Potentiometric differential microdetector with interchange solid membranes used in flow injection analysis

被引:4
作者
Olenic, L
Hopirtean, E
Olenic, L
机构
[1] Institute of Chemistry 'Raluca Ripan', 3400 Cluj-Napoca, Fîntînele str.
关键词
flow injection analysis; ion-selective membranes for silver(I); copper(II) and sulfide; flow-through electrode; potentiometric differential microdetector;
D O I
10.1039/a605410b
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The construction and characteristics of two potentiometric differential microdetectors equipped with two similar ion-selective membranes are described, The two membranes used were of Ag2S for Ag-I and S2- and CuS-Ag2S for Cu-II. The active volume of the measuring cell was about 3 mu l. Calibration plots were linear, with almost theoretical slopes, down to 10(-6) mol dm(-3) Ag-I, 5 x 10(-6) mol dm(-3) Cu-II and 10(-5) mol dm(-3) S2-. The throughput was 30-60 samples h(-1) depending on the sample. The errors in the determination of the above-mentioned ions were below 3.9%.
引用
收藏
页码:107 / 109
页数:3
相关论文