Template based Attenuation Correction for PET in MR-PET Scanners

被引:0
|
作者
Kops, Elena Rota [1 ]
Herzog, Hans [1 ]
机构
[1] Forschungszentrum Julich, Julich, Germany
关键词
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This work investigates a template based procedure for attenuation correction (TBA) of PET scans acquired in future hybrid MR-PET scanners which will not offer a measured attenuation correction. A previous report of our group described a method (TBA-SPM) how individual attenuation maps can be obtained from an attenuation template which is spatially normalized to the SPM2 standard brain shape. Attenuation maps of females and males obtained from PET transmission scans were used as input for this template. The present study replaces the template referring to SPM2 by a female and a male attenuation template (fAT and mAT), each based on four measured attenuation images (TBA-f&m). The corresponding T1-MR templates (fMR and mMR) were also available. Thus, possible morphological gender-related differences, not considered when the standardized SPM2 (1)brain shape is used, may be taken into account. To examine this approach PET scans of 15 female and 15 male subjects of an ongoing study were attenuation corrected using the templates fAT and mAT. For this purpose and depending on the subject's gender the fMR or mMR templates were warped onto the individual MR image. The resulting warping matrix was applied to fAT or mAT, respectively. The individualized attenuation maps were used to reconstruct the PET emission data. These were compared to PET images attenuation corrected with the conventional PET based transmission data (PBA). While the relative differences between PBA and TBA-f&m reconstructed images averaged over each group and all regions of interest were 0.57% +/- 3.76% for females and -0.59% +/- 3-56% for males, the corresponding values obtained with the TBA-SPM method showed an overestimation with similar standard deviations (2.39% +/- 3.76% for females and 2.42% +/- 3.37% for males). In conclusion, the alter-native gender-related template method TBA-f&m gives acceptable results with no significant differences between the genders.
引用
收藏
页码:3061 / 3064
页数:4
相关论文
共 50 条
  • [1] Template based attenuation correction for brain MR-PET scanners
    Kops, E. Rota
    Herzog, H.
    JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 2009, 29 : S337 - S337
  • [2] Alternative methods for attenuation correction for PET images in MR-PET scanners
    Kops, Elena Rota
    Herzog, Hans
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4327 - 4330
  • [3] Attenuation Correction in MR-PET Scanners with Segmented T1-weighted MR Images
    Kops, Elena Rota
    Wagenknecht, Gudrun
    Scheins, Juergen
    Tellmann, Lutz
    Herzog, Hans
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 2530 - +
  • [4] Comparison template-based with CT-based attenuation correction for hybrid MR/PET scanners
    Elena Rota Kops
    Hans Herzog
    N Jon Shah
    EJNMMI Physics, 1 (Suppl 1)
  • [5] Hybrid approach for attenuation correction in PET/MR scanners
    Santos Ribeiro, A.
    Mops, E. Rota
    Herzog, H.
    Almeida, P.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2014, 734 : 166 - 170
  • [6] Comparison of Template-Based Versus CT-Based Attenuation Correction for Hybrid MR/PET Scanners
    Kops, Elena Rota
    Hautzel, Hubertus
    Herzog, Hans
    Antoch, Gerald
    Shah, N. Jon
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (05) : 2115 - 2121
  • [7] Value of a Dixon based MR-PET attenuation correction sequence for the localization and evaluation of PET positive lesions
    Eiber, Matthias
    Martinez-Moeller, Axel
    Souvatzoglou, Michael
    Rummeny, Ernst
    Schwaiger, Markus
    Beer, Ambros
    Nekolla, Stephan
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [8] Errors in MR-based attenuation correction for brain imaging with PET/MR scanners
    Kops, Elena Rota
    Herzog, Hans
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 702 : 104 - 107
  • [9] Attenuation Correction Synthesis for Hybrid PET-MR Scanners
    Burgos, Ninon
    Cardoso, Manuel Jorge
    Modat, Marc
    Pedemonte, Stefano
    Dickson, John
    Barnes, Anna
    Duncan, John S.
    Atkinson, David
    Arridge, Simon R.
    Hutton, Brian F.
    Ourselin, Sebastien
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION (MICCAI 2013), PT I, 2013, 8149 : 147 - 154
  • [10] Attenuation correction for hybrid MR/PET scanners: a comparison study
    Elena Rota Kops
    Andre Santos Ribeiro
    Liliana Caldeira
    Hubertus Hautzel
    Mathias Lukas
    Gerald Antoch
    Christoph Lerche
    Jon Shah
    EJNMMI Physics, 2 (Suppl 1)