Femtomolar detection of cardiac troponin I using a novel label-free and reagent-free dendrimer enhanced impedimetric immunosensor

被引:51
|
作者
Akter, Rashida [1 ]
Jeong, Bongjin [1 ]
Lee, Yong-Mi [1 ]
Choi, Jong-Soon [1 ,2 ]
Rahman, Md. Aminur [1 ]
机构
[1] Chungnam Natl Univ, Grad Sch Analyt Sci & Technol, GRAST, Daejeon 34134, South Korea
[2] Korea Basic Sci Inst, Div Life Sci, Daejeon 34133, South Korea
基金
新加坡国家研究基金会;
关键词
Impedimetric immunosensor; Dendrimer; 3; 5; '-Tetramethylbenzidine; Protein biomarkers; Cardiac troponin I; Human serum; ELECTROCHEMICAL IMMUNOSENSOR; GOLD NANOPARTICLES; CANCER BIOMARKERS; GRAPHENE; PROTEIN; BIOSENSOR; PLATFORM; ARRAY;
D O I
10.1016/j.bios.2017.01.021
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A novel highly sensitive dendrimer coupled impedimetric immunosensor was developed for the label-free and reagent-free detection of cardiac troponin I (TnI) in serum samples. The immunosensor probe was fabricated by covalently attaching carboxylic acid-functionalized third generation (G3) poly (amidoamine) (PAMAM) dendrimer (Den) on the 3, 3', 5, 5'-tetramethyl benzidine (TMB) modified 6-mercaptohexanoic acid (MHA) self-assembled monolayer (SAM) on a gold (Au) electrode. Monoclonal anti-TnI antibody was then covalently immobilized on the Den and TMB attached MHA SAM modified surface. TMB was used as an internal surface redox couple for generating signal which also allowed to avoid the use of an external one (i.e. ferricyanide couple) in solution during the impedance measurement for monitoring the antibody-antigen binding. On the other hand, Den was used as a signal enhancer by immobilizing more anti-body on the immunosensor probe. The immunosensor probe was characterized using X-ray photoelectron spectroscopy (XPS), quartz crystal microbalance (QCM), cyclic voltammetry (CV), and electrochemical impedance spectroscopy (EIS) techniques. The TnI detection in diluted serum was based on the measurement of charge transfer resistance (R-ct) of the electron transfer process of the surface -attached TMB before and after immunobinding. Under the optimized condition, the proposed immunosensor could detect human TnI in diluted serum samples as low as 11.7 fM with a wide linear dynamic range, good stability, and excellent specificity. The validity of the proposed method was tested in various TnI spiked human undiluted serum samples and was compared with the enzyme -linked immunosorbent assay (ELISA). The results suggested that the proposed immunosensor could be a useful tool for practical applications in clinical diagnosis.
引用
收藏
页码:637 / 643
页数:7
相关论文
共 50 条
  • [21] Label-Free QCM Immunosensor for the Detection of Ochratoxin A
    Pirincci, Serife Seyda
    Ertekin, Ozlem
    Laguna, Duygu Ercan
    Ozen, Fehime Seyma
    Ozturk, Zafer Ziya
    Ozturk, Selma
    SENSORS, 2018, 18 (04)
  • [22] Label-free impedimetric immunosensor for point-of-care detection of COVID-19 antibodies
    Shoute, Lian C. T.
    Abdelrasoul, Gaser N.
    Ma, Yuhao
    Duarte, Pedro A.
    Edwards, Cole
    Zhuo, Ran
    Zeng, Jie
    Feng, Yiwei
    Charlton, Carmen L.
    Kanji, Jamil N.
    Babiuk, Shawn
    Chen, Jie
    MICROSYSTEMS & NANOENGINEERING, 2023, 9 (01)
  • [23] Label-free, capacitive immunosensor for protein detection
    Gul, Ozgur
    Heves, Emre
    Kaynak, Mehmet
    Basaga, Huveyda
    Gurbuz, Yasar
    2006 IEEE SENSORS, VOLS 1-3, 2006, : 600 - +
  • [24] Label-free impedimetric immunosensor for ultrasensitive detection of cancer marker Murine double minute 2 in brain tissue
    Elshafey, Reda
    Tlili, Chaker
    Abulrob, Abedelnasser
    Tavares, Ana C.
    Zourob, Mohammed
    BIOSENSORS & BIOELECTRONICS, 2013, 39 (01) : 220 - 225
  • [25] A novel label-free impedimetric immunosensor for sensitive detection of prostate specific antigen using Au nanoparticles/MWCNTs- graphene quantum dots nanocomposite
    Ghanavati, Mehrnaz
    Tadayon, Fariba
    Bagheri, Hasan
    MICROCHEMICAL JOURNAL, 2020, 159
  • [26] Paper based point of care immunosensor for the impedimetric detection of cardiac troponin I biomarker
    Vasantham, Shreyas
    Alhans, Ruby
    Singhal, Chaitali
    Nagabooshanam, Shalini
    Nissar, Sumaya
    Basu, Tinku
    Ray, Sekhar C.
    Wadhwa, Shikha
    Narang, Jagriti
    Mathur, Ashish
    BIOMEDICAL MICRODEVICES, 2019, 22 (01)
  • [27] Ultrasensitive label-free optical microfiber coupler biosensor for detection of cardiac troponin I based on interference turning point effect
    Zhou, Wenchao
    Li, Kaiwei
    Wei, Youlian
    Hao, Peng
    Chi, Mingbo
    Liu, Yongshun
    Wu, Yihui
    BIOSENSORS & BIOELECTRONICS, 2018, 106 : 99 - 104
  • [28] Label-free electrochemical aptasensor for femtomolar detection of 17β-estradiol
    Zhu, Bicheng
    Alsager, Omar A.
    Kumar, Shalen
    Hodgkiss, Justin M.
    Travas-Sejdic, Jadranka
    BIOSENSORS & BIOELECTRONICS, 2015, 70 : 398 - 403
  • [29] Label-free Voltammetric Immunosensor for Prostate Specific Antigen Detection
    Oliveira, Nelia
    Costa-Rama, Estefania
    Viswanathan, Subramanian
    Delerue-Matos, Cristina
    Pereira, Lourdes
    Morais, Simone
    ELECTROANALYSIS, 2018, 30 (11) : 2604 - 2611
  • [30] Multilayered Mesoporous Composite Nanostructures for Highly Sensitive Label-Free Quantification of Cardiac Troponin-I
    Saeidi, Mohsen
    Amidian, Mohammad Ali
    Sheybanikashani, Sana
    Mahdavi, Hossein
    Alimohammadi, Homayoon
    Syedmoradi, Leila
    Mohandes, Fatemeh
    Zarrabi, Ali
    Tamjid, Elnaz
    Omidfar, Kobra
    Simchi, Abdolreza
    BIOSENSORS-BASEL, 2022, 12 (05):