synchrotron X-ray monochromators under heat load. Part 2. Application of the Takagi-Taupin diffraction theory [J].
Mocella, V ;
Ferrero, C ;
Freund, AK ;
Hoszowska, J ;
Zhang, L ;
Epelboin, Y .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 467 :414-417
[6]  
ROGERS CS, 1995, REV SCI INSTRUM, V66, P2494
[7]   The performance of a cryogenically cooled monochromator for an in-vacuum undulator beamline [J].
Zhang, L ;
Lee, WK ;
Wulff, M ;
Eybert, L .
JOURNAL OF SYNCHROTRON RADIATION, 2003, 10 :313-319
[8]  
ZHANG L, 1993, P SOC PHOTO-OPT INS, V1997, P223, DOI 10.1117/12.163803
[9]   Parkin: clinical aspects and neurobiology [J].
Zhang, Y ;
Dawson, VL ;
Dawson, TM .
CLINICAL NEUROSCIENCE RESEARCH, 2001, 1 (06) :467-482