A newly released processless Gafchromic EBT3 film was investigated for its response to ultraviolet radiation (UVR) at three different wavelengths: 365 (UVA), 302 (UVB) and 254nm (UVC). The change in the film's optical density was dependent on the radiance exposure and UV wavelength and was found to correlate with the UV intensity. The dynamic range of the EBT3 films is similar to 5 to similar to 60J/cm(2) for UVA and UVB and similar to 5 to similar to 300J/cm(2) for UVC. The dose sensitivity of EBT3 films for UVA and UVB radiation was approximately 10 times higher than that for UVC radiation. The results indicate the suitability of EBT3 films for quantitative and qualitative measurements of UVR exposure.
引用
收藏
页码:249 / 255
页数:7
相关论文
共 22 条
[1]  
Abdel-Rehim F., 1992, J PHOTOCH PHOTOBIO A, V64, P123
[2]   Sunscreen use and intentional exposure to ultraviolet A and B radiation:: a double blind randomized trial using personal dosimeters [J].
Autier, P ;
Doré, JF ;
Reis, AC ;
Grivegnée, A ;
Ollivaud, L ;
Truchetet, F ;
Chamoun, E ;
Rotmensz, N ;
Severi, G ;
Césarini, JP .
BRITISH JOURNAL OF CANCER, 2000, 83 (09) :1243-1248
[3]   Measuring solar UV radiation with EBT radiochromic film [J].
Butson, Ethan T. ;
Cheung, Tsang ;
Yu, Peter K. N. ;
Butson, Martin J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (20) :N487-N493
[4]   Ultraviolet radiation dosimetry with radiochromic film [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN ;
Abbati, D ;
Greenoak, GE .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (07) :1863-1868
[5]   Radiochromic film for medical radiation dosimetry [J].
Butson, MJ ;
Yu, PKN ;
Cheung, T ;
Metcalfe, P .
MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2003, 41 (3-5) :61-120
[6]  
CASALE GR, 2009, BIOPHYS BIOENG LETT, V2, P1
[7]   PERSONNEL MONITORING OF EXPOSURE TO UV-RADIATION [J].
CHALLONER, AVJ ;
CORLESS, D ;
DAVIS, A ;
DEANE, GHW ;
DIFFEY, BL ;
GUPTA, SP ;
MAGNUS, IA .
CLINICAL AND EXPERIMENTAL DERMATOLOGY, 1976, 1 (02) :175-179
[8]   Dosimetric properties of improved GafChromic films for seven different digitizers [J].
Devic, S ;
Seuntjens, J ;
Hegyi, G ;
Podgorsak, EB ;
Soares, CG ;
Kirov, AS ;
Ali, I ;
Williamson, JF ;
Elizondo, A .
MEDICAL PHYSICS, 2004, 31 (09) :2392-2401
[9]   Sources and measurement of ultraviolet radiation [J].
Diffey, BL .
METHODS, 2002, 28 (01) :4-13
[10]   ULTRAVIOLET-RADIATION PHYSICS AND THE SKIN [J].
DIFFEY, BL .
PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (03) :405-426