Changes in the electrical properties of the electrode-skin-underlying tissue composite during a week-long programme of neuromuscular electrical stimulation

被引:21
作者
Birlea, S. I. [1 ,2 ]
Breen, P. P. [1 ,2 ]
Corley, G. J. [1 ,2 ]
Birlea, N. M. [3 ]
Quondamatteo, F. [4 ]
OLaighin, G. [1 ,2 ]
机构
[1] NUI Galway, Sch Engn & Informat, Galway, Ireland
[2] NUI Galway, Natl Ctr Biomed Engn Sci, Galway, Ireland
[3] Tech Univ Cluj Napoca, Dept Phys, Cluj Napoca 400020, Romania
[4] NUI Galway, Sch Med, Galway, Ireland
关键词
tissue impedance; human skin electrical properties; stratum corneum; electrode-skin interface; neuromuscular electrical stimulation; DIELECTRIC-PROPERTIES; BLOOD-FLOW; PHYSIOLOGICAL-RESPONSES; SHOULDER PAIN; FREQUENCY; IMPEDANCE; HEMIPLEGIA; RESISTANCE; THICKNESS; COMFORT;
D O I
10.1088/0967-3334/35/2/231
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Particular neuromuscular electrical stimulation (NMES) applications require the use of the same electrodes over a long duration (>1 day) without having access to them. Under such circumstance the quality of the electrode-skin contact cannot be assessed. We used the NMES signal itself to assess the quality of the electrode-skin contact and the electrical properties of the underlying tissues over a week. A 14% decrease in the skin's stratum corneum resistance (from 20 to 17 k Omega) and a 15% decrease in the resistance of the electrodes and underlying tissues (from 550 to 460 Omega) were observed in the 14 healthy subjects investigated. A follow-on investigation of the effect of exercise-induced sweating on the electrical properties of the electrode-skin-underlying tissue composite during NMES indicated a correlation between the decrease in the resistance values observed over the course of the week and the accumulation of sweat at the electrode-skin interface. The value of the capacitance representing the dielectric properties of the skin's stratum corneum increased after exercise-induced sweating but did not change significantly over the course of the week. We conclude that valuable information about the electrode-skin-underlying tissue composite can be gathered using the NMES signal itself, and suggest that this is a practical, safe and relatively simple method for monitoring these electrical properties during long-term stimulation.
引用
收藏
页码:231 / 252
页数:22
相关论文
共 72 条
  • [21] The effect of surface neuromuscular electrical stimulation and compression hosiery applied to the lower limb, on the comfort and blood flow of healthy subjects
    Corley, Gavin J.
    Breen, Paul P.
    Grace, Pierce A.
    OLaighin, Gearoid
    [J]. 2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8, 2008, : 703 - +
  • [22] Effect of temperature and sweating on bioimpedance measurements
    Cornish, BH
    Thomas, BJ
    Ward, LC
    [J]. APPLIED RADIATION AND ISOTOPES, 1998, 49 (5-6) : 475 - 476
  • [23] Visual imaging of ion distribution in human epidermis
    Denda, M
    Hosoi, J
    Asida, Y
    [J]. BIOCHEMICAL AND BIOPHYSICAL RESEARCH COMMUNICATIONS, 2000, 272 (01) : 134 - 137
  • [24] Dorgan S J, 1999, IEEE Trans Rehabil Eng, V7, P341, DOI 10.1109/86.788470
  • [25] THE EFFECTS OF NEUROMUSCULAR STIMULATION ON MUSCLE TONE AT THE KNEE IN PARAPLEGIA
    DOUGLAS, AJ
    WALSH, EG
    WRIGHT, GW
    CREASEY, GH
    EDMOND, P
    [J]. EXPERIMENTAL PHYSIOLOGY, 1991, 76 (03) : 357 - 367
  • [26] In vivo estimation of stratum corneum thickness from water concentration profiles obtained with Raman spectroscopy
    Egawa, Mariko
    Hirao, Tetsuji
    Takahashi, Motoji
    [J]. ACTA DERMATO-VENEREOLOGICA, 2007, 87 (01) : 4 - 8
  • [27] DIELECTRIC PROPERTIES OF THE HUMAN BODY FOR WAVE-LENGTHS IN THE 1-10 CM. RANGE
    ENGLAND, TS
    [J]. NATURE, 1950, 166 (4220) : 480 - 480
  • [28] IONTOPHORETIC DELIVERY OF DRUGS THROUGH MEMBRANES INCLUDING HUMAN STRATUM-CORNEUM
    FOLEY, D
    CORISH, J
    CORRIGAN, OI
    [J]. SOLID STATE IONICS, 1992, 53 : 184 - 196
  • [29] The dielectric properties of biological tissues .2. Measurements in the frequency range 10 Hz to 20 GHz
    Gabriel, S
    Lau, RW
    Gabriel, C
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) : 2251 - 2269
  • [30] Control of urinary bladder function with devices: successes and failures
    Gaunt, RA
    Prochazka, A
    [J]. AUTONOMIC DYSFUNCTION AFTER SPINAL CORD INJURY, 2006, 152 : 163 - 194