Finite element analysis of stress around a sternum screw used to prevent sternal dehiscence after heart surgery

被引:8
作者
Jutley, RS
Watson, MA
Shepherd, DET
Hukins, DWL
机构
[1] Univ Aberdeen, Dept Biomed Phys & Bioengn, Aberdeen AB25 2ZD, Scotland
[2] Grampian Univ Hosp NHS Trust, Dept Cardiothorac Surg, Aberdeen, Scotland
关键词
bone; dehiscence; finite element analysis; screw; sternotomy; sternum;
D O I
10.1243/09544110260216586
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The sternum screw has been proposed as a means of preventing sternal dehiscence, following heart surgery, by increasing the contact area between the wire used to close the median sternotomy and the surrounding bone; as a result, the contact stress is reduced. A finite element model was constructed of a cylindrical wire or screw passing through a block of sternum which consisted of cancellous bone sandwiched within a cortical shell. The thickness of the cortical shell and the material properties of bone were varied between reasonable values. The stress distribution in the sternum was calculated for each model when the wire was subjected to a tension (250 N) which would be required for six wires to withstand a strong cough (40 kPa). Results were validated by comparison with a simple analytical model in which the bone and wire were considered incompressible. They show that the screw reduces the contact stress to almost one-seventh of its value when wire is used alone. Contact stresses are especially high if the cortical shell is thin. The high stress in the bone around a screw falls off within a few millimetres. As a result, no problems are anticipated in placing six screws in each half-sternum so that the sternotomy may be closed with the usual six wires.
引用
收藏
页码:315 / 321
页数:7
相关论文
共 26 条
[1]  
ABENDSCHEIN W, 1970, CLIN ORTHOP RELAT R, P294
[2]  
ADAMS V, 1999, BENCHMARK APR, P8
[3]  
[Anonymous], J KYOTO PREF MED U
[4]  
Ashley GT, 1956, J Forensic Med, V3, P27
[5]   ELASTIC-MODULUS OF TRABECULAR BONE MATERIAL [J].
ASHMAN, RB ;
RHO, JY .
JOURNAL OF BIOMECHANICS, 1988, 21 (03) :177-181
[6]   A biomechanical study of median sternotomy closure techniques [J].
Casha, AR ;
Yang, L ;
Kay, PH ;
Saleh, M ;
Cooper, GJ .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1999, 15 (03) :365-369
[7]   Fatigue testing median sternotomy closures [J].
Casha, AR ;
Gauci, M ;
Yang, L ;
Saleh, M ;
Kay, PH ;
Cooper, GJ .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2001, 19 (03) :249-253
[8]   The peak CT number of profiles perpendicular to the vertebral cortical shell may be a useful indicator of the integrity of the cortical shell [J].
Dougherty, G .
MEDICAL ENGINEERING & PHYSICS, 2000, 22 (07) :487-491
[9]  
Gere J., 1987, MECH MAT
[10]   Evaluation of cortical bone by computed tomography [J].
Hangartner, TN ;
Gilsanz, V .
JOURNAL OF BONE AND MINERAL RESEARCH, 1996, 11 (10) :1518-1525