Analytical control of an esterification batch reaction between glycerine and fatty acids by near-infrared spectroscopy

被引:17
作者
Blanco, M [1 ]
Beneyto, R
Castillo, M
Porcel, M
机构
[1] Univ Autonoma Barcelona, Dept Chem, Fac Sci, Bellaterra 08193, Barcelona, Spain
[2] ICI Spain SA, Dev & Qual Assurance Dept, Fogars De La Selva 08495, Barcelona, Spain
关键词
NIR; control esterification process; multivariate calibration;
D O I
10.1016/j.aca.2004.06.003
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Near-infrared spectroscopy was used to control an esterification reaction between glycerine and middle- or long-chain fatty acids performed in a laboratory-scale reactor. The process involves the initial formation of monoglycerides, which is followed by that of di- and triglycerides as well as transesterification. Establishing the end point of the process is critical with a view to ensuring that the end product will have the composition required for its intended use. PLS calibration was applied to industrial and laboratory-scale batch samples, and laboratory samples were additionally used to extend calibration ranges and avoid correlation between the concentration of the batch samples. In this way, PLS calibration models for glycerine, fatty acids, water, and mono-, di- and triglycerides, were constructed. The proposed method allows the reaction to be monitored in real time, thereby avoiding long analysis times, excessive reagent consumption and the obtainment of out-of-specitication products. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:143 / 148
页数:6
相关论文
共 9 条
[1]   On-line monitoring and quantification of a process reaction by near-infrared spectroscopy. Catalysed esterification of butan-1-ol by acetic acid [J].
Blanco, M ;
Serrano, D .
ANALYST, 2000, 125 (11) :2059-2064
[2]   Real time quantitation of a chemical reaction by fiber optic near-infrared spectroscopy [J].
Coffey, C ;
Cooley, BE ;
Walker, DS .
ANALYTICA CHIMICA ACTA, 1999, 395 (03) :335-341
[3]   Monitoring, of a batch organic synthesis by near-infrared spectroscopy:: modeling and interpretation of three-way data [J].
Geladi, P ;
Forsström, J .
JOURNAL OF CHEMOMETRICS, 2002, 16 (07) :329-338
[4]   Determination of acid value, hydroxyl Value and water content in reactions between dicarboxylic acids and diols using near-infrared spectroscopy and non-linear partial least squares regression [J].
Heikka, RA ;
Immonen, KT ;
Minkkinen, PO ;
Paatero, EYO ;
Salmi, TO .
ANALYTICA CHIMICA ACTA, 1997, 349 (1-3) :287-294
[5]   Response of flame ionization detectors to different homologous series [J].
Kállai, M ;
Veres, Z ;
Balla, J .
CHROMATOGRAPHIA, 2001, 54 (7-8) :511-517
[6]   Comparison of in-line NIR, Raman and UV-visible spectrometries, and at-line NMR spectrometry for the monitoring of an esterification reaction [J].
McGill, CA ;
Nordon, A ;
Littlejohn, D .
ANALYST, 2002, 127 (02) :287-292
[7]  
MILLER KL, 1994, P ANN S INSTRU PROCE, P53
[8]  
MOCKEL WD, 1992, P SPIE INT SOC OPTIC, P1681
[9]  
2002, DIRECTORATE QUALITY