Target position uncertainty during visually guided deep-inspiration breath-hold radiotherapy in locally advanced lung cancer

被引:33
作者
Rydhog, Jonas Scherman [1 ,2 ]
de Blanck, Steen Riisgaard [1 ]
Josipovic, Mirjana [1 ,2 ]
Jolck, Rasmus Irming [3 ,4 ]
Larsen, Klaus Richter [5 ]
Clementsen, Paul [6 ,7 ]
Andersen, Thomas Lars [3 ]
Poulsen, Per Rugaard [8 ]
Persson, Gitte Fredberg [1 ]
af Rosenschold, Per Munck [1 ,2 ]
机构
[1] Rigshosp, Sect Radiotherapy, Dept Oncol, DK-3994 Copenhagen, Denmark
[2] Univ Copenhagen, Niels Bohr Inst, Copenhagen, Denmark
[3] Tech Univ Denmark, Ctr Nanomed & Theranost, Dept Micro and Nanotechnol, DTU Nanotech, Lyngby, Denmark
[4] Nanovi Radiotherapy AS, Lyngby, Denmark
[5] Bispebjerg Hosp, Dept Clin Med, Copenhagen, Denmark
[6] Zealand Univ Hosp, Dept Internal Med, Roskilde, Denmark
[7] Rigshosp, CAMES, Copenhagen, Denmark
[8] Aarhus Univ Hosp, Dept Oncol, Aarhus, Denmark
关键词
Image-guided radiotherapy; Liquid fiducial marker; Image based tracking; RESPIRATORY MOTION; CONTROL ABC; LYMPH-NODE; TUMOR; INTERFRACTION; INTRAFRACTION; MANAGEMENT; PHANTOM; MARKER;
D O I
10.1016/j.radonc.2017.02.003
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The purpose of this study was to estimate the uncertainty in voluntary deep-inspiration breath hold (DISH) radiotherapy for locally advanced non-small cell lung cancer (NSCLC) patients. Methods: Perpendicular fluoroscopic movies were acquired in free breathing (FB) and DIBH during a course of visually guided DIBH radiotherapy of nine patients with NSCLC. Patients had liquid markers injected in mediastinal lymph nodes and primary tumours. Excursion, systematic- and random errors, and inter-breath-hold position uncertainty were investigated using an image based tracking algorithm. Results: A mean reduction of 2-6 mm in marker excursion in DIBH versus FB was seen in the anterior posterior (AP), left-right (LR) and cranio-caudal (CC) directions. Lymph node motion during DIBH originated from cardiac motion. The systematic-(standard deviation (SD) of all the mean marker positions) and random errors (root-mean-square of the intra-BH SD) during DIBH were 0.5 and 0.3 mm (AP), 0.5 and 0.3 mm (LR), 0.8 and 0.4 mm (CC), respectively. The mean inter-breath-hold shifts were -0.3 mm (AP), -0.2 mm (LR), and -0.2 mm (CC). Conclusion: Intra- and inter-breath-hold uncertainty of tumours and lymph nodes were small in visually guided breath-hold radiotherapy of NSCLC. Target motion could be substantially reduced, but not eliminated, using visually guided DIBH. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:78 / 84
页数:7
相关论文
共 29 条
  • [1] [Anonymous], 1993, 50 ICRU OXF U PRESS
  • [2] [Anonymous], 1999, 62 ICRU OXF U PRESS
  • [3] Assessment of gross tumor volume regression and motion changes during radiotherapy for non-small-cell lung cancer as measured by four-dimensional computed tomography
    Britton, Keith R.
    Starkschall, George
    Tucker, Susan L.
    Pan, Tinsu
    Nelson, Christopher
    Chang, Joe Y.
    Cox, James D.
    Mohan, Radhe
    Komaki, Ritsuko
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (04): : 1036 - 1046
  • [4] THE USE OF THE ACTIVE BREATHING COORDINATOR THROUGHOUT RADICAL NON SMALL-CELL LUNG CANCER (NSCLC) RADIOTHERAPY
    Brock, Juliet
    McNair, Helen A.
    Panakis, Niki
    Symonds-Tayler, Richard
    Evans, Phil M.
    Brada, Michael
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : 369 - 375
  • [5] INTERFRACTION AND INTRAFRACTION CHANGES IN AMPLITUDE OF BREATHING MOTION IN STEREOTACTIC LIVER RADIOTHERAPY
    Case, Robert B.
    Moseley, Douglas J.
    Sonke, Jan Jakob
    Eccles, Cynthia L.
    Dinniwell, Robert E.
    Kim, John
    Bezjak, Andrea
    Milosevic, Michael
    Brock, Kristy K.
    Dawson, Laura A.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 77 (03): : 918 - 925
  • [6] Respiratory motion-management in stereotactic body radiation therapy for lung cancer - A dosimetric comparison in an anthropomorphic lung phantom (LuCa)
    Ehrbar, Stefanie
    Perrin, Rosalind
    Peroni, Marta
    Bernatowicz, Kinga
    Parkel, Thomas
    Pytko, Izabela
    Klock, Stephan
    Guckenberger, Matthias
    Tanadini-Lang, Stephanie
    Weber, Damien Charles
    Lomax, Antony
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 121 (02) : 328 - 334
  • [7] ANATOMIC AND PATHOLOGIC VARIABILITY DURING RADIOTHERAPY FOR A HYBRID ACTIVE BREATH-HOLD GATING TECHNIQUE
    Glide-Hurst, Carri K.
    Gopan, Ellen
    Hugo, Geoffrey D.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 77 (03): : 910 - 917
  • [8] Deep inspiration breath-hold technique for lung tumors: The potential value of target immobilization and reduced lung density in dose escalation
    Hanley, J
    Debois, MM
    Mah, D
    Mageras, GS
    Raben, A
    Rosenzweig, K
    Mychalczak, B
    Schwartz, LH
    Gloeggler, PJ
    Lutz, W
    Ling, CC
    Leibel, SA
    Fuks, Z
    Kutcher, GJ
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (03): : 603 - 611
  • [9] ICRU, 2010, J. ICRU, V10, DOI [10.1093/jicru_ndq002, DOI 10.1093/JICRU_NDQ002, 10.1093/jicrundq002, DOI 10.1093/JICRUNDQ002]
  • [10] Deep inspiration breath-hold radiotherapy for lung cancer: impact on image quality and registration uncertainty in cone beam CT image guidance
    Josipovic, Mirjana
    Persson, Gitte F.
    Bangsgaard, Jens P.
    Specht, Lena
    Aznar, Marianne C.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1068)