Evaluation of a lower-powered analyzer and sampling system for eddy-covariance measurements of nitrous oxide fluxes

被引:10
作者
Brown, Shannon E. [1 ]
Sargent, Steve [2 ]
Wagner-Riddle, Claudia [1 ]
机构
[1] Univ Guelph, Sch Environm Sci, Guelph, ON, Canada
[2] Campbell Sci Inc, Logan, UT USA
关键词
FREQUENCY-RESPONSE CORRECTIONS; GAS ANALYZERS; SPECTRAL CHARACTERISTICS; MANAGED GRASSLAND; DIODE-LASER; N2O FLUXES; EMISSIONS; FIELD; PATH; CO2;
D O I
10.5194/amt-11-1583-2018
中图分类号
P4 [大气科学(气象学)];
学科分类号
0706 ; 070601 ;
摘要
Nitrous oxide (N2O) fluxes measured using the eddy-covariance method capture the spatial and temporal heterogeneity of N2O emissions. Most closed-path tracegas analyzers for eddy-covariance measurements have large-volume, multi-pass absorption cells that necessitate high flow rates for ample frequency response, thus requiring high-power sample pumps. Other sampling system components, including rain caps, filters, dryers, and tubing, can also degrade system frequency response. This field trial tested the performance of a closed-path eddy-covariance system for N2O flux measurements with improvements to use less power while maintaining the frequency response. The new system consists of a thermoelectrically cooled tunable diode laser absorption spectrometer configured to measure both N2O and carbon dioxide (CO2) The system features a relatively small, single-pass sample cell (200 mL) that provides good frequency response with a lower-powered pump (similar to 250 W). A new filterless intake removes particulates from the sample air stream with no additional mixing volume that could degrade frequency response. A single-tube dryer removes water vapour from the sample to avoid the need for density or spectroscopic corrections, while maintaining frequency response. This eddy-covariance system was collocated with a previous tunable diode laser absorption spectrometer model to compare N2O and CO2 flux measurements for two full growing seasons (May 2015 to October 2016) in a fertilized cornfield in Southern Ontario, Canada. Both spectrometers were placed outdoors at the base of the sampling tower, demonstrating ruggedness for a range of environmental conditions (minimum to maximum daily temperature range: -26.1 to 31.6 degrees C). The new system rarely required maintenance. An in situ frequency-response test demonstrated that the cutoff frequency of the new system was better than the old system (3.5 Hz compared to 2.30 Hz) and similar to that of a closed-path CO2 eddy-covariance system (4.05 Hz), using shorter tubing and no dryer, that was also collocated at the site. Values of the N2O fluxes were similar between the two spectrometer systems (slope = 1.01, r(2) = 0.96); CO2 fluxes as measured by the short-tubed eddy-covariance system and the two spectrometer systems correlated well (slope = 1.03, r(2) = 0.998). The new lower-powered tunable diode laser absorption spectrometer configuration with the filterless intake and singletube dryer showed promise for deployment in remote areas.
引用
收藏
页码:1583 / 1597
页数:15
相关论文
共 61 条
  • [1] Abalos D., 2015, GLOBAL CHANGE BIOL, V3, P1244
  • [2] [Anonymous], 2012, EDDY COVARIANCE PRAC
  • [3] Aubinet M, 2000, ADV ECOL RES, V30, P113, DOI 10.1016/S0065-2504(08)60018-5
  • [4] Dimensioning IRGA gas sampling systems: laboratory and field experiments
    Aubinet, Marc
    Joly, Lilian
    Loustau, Denis
    De Ligne, Anne
    Chopin, Henri
    Cousin, Julien
    Chauvin, Nicolas
    Decarpenterie, Thomas
    Gross, Patrick
    [J]. ATMOSPHERIC MEASUREMENT TECHNIQUES, 2016, 9 (03) : 1361 - 1367
  • [5] Determining the sea-air flux of dimethylsulfide by eddy correlation using mass spectrometry
    Blomquist, B. W.
    Huebert, B. J.
    Fairall, C. W.
    Faloona, I. C.
    [J]. ATMOSPHERIC MEASUREMENT TECHNIQUES, 2010, 3 (01) : 1 - 20
  • [6] Estimating a Lagrangian Length Scale Using Measurements of CO2 in a Plant Canopy
    Brown, Shannon E.
    Warland, Jon S.
    Santos, Eduardo A.
    Wagner-Riddle, Claudia
    Staebler, Ralf
    Wilton, Meaghan
    [J]. BOUNDARY-LAYER METEOROLOGY, 2013, 147 (01) : 83 - 102
  • [7] Novel design of an enclosed CO2/H2O gas analyser for eddy covariance flux measurements
    Burba, G. G.
    Mcdermitt, D. K.
    Anderson, D. J.
    Furtaw, M. D.
    Eckles, R. D.
    [J]. TELLUS SERIES B-CHEMICAL AND PHYSICAL METEOROLOGY, 2010, 62 (05): : 743 - 748
  • [8] Differences in field-scale N2O flux linked to crop residue removal under two tillage systems in cold climates
    Congreves, Katelyn A.
    Brown, Shannon E.
    Nemeth, Deanna D.
    Dunfield, Kari E.
    Wagner-Riddle, Claudia
    [J]. GLOBAL CHANGE BIOLOGY BIOENERGY, 2017, 9 (04): : 666 - 680
  • [9] The contribution of manure and fertilizer nitrogen to atmospheric nitrous oxide since 1860
    Davidson, Eric A.
    [J]. NATURE GEOSCIENCE, 2009, 2 (09) : 659 - 662
  • [10] Comparing laser-based open- and closed-path gas analyzers to measure methane fluxes using the eddy covariance method
    Detto, Matteo
    Verfaillie, Joseph
    Anderson, Frank
    Xu, Liukang
    Baldocchi, Dennis
    [J]. AGRICULTURAL AND FOREST METEOROLOGY, 2011, 151 (10) : 1312 - 1324