Surface guided imaging during stereotactic radiosurgery with automated delivery

被引:28
作者
Covington, Elizabeth L. [1 ]
Stanley, Dennis N. [1 ]
Fiveash, John B. [1 ]
Thomas, Evan M. [1 ]
Marcrom, Samuel R. [1 ]
Bredel, Marcus [1 ]
Willey, Christopher D. [1 ]
Riley, Kristen O. [2 ]
Popple, Richard A. [1 ]
机构
[1] Univ Alabama Birmingham, Dept Radiat Oncol, Birmingham, AL 35294 USA
[2] Univ Alabama Birmingham, Dept Neurosurg, Birmingham, AL USA
关键词
intrafraction motion; optical surface imaging; SGRT; stereotactic radiosurgery; surface guided radiotherapy; INSPIRATION BREATH-HOLD; CLINICAL-EXPERIENCE; CANCER PATIENTS; REAL-TIME; RADIOTHERAPY; FRAMELESS; ACCURACY;
D O I
10.1002/acm2.13066
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To report on the use of surface guided imaging during frameless intracranial stereotactic radiotherapy with automated delivery via HyperArc(TM) (Varian Medical Systems, Palo Alto, CA). Methods All patients received intracranial radiotherapy with HyperArc(TM) and were monitored for intrafraction motion by the AlignRT (R) (VisionRT, London, UK) surface imaging (SI) system. Immobilization was with the Encompass(TM) (Qfix, Avondale, PA) aquaplast mask device. AlignRT (R) log files were correlated with trajectory log files to correlate treatment parameters with SI reported offsets. SI reported offsets were correlated with gantry angle and analyzed for performance issues at non-zero couch angles and during camera-pod blockage during gantry motion. Demographics in the treatment management system were used to identify race and determine if differences in SI reported offsets are due to skin tone settings. Results A total of 981 fractions were monitored over 14 months and 819 were analyzed. The median AlignRT (R) reported motion from beginning to the end of treatment was 0.24 mm. The median offset before beam on at non-zero couch angles was 0.55 mm. During gantry motion when camera pods are blocked, the median magnitude was below 1 mm. Median magnitude of offsets at non-zero couch angles was not found to be significantly different for patients stratified by race. Conclusions Surface image guidance is a viable alternative to scheduled mid-treatment imaging for monitoring intrafraction motion during stereotactic radiosurgery with automated delivery.
引用
收藏
页码:90 / 95
页数:6
相关论文
共 18 条
[1]   Accuracy Evaluation of a 3-Dimensional Surface Imaging System for Guidance in Deep-Inspiration Breath-Hold Radiation Therapy [J].
Alderliesten, Tanja ;
Sonke, Jan-Jakob ;
Betgen, Anja ;
Honnef, Joeri ;
van Vliet-Vroegindeweij, Corine ;
Remeijer, Peter .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 85 (02) :536-542
[2]   Assessment of set-up variability during deep inspiration breath hold radiotherapy for breast cancer patients by 3D-surface imaging [J].
Betgen, Anja ;
Alderliesten, Tanja ;
Sonke, Jan-Jakob ;
van Vliet-Vroegindeweij, Corine ;
Bartelink, Harry ;
Remeijer, Peter .
RADIOTHERAPY AND ONCOLOGY, 2013, 106 (02) :225-230
[3]   Initial clinical experience with a frameless and maskless stereotactic radiosurgery treatment [J].
Cervino, Laura I. ;
Detorie, Nicole ;
Taylor, Matthew ;
Lawson, Joshua D. ;
Harry, Taylor ;
Murphy, Kevin T. ;
Mundt, Arno J. ;
Jiang, Steve B. ;
Pawlicki, Todd A. .
PRACTICAL RADIATION ONCOLOGY, 2012, 2 (01) :54-62
[4]   Evaluation of radiotherapy setup accuracy for head and neck cancer using a 3-D surface imaging system [J].
Cho, H. -L. ;
Park, E. -T. ;
Kim, J. -Y. ;
Kwak, K. -S. ;
Kim, C. -J. ;
Ahn, K. -J. ;
Suh, T. -S. ;
Lee, Y. -K. ;
Kim, S. -W. ;
Kim, J. -K. ;
Lim, S. ;
Choi, Y. -M. ;
Park, S. -K. .
JOURNAL OF INSTRUMENTATION, 2013, 8
[5]   Optical surface guidance for submillimeter monitoring of patient position during frameless stereotactic radiotherapy [J].
Covington, Elizabeth L. ;
Fiveash, John B. ;
Wu, Xingen ;
Brezovich, Ivan ;
Willey, Christopher D. ;
Riley, Kristen ;
Popple, Richard A. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (06) :91-98
[6]   Dosimetric effects of intrafractional isocenter variation during deep inspiration breath-hold for breast cancer patients using surface-guided radiotherapy [J].
Kugele, Malin ;
Edvardsson, Anneli ;
Berg, Lovisa ;
Alkner, Sara ;
Ljus, Carina Andersson ;
Ceberg, Sofie .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01) :25-38
[7]   Clinical efficacy and safety of surface imaging guided radiosurgery (SIG-RS) in the treatment of benign skull base tumors [J].
Lau, Steven K. M. ;
Patel, Kunal ;
Kim, Teddy ;
Knipprath, Erik ;
Kim, Gwe-Ya ;
Cervino, Laura I. ;
Lawson, Joshua D. ;
Murphy, Kevin T. ;
Sanghvi, Parag ;
Carter, Bob S. ;
Chen, Clark C. .
JOURNAL OF NEURO-ONCOLOGY, 2017, 132 (02) :307-312
[8]   Motion monitoring for cranial frameless stereotactic radiosurgery using video-based three-dimensional optical surface imaging [J].
Li, Guang ;
Ballangrud, Ase ;
Kuo, Li Cheng ;
Kang, Hyejoo ;
Kirov, Assen ;
Lovelock, Michael ;
Yamada, Yoshiya ;
Mechalakos, James ;
Amols, Howard .
MEDICAL PHYSICS, 2011, 38 (07) :3981-3994
[9]   Real-time 3D-surface-guided head refixation useful for fractionated stereotactic radiotherapy [J].
Li, SD ;
Liu, DZ ;
Yin, GJ ;
Zhuang, P ;
Geng, J .
MEDICAL PHYSICS, 2006, 33 (02) :492-503
[10]   Assessment of interfractional variation of the breast surface following conventional patient positioning for whole-breast radiotherapy [J].
Padilla, Laura ;
Kang, Hyejoo ;
Washington, Maxine ;
Hasan, Yasmin ;
Chmura, Steve J. ;
Al-Hallaq, Hania .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (05) :177-189