Rapid quantification of steroid patterns in human serum by on-line solid phase extraction combined with liquid chromatography-triple quadrupole linear ion trap mass spectrometry

被引:99
作者
Ceglarek, Uta [1 ]
Kortz, Linda [1 ]
Leichtle, Alexander [1 ]
Fiedler, Georg M. [1 ]
Kratzsch, Juergen [1 ]
Thiery, Joachim [1 ]
机构
[1] Univ Hosp Leipzig, Inst Lab Med Clin Chem & Mol Diagnost, D-04103 Leipzig, Germany
关键词
Steroids; Steroid patterns; On-line SPE-LC-MS/MS; Linear ion trap; Mass spectrometry; TESTOSTERONE; IMMUNOASSAY; METABOLITES; ESTROGENS; PROFILES;
D O I
10.1016/j.cca.2008.11.022
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The determination of steroids is important for the diagnosis and monitoring of endocrine diseases and infertility workup. We developed a rapid and reliable mass spectrometric method for the simultaneous quantification of steroid patterns in human serum. Methods: An on-line solid phase extraction (SPE)-liquid chromatography-triple quadrupole linear ion trap (LCQ-Trap) method utilizing atmospheric pressure chemical ionization was developed. Following protein precipitation of 100 mu L serum, on-line SPE and chromatographic separation was performed for 13 steroids in 1.8 min. Analytes were confirmed by the characteristic fragment patterns. Results:The total run time of the method was 4 min. Detection limits ranged between 0.02 mu g/L(testosterone) and 9 mu g/L (dehydroepiandrosterone sulfate). The method was linear up to 7000 mu g/L for dehydroepiandrosterone sulfate, 500 mu g/L for cortisol, 125 mu g/L for 11-deoxycortisol, and 25 mu g/L for aldosterone, 17-hydroxyprogesterone, progesterone, testosterone, androstenedione and beta-estradiol, respectively. Accuracy ranged between 80 and 114% Between-day variance at three different concentration levels was <15% Excellent correlations with immunoassays were observed for testosterone, cortisol and beta-estradiol with Pearson's correlation coefficient r=0.967,0.963, and 0.998, respectively. Conclusion: The novel on-line SPE-LC-MS/QTrap platform offers a very fast, reliable, and sensitive quantification of steroid patterns and fulfils the quality criteria for routine laboratory application. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:114 / 118
页数:5
相关论文
共 18 条
  • [1] Ion suppression in mass spectrometry
    Annesley, TM
    [J]. CLINICAL CHEMISTRY, 2003, 49 (07) : 1041 - 1044
  • [2] The new era of automated immunoassay
    Bock, JL
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2000, 113 (05) : 628 - 646
  • [3] Guo TD, 2004, ARCH PATHOL LAB MED, V128, P469
  • [4] Rapid measurement of estrogens and their metabolites in human serum by liquid chromatography-tandem mass spectrometry without derivatization
    Guo, Tiedong
    Gu, Jianghong
    Soldin, Offie P.
    Singh, Ravinder J.
    Soldin, Steven J.
    [J]. CLINICAL BIOCHEMISTRY, 2008, 41 (09) : 736 - 741
  • [5] Simultaneous determination of 12 steroids by isotope dilution liquid chromatography-photo spray ionization tandem mass spectrometry
    Guo, Tiedong
    Taylor, Robert L.
    Singh, Ravinder J.
    Soldin, Steven J.
    [J]. CLINICA CHIMICA ACTA, 2006, 372 (1-2) : 76 - 82
  • [6] Immunoassays for testosterone in women: Better than a guess?
    Herold, DA
    Fitzgerald, RL
    [J]. CLINICAL CHEMISTRY, 2003, 49 (08) : 1250 - 1251
  • [7] Use of steroid profiles in determining the cause of adrenal insufficiency
    Holst, Jennifer P.
    Soldin, Steven J.
    Tractenberg, Rochelle E.
    Guo, Tiedong
    Kundra, Priya
    Verbalis, Joseph G.
    Jonklaas, Jacqueline
    [J]. STEROIDS, 2007, 72 (01) : 71 - 84
  • [8] Steroid hormones: relevance and measurement in the clinical laboratory
    Holst, JP
    Soldin, OP
    Guo, TD
    Soldin, SJ
    [J]. CLINICS IN LABORATORY MEDICINE, 2004, 24 (01) : 105 - +
  • [9] Interference in immunoassay is an underestimated problem
    Ismail, AAA
    Walker, PL
    Cawood, ML
    Barth, JH
    [J]. ANNALS OF CLINICAL BIOCHEMISTRY, 2002, 39 : 366 - 373
  • [10] Kao PC, 2001, ANN CLIN LAB SCI, V31, P199