Large volume injection in gas chromatography using the through oven transfer adsorption desorption interface operating under vacuum

被引:4
|
作者
Aragon, Alvaro [1 ]
Toledano, Rosa M. [1 ]
Gea, Sara [1 ]
Cortes, Jose M. [1 ]
Vazquez, Ana M. [2 ]
Villen, Jesus [1 ]
机构
[1] Univ Castilla La Mancha, Escuela Tecn Super Ingenieros Agronomos, Albacete 02071, Spain
[2] Univ Castilla La Mancha, Fac Educ, Albacete 02071, Spain
关键词
Large volume gas chromatography injection; Vacuum; Tenax; Polydimethylsiloxane; Through Oven Transfer Adsorption; Desorption interface; PHASE LIQUID-CHROMATOGRAPHY; PROGRAMMED-TEMPERATURE VAPORIZER; PESTICIDE-RESIDUE ANALYSIS; TOTAD INTERFACE; EDIBLE OILS; LC-GC; WATER; PRECOLUMNS;
D O I
10.1016/j.talanta.2014.01.064
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The present work describes a modification of the Through Oven Transfer Adsorption Desorption (TOTAD) interface, consisting of coupling a vacuum system to reduce the consumption of the helium needed to totally remove the eluent for large volume injection (LVI) in gas chromatography (GC). Two different retention materials in the liner of the TOTAD interface were evaluated: Tenax TA, which was seen to be unsuitable for working under vacuum conditions, and polydimethylsiloxane (PDMS), which provided satisfactory repeatability as well as a good sensitivity. No variability was observed in the retention times in either case. Solutions containing organophosphorous pesticides in two different solvents, a polar (methanol/water) and a non-polar (hexane) solvent, were used to evaluate the modification. The vacuum system coupled to the TOTAD interface allowed up to 90% helium to be saved without affecting the performance. (c) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:39 / 44
页数:6
相关论文
共 22 条
  • [21] Optimisation of programmable temperature vaporizer-based large volume injection for determination of pesticide residues in fruits and vegetables using gas chromatography-mass spectrometry
    Stajnbaher, Darinka
    Zupancic-Kralj, Lucija
    JOURNAL OF CHROMATOGRAPHY A, 2008, 1190 (1-2) : 316 - 326
  • [22] Determination of cyclic volatile methylsiloxanes in water, sediment, soil, biota, and biosolid using large-volume injection-gas chromatography-mass spectrometry
    Wang, De-Gao
    Alaee, Mehran
    Steer, Helena
    Tait, Tara
    Williams, Zackery
    Brimble, Samantha
    Svoboda, Lewina
    Barresi, Enzo
    DeJong, Maryl
    Schachtschneider, JoAnne
    Kaminski, Ed
    Norwood, Warren
    Sverko, Ed
    CHEMOSPHERE, 2013, 93 (05) : 741 - 748