Rapid Pacing Using the 0.035-in. Retrograde Left Ventricular Support Wire in 208 Cases of Transcatheter Aortic Valve Implantation and Balloon Aortic Valvuloplasty

被引:45
作者
Hilling-Smith, Roland [1 ]
Cockburn, James [1 ]
Dooley, Maureen [1 ]
Parker, Jessica [1 ]
Newton, Andrea [1 ]
Hill, Andrew [1 ]
Trivedi, Uday [1 ]
de Belder, Adam [1 ]
Hildick-Smith, David [1 ]
机构
[1] Brighton & Sussex Univ Hosp, Royal Sussex Cty Hosp, Sussex Cardiac Ctr, Brighton BN2 5BE, E Sussex, England
关键词
TAVI; TAVR; Temporary Pacing Wire; Balloon valvuloplasty; pacemaker; PERMANENT PACEMAKER; PREDICTORS; STENOSIS;
D O I
10.1002/ccd.26720
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction and Method: Transcatheter aortic valve implantation (TAVI) and balloon aortic valvuloplasty (BAV) are now well established percutaneous procedures. These procedures almost always require bursts of rapid ventricular pacing to temporarily reduce cardiac output to facilitate the procedure, usually done via a temporary pacing wire inserted into the right ventricle. We describe a case series of 132 cases of TAVI and 76 BAV done using ventricular pacing via the left ventricular lead by simply connecting one electrode to the patient's skin and one electrode through the left ventricular (LV) wire. Results: All of the 132 TAVI cases (a mixture of Edwards Sapien, Medtronic CoreValve and Boston Scientific Lotus) and 76 BAV were successfully performed using pacing through the LV wire. No BAV patients required temporary pacing wire (TPW) or permanent pacemaker (PPM) insertion. Of the TAVI patients, 6 (4.5%) required TPW during the procedure due to complete heart block to facilitate removal of the LV wire. 1 patient (0.8%) required a PPM urgently due to complete heart block and haemodynamic instability. Twenty eight Patients (21.2%) required PPM following TAVI, 9 of which were within the first 24 hr. Average time to pacemaker implantation was 3.7 days. Conclusion: Rapid ventricular pacing via the LV wire is a simple, safe and effective strategy for percutaneous aortic valve intervention and balloon aortic valvuloplasty. It eliminates the need for a temporary pacing wire with its attendant risks in the vast majority of cases. Furthermore, most pacemakers following TAVI are required late, after the first 24 hr period, by which time the TPW has already usually been removed. (C) 2016 Wiley Periodicals, Inc.
引用
收藏
页码:783 / 786
页数:4
相关论文
共 6 条
[1]   Predictors for New-Onset Complete Heart Block After Transcatheter Aortic Valve Implantation [J].
Bleiziffer, Sabine ;
Ruge, Hendrik ;
Hoerer, Juergen ;
Hutter, Andrea ;
Geisbuesch, Sarah ;
Brockmann, Gernot ;
Mazzitelli, Domenico ;
Bauernschmitt, Robert ;
Lange, Ruediger .
JACC-CARDIOVASCULAR INTERVENTIONS, 2010, 3 (05) :524-530
[2]   Transcatheter Aortic-Valve Implantation for Aortic Stenosis in Patients Who Cannot Undergo Surgery. [J].
Leon, Martin B. ;
Smith, Craig R. ;
Mack, Michael ;
Miller, D. Craig ;
Moses, Jeffrey W. ;
Svensson, Lars G. ;
Tuzcu, E. Murat ;
Webb, John G. ;
Fontana, Gregory P. ;
Makkar, Raj R. ;
Brown, David L. ;
Block, Peter C. ;
Guyton, Robert A. ;
Pichard, Augusto D. ;
Bavaria, Joseph E. ;
Herrmann, Howard C. ;
Douglas, Pamela S. ;
Petersen, John L. ;
Akin, Jodi J. ;
Anderson, William N. ;
Wang, Duolao ;
Pocock, Stuart .
NEW ENGLAND JOURNAL OF MEDICINE, 2010, 363 (17) :1597-1607
[3]  
McCann Peter, 2007, Indian Pacing Electrophysiol J, V7, P40
[4]   CORONARY PACING DURING PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY [J].
MEIER, B ;
RUTISHAUSER, W .
CIRCULATION, 1985, 71 (03) :557-561
[5]   Factors Predicting and Having an Impact on the Need for a Permanent Pacemaker After CoreValve Prosthesis Implantation Using the New Accutrak Delivery Catheter System [J].
Munoz-Garcia, Antonio J. ;
Hernandez-Garcia, Jose M. ;
Jimenez-Navarro, Manuel F. ;
Alonso-Briales, Juan H. ;
Dominguez-Franco, Antonio J. ;
Fernandez-Pastor, Julia ;
Pena Hernandez, Jose ;
Barrera Cordero, Alberto ;
Alzueta Rodriguez, Javier ;
de Teresa-Galvan, Eduardo .
JACC-CARDIOVASCULAR INTERVENTIONS, 2012, 5 (05) :533-539
[6]   Predictors of Permanent Pacemaker Implantation in Patients With Severe Aortic Stenosis Undergoing TAVR A Meta-Analysis [J].
Siontis, George C. M. ;
Jueni, Peter ;
Pilgrim, Thomas ;
Stortecky, Stefan ;
Buellesfeld, Lutz ;
Meier, Bernhard ;
Wenaweser, Peter ;
Windecker, Stephan .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2014, 64 (02) :129-140