The influence of various automated devices on the shaping ability of Mtwo rotary nickel-titanium instruments

被引:20
作者
Buerklein, S. [1 ]
Schaefer, E. [1 ]
机构
[1] Univ Munster, Dept Operat Dent, D-48149 Munster, Germany
关键词
canal straightening; Mtwo direct; resin blocks; rotary instruments; torque control;
D O I
10.1111/j.1365-2591.2006.01171.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim To compare the shaping ability of Mtwo rotary nickel-titanium instruments in simulated curved canals and in curved canals of extracted teeth when set into permanent rotation with two different torque-limited automated devices. Methodology Root canal instrumentation was performed with two different torque-limited automated devices (Mtwo direct handpiece and Endo IT motor) using Mtwo rotary nickel-titanium instruments up to size 35. Simulated canals: 28 degrees and 35 degrees curved canals in resin blocks (n = 20 canals in each group) were prepared. Pre- and post-instrumentation images were recorded and assessment of canal shape was completed with a computer image analysis program. Extracted teeth: a total of 40 curved root canals were divided into two groups, which were balanced with respect to the angle and the radius of canal curvature. Straightening of the canal curvatures was determined with a computer image analysis program. Incidence of canal aberrations, preparation time, changes of working length and instrument failures were recorded both in simulated and real canals and were analysed using the t-test and a chi-square test. Results In simulated canals, instrumentation with Endo IT was significantly faster than with Mtwo direct (P < 0.05). No significant differences between the two devices were noted when all other parameters were assessed (P > 0.05). Only one instrument separated (Mtwo direct) during the enlargement of 28 degrees curved simulated canals. All systems maintained working distance well. Conclusions Both systems respected original root canal curvature well and were safe, indicating that the Mtwo direct handpiece is suitable for preparing curved root canals with the Mtwo instruments.
引用
收藏
页码:945 / 951
页数:7
相关论文
共 50 条
[31]   Comparison of preparation of curved root canals with Hyflex CM and Revo-S rotary nickel-titanium instruments [J].
Buerklein, S. ;
Boerjes, L. ;
Schaefer, E. .
INTERNATIONAL ENDODONTIC JOURNAL, 2014, 47 (05) :470-476
[32]   Nickel-titanium instruments in endodontics: a concise review of the state of the art [J].
Gavini, Giulio ;
dos Santos, Marcelo ;
Caldeira, Celso Luis ;
de Lima Machado, Manoel Eduardo ;
Freire, Laila Gonzales ;
Iglecias, Elaine Faga ;
Peters, Ove Andrea ;
de Miranda Candeiro, George Taccio .
BRAZILIAN ORAL RESEARCH, 2018, 32 :44-65
[33]   Heat Treatment and Surface Treatment of Nickel-Titanium Endodontic Instruments [J].
Kwak, Sang Won ;
Shen, Ya ;
Liu, He ;
Wang, Zhejun ;
Kim, Hyeon-Cheol ;
Haapasalo, Markus .
FRONTIERS IN DENTAL MEDICINE, 2021, 2
[34]   Evaluation of the mechanical properties of different nickel-titanium retreatment instruments [J].
Weissheimer, Theodoro ;
Heck, Luana ;
Calefi, Pedro Henrique Souza ;
Alcalde, Murilo Priori ;
da Rosa, Ricardo Abreu ;
Vivan, Rodrigo Ricci ;
Duarte, Marco Antonio Hungaro ;
So, Marcus Vinicius Reis .
AUSTRALIAN ENDODONTIC JOURNAL, 2021, 47 (02) :265-272
[35]   Influence of Heat Treatment of Nickel-Titanium Instruments on Cyclic Fatigue Resistance in Simulated Curved Canals [J].
Campos, Gustavo Oliveira ;
Fontana, Carlos Eduardo ;
Leal Vieira, Victor Talarico ;
Elias, Carlos Nelson ;
de Martin, Alexandre Sigrist ;
da Silveira Bueno, Carlos Eduardo .
EUROPEAN JOURNAL OF DENTISTRY, 2023, 17 (02) :472-477
[36]   Influence of heat-treatment on torsional resistance to fracture of nickel-titanium endodontic instruments. [J].
Lo Savio, F. ;
Pedulla, E. ;
Rapisarda, E. ;
La Rosa, G. .
21ST EUROPEAN CONFERENCE ON FRACTURE, (ECF21), 2016, 2 :1311-1318
[37]   Canal shapes produced sequentially during instrumentation with Quantec SC rotary nickel-titanium instruments: a study in simulated canals [J].
Griffiths, IT ;
Chassot, AL ;
Nascimento, MF ;
Bryant, ST ;
Dummer, PMH .
INTERNATIONAL ENDODONTIC JOURNAL, 2001, 34 (02) :107-112
[38]   Canal shapes produced sequentially during instrumentation with Quantec LX rotary nickel-titanium instruments: a study in simulated canals [J].
Griffiths, IT ;
Bryant, ST ;
Dummer, PMH .
INTERNATIONAL ENDODONTIC JOURNAL, 2000, 33 (04) :346-354
[39]   Shaping Ability of Different Nickel-Titanium Systems in Simulated S-shaped Canals with and without Glide Path [J].
Buerklein, Sebastian ;
Poschmann, Thomas ;
Schaefen, Edgar .
JOURNAL OF ENDODONTICS, 2014, 40 (08) :1231-1234
[40]   Influence of Coronal Flaring on the Shaping Ability of Two Heat-Treated Nickel-Titanium Endodontic Files: A Micro-Computed Tomographic Study [J].
Hawi, Nadine ;
Pedulla, Eugenio ;
La Rosa, Giusy Rita Maria ;
Conte, Gianluca ;
Nehme, Walid ;
Neelakantan, Prasanna .
JOURNAL OF CLINICAL MEDICINE, 2023, 12 (01)