Determination of trace elements in residual oil by high-resolution inductively coupled plasma mass spectrometry

被引:42
作者
Xie, Hualin [1 ,2 ]
Huang, Kelong [1 ]
Liu, Jinchun [2 ]
Nie, Xidu [2 ]
Fu, Liang [3 ]
机构
[1] Cent S Univ, Sch Chem & Chem Engn, Changsha 410083, Peoples R China
[2] Hunan Inst Technol, Dept Chem, Hengyang 421008, Peoples R China
[3] Nanhua Univ, Coll Chem & Chem Engn, Hengyang 410001, Peoples R China
关键词
Residual oil; HR-ICP-MS; Trace elements; Spectral interferences; Matrix effect; IMPURITIES;
D O I
10.1007/s00216-009-2658-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An analytical method using high-resolution inductively coupled plasma mass spectrometry (HR-ICP-MS) for rapid simultaneous determination of 20 elements, including Na, Mg, Al, Ca, Ti, V, Mn, Fe, Co, Ni, Cu, Zn, Ga, As, Mo, Cd, Sn, Sb, Ba, and Pb elements, in residual oil was described. The sample was dissolved in HNO3 by microwave digestion, and then the above 20 elements in the solution were analyzed directly by HR-ICP-MS. Most of the spectral interferences could be avoided by measuring in the high-resolution mode. The matrix effect caused by the sample-digesting solution and corrected by Sc, Rh, and Bi as the internal standard elements was studied in detail. The optimum condition of the determination was also tested and discussed. The result showed that the detection limits of the method were in the range of 0.014 to 11.6 mu g L-1; the relative standard deviation was less than 3.8% and recoveries in the samples were in the range of 88.4% to 108.0%. This method can be used to determine the trace elements in residual oil with the features of accurate, rapid, and convenient determination.
引用
收藏
页码:2075 / 2080
页数:6
相关论文
共 10 条
[1]  
Hu JY, 2002, ANAL CHIM ACTA, V451, P329
[2]   PRE-CONCENTRATION AND X-RAY-FLUORESCENCE DETERMINATION OF VANADIUM, NICKEL AND IRON IN RESIDUAL FUEL OILS AND IN PARTICULATE MATERIAL FROM OIL-FIRED SOURCES [J].
IWASAKI, K ;
TANAKA, K .
ANALYTICA CHIMICA ACTA, 1982, 136 (APR) :293-299
[3]  
MARIELA N, 2005, SPECTROCHIM ACTA, V60, P615
[4]  
MAURIZIO B, 1995, J ANAL ATOM SPECTROM, V10, P555
[5]  
MIGUEL M, 1995, J ANAL ATOM SPECTROM, V9, P237
[6]   Determination of trace impurities in high-purity aluminium oxide by high resolution inductively coupled plasma mass spectrometry [J].
Nakane, K ;
Uwamino, Y ;
Morikawa, H ;
Tsuge, A ;
Ishizuka, T .
ANALYTICA CHIMICA ACTA, 1998, 369 (1-2) :79-85
[7]   THE USE OF INTERNAL STANDARDS IN ICP MS [J].
VANHAECKE, F ;
VANHOE, H ;
DAMS, R ;
VANDECASTEELE, C .
TALANTA, 1992, 39 (07) :737-742
[8]   Microwave digestion of "residual fuel oil" (NIST SRM 1634b) for the determination of trace elements by inductively coupled plasma-mass spectrometry [J].
Wondimu, T ;
Goessler, W ;
Irgolic, KJ .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 367 (01) :35-42
[9]  
Xie HL, 2006, CHINESE CHEM LETT, V17, P1077
[10]   Determination of trace impurities in high-purity cadmium by high-resolution inductively coupled plasma mass spectrometry [J].
Xie, Hualin ;
Nie, Xidu .
ANALYTICAL SCIENCES, 2006, 22 (10) :1371-1374