Evaluation of high-performance liquid chromatography column retentivity using macromolecular probes.: II.: Silanophilic interactivity traced by highly polar polymers

被引:22
作者
Berek, D [1 ]
Tarbajovská, J [1 ]
机构
[1] Slovak Acad Sci, Inst Polymer, Lab Liquid Chromatog, Bratislava 84236, Slovakia
关键词
stationary phases; LC; macromolecular probes; column retentivity; poly(methyl methacrylate); polystyrene; polymers;
D O I
10.1016/S0021-9673(02)01150-0
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A new method of HPLC column retentivity testing utilizes polymeric probes instead of conventional sets of low molar mass substances. The procedure allows at least semiquantitative, separate and independent evaluation of adsorption and partition properties of column packings. In this present work, the method is applied for comparison of the polar interactivities of selected silica gel C-18 HPLC columns. It is shown that free silanols which remained on the surface of the end-capped silica C-18 column packings are accessible for interaction with highly polar macromolecules. High molar mass polymeric test probes are adsorbed on the surface silanols and their retention volumes increase. As result, deviations from regular size-exclusion chromatographic (SEC) behavior are observed. The extent of retention volume changes depends on both the nature of polymer probes and on column packing type. Adsorption of macromolecules can be suppressed by addition of a highly polar substance to the mobile phase. The amount of polar additive which is needed to attain regular SEC elution of the polymer probe depends on the column packing type and can be used as a characteristic of silanophilic column interactivity. Courses of dependences of retention volumes on sizes of macromolecules indicate the presence of "U-turn" adsorption which allows two and more silanols situated among C, groups to be occupied simultaneously with the same macromolecule. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:27 / 37
页数:11
相关论文
共 24 条
[1]  
Belenkii BG, 1996, ACS SYM SER, V635, P274
[2]   Coupled liquid chromatographic techniques for the separation of complex polymers [J].
Berek, D .
PROGRESS IN POLYMER SCIENCE, 2000, 25 (07) :873-908
[3]  
Berek D, 1998, J POLYM SCI POL CHEM, V36, P1363, DOI 10.1002/(SICI)1099-0518(19980715)36:9<1363::AID-POLA3>3.0.CO
[4]  
2-U
[5]  
BEREK D, 1975, MAKROMOL CHEM, V176, P391
[6]   Evaluation of high-performance liquid chromatography column retentivity using macromolecular probes I [J].
Berek, D .
JOURNAL OF CHROMATOGRAPHY A, 2002, 950 (1-2) :75-80
[7]  
Berek D., 1999, COLUMN HDB SIZE EXCL, P445
[8]   POLYMER-SUBSTRATE INTERACTIONS IN SIZE EXCLUSION CHROMATOGRAPHY WITH SILICA-GELS AND PURE SOLVENTS [J].
CHIANTORE, O .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1984, 7 (01) :1-11
[9]   Comparative study of test methods for reversed-phase columns for high-performance liquid chromatography [J].
Claessens, HA ;
van Straten, MA ;
Cramers, CA ;
Jezierska, M ;
Buszewski, B .
JOURNAL OF CHROMATOGRAPHY A, 1998, 826 (02) :135-156
[10]   Trends and progress in the characterization of stationary phases for reversed-phase liquid chromatography [J].
Claessens, HA .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2001, 20 (10) :563-583