Quality assessment of localization technique performance in small volume in vivo 1H MR spectroscopy

被引:10
作者
Burtscher, IM [1 ]
Johansson, E
Holtås, S
Ståhlberg, F
机构
[1] Univ Lund Hosp, Dept Radiol, S-22185 Lund, Sweden
[2] Univ Lund Hosp, Dept Radiat Phys, S-22185 Lund, Sweden
关键词
proton MRS; phantom; localization techniques; quality assessment;
D O I
10.1016/S0730-725X(99)00076-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new phantom and evaluation method for experimental evaluation of H-1-magnetic resonance spectroscopy single volume localization techniques regarding signal contamination (C), defined as the part of the signal originating outside the volume of interest, is presented. The quality assessment method is based on a spherical phantom with an oil/water interface in order to reduce susceptibility effects, and applied for stimulated-echo acquisition method (STEAM) and spin-echo (SE) sequences, echo times of 270, 135, and 10 ms, and cubic volumes of interest (VOI) of 1(3), 1.5(3), 2(3), 2.5(3), and 3(3) cm(3), To be able to mimic measurements of the contamination in three dimensions the physical gradients representing the three orthogonal directions for slice selection were shifted in the pulse sequences. Contamination values in one dimension differed between 6.5% and 8.4% in SE sequences, and between 0.7% and 13.8% in STEAM sequences; In STEAM sequences a decrease of C with increasing VOI size was observed while SE sequences showed comparable C values for the different VOI sizes tested. The total contamination in three dimensions were 19% and 18% in SE and STEAM sequences with a TE of 270 ms, and 7% in a STEAM sequence with a TE of 10 ms, respectively; The presented evaluation method is easily applied to the new phantom and showed high reproducibility, (C) 1999 Elsevier Science Inc.
引用
收藏
页码:1511 / 1519
页数:9
相关论文
共 21 条
[1]  
BOCKHORST KHJ, 1997, 5 ANN M INT SOC MAGN, P1127
[2]   QUALITY ASSESSMENT IN IN-VIVO NMR-SPECTROSCOPY .2. A PROTOCOL FOR QUALITY ASSESSMENT [J].
BOVEE, WMMJ ;
KEEVIL, SF ;
LEACH, MO ;
PODO, F .
MAGNETIC RESONANCE IMAGING, 1995, 13 (01) :123-129
[3]   QUALITY ASSESSMENT IN IN-VIVO NMR-SPECTROSCOPY .6. MULTICENTER QUANTIFICATION OF MRS TEST SIGNALS [J].
DEBEER, R ;
BACHERTBAUMANN, P ;
BOVEE, WMMJ ;
CADY, E ;
CHAMBRON, J ;
DOMMISSE, R ;
VANECHTELD, CJA ;
MATHURDEVRE, R ;
WILLIAMS, SR .
MAGNETIC RESONANCE IMAGING, 1995, 13 (01) :169-176
[4]   LOCALIZED PROTON SPECTROSCOPY USING STIMULATED ECHOES [J].
FRAHM, J ;
MERBOLDT, KD ;
HANICKE, W .
JOURNAL OF MAGNETIC RESONANCE, 1987, 72 (03) :502-508
[5]   QUALITY ASSESSMENT IN IN-VIVO NMR-SPECTROSCOPY .5. MULTICENTER EVALUATION OF PROTOTYPE TEST OBJECTS AND PROTOCOLS FOR PERFORMANCE ASSESSMENT IN SMALL-BORE MRS EQUIPMENT [J].
HOWE, FA ;
CANESE, R ;
PODO, F ;
VIKHOFF, B ;
SLOTBOOM, J ;
GRIFFITHS, JR ;
HENRIKSEN, O ;
BOVEE, WMMJ .
MAGNETIC RESONANCE IMAGING, 1995, 13 (01) :159-167
[6]   EXPERIMENTAL CHARACTERIZATION OF THE ISIS TECHNIQUE FOR VOLUME SELECTED NMR-SPECTROSCOPY [J].
KEEVIL, SF ;
PORTER, DA ;
SMITH, MA .
NMR IN BIOMEDICINE, 1992, 5 (04) :200-208
[7]   QUALITY ASSESSMENT IN IN-VIVO NMR-SPECTROSCOPY .4. A MULTICENTER TRIAL OF TEST OBJECTS AND PROTOCOLS FOR PERFORMANCE ASSESSMENT IN CLINICAL NMR-SPECTROSCOPY [J].
KEEVIL, SF ;
BARBIROLI, B ;
COLLINS, DJ ;
DANIELSEN, ER ;
HENNIG, J ;
HENRIKSEN, O ;
LEACH, MO ;
LONGO, R ;
LOWRY, M ;
MOORE, C ;
MOSER, E ;
SEGEBARTH, C ;
BOVEE, WMMJ ;
PODO, F .
MAGNETIC RESONANCE IMAGING, 1995, 13 (01) :139-157
[8]   A METHOD FOR CHARACTERIZING LOCALIZATION TECHNIQUES IN VOLUME SELECTED NUCLEAR-MAGNETIC-RESONANCE SPECTROSCOPY [J].
KEEVIL, SF ;
PORTER, DA ;
SMITH, MA .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (07) :821-834
[9]   COMPUTER-SIMULATION OF MRS LOCALIZATION TECHNIQUES - AN ANALYSIS OF ISIS [J].
LAWRY, TJ ;
KARCZMAR, GS ;
WEINER, MW ;
MATSON, GB .
MAGNETIC RESONANCE IN MEDICINE, 1989, 9 (03) :299-314
[10]   QUALITY ASSESSMENT IN IN-VIVO NMR-SPECTROSCOPY .3. CLINICAL-TEST OBJECTS - DESIGN, CONSTRUCTION, AND SOLUTIONS [J].
LEACH, MO ;
COLLINS, DJ ;
KEEVIL, S ;
ROWLAND, I ;
SMITH, MA ;
HENRIKSEN, O ;
BOVEE, WMMJ ;
PODO, F .
MAGNETIC RESONANCE IMAGING, 1995, 13 (01) :131-137