Background: Despite advances in surgical treatment options, failure rates of rotator cuff repair have continued to range from 20% to 90%. Hence, there is a need for new repair strategies that provide effective mechanical reinforcement of rotator cuff repair as well as stimulate and enhance the intrinsic healing potential of the patient. The purpose of this study was to evaluate the extent to which augmentation of acute repair of rotator cuff tendons with a newly designed poly-L-lactide repair device would improve functional and biomechanical outcomes in a canine model. Methods: Eight adult, male mongrel dogs (25 to 30 kg) underwent bilateral shoulder surgery. One shoulder underwent tendon release and repair only, and the other was subjected to release and repair followed by augmentation with the repair device. At twelve weeks, tendon retraction, cross-sectional area, stiffness, and ultimate load of the repair site were measured. Augmented repairs underwent histologic assessment of biocompatibility. In addition, eight pairs of canine cadaver shoulders underwent infraspinatus injury and repair with and without device augmentation with use of identical surgical procedures and served as time-zero biomechanical controls. Eight unpaired, canine cadaver shoulders were included as normal biomechanical controls. Results: At time zero, repair augmentation significantly increased the ultimate load (23%) (p = 0.034) but not the stiffness of the canine infraspinatus tendon repair. At twelve weeks, the poly-L-lactide scaffold was observed to be histologically biocompatible, and augmented repairs demonstrated significantly less tendon retraction (p = 0.008) and significantly greater cross-sectional area (137%), stiffness (26%), and ultimate load (35%) than did repairs that had not been augmented (p < 0.001, p = 0.002, and p = 0.009, respectively). Conclusions: While limiting but not eliminating tendon repair retraction, the augmentation device provided a tendon-bone bridge and scaffold for host tissue deposition and ingrowth, resulting in improved biomechanical function of the repair at twelve weeks. Clinical Relevance: The augmentation device, applied in a similar manner as described in the present study, might offer a functional benefit to patients undergoing rotator cuff repair.
引用
收藏
页码:1159 / 1171
页数:13
相关论文
共 64 条
[1]  
Accousti Kenneth J, 2007, Instr Course Lect, V56, P3
[2]   Rotator cuff repair using an acellular dermal matrix graft: An in vivo study in a canine model [J].
Adams, Julie E. ;
Zobitz, Mark E. ;
Reach, John S., Jr. ;
an, Kai-N An ;
Steinmann, Scott P. .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2006, 22 (07) :700-709
[3]  
*AM AC ORTH SURG, NUMB PAT NUM PROC AV
[4]   Tensile properties and biological response of poly(L-lactic acid) felt graft: An experimental trial for Rotator-Cuff reconstruction [J].
Aoki, M ;
Miyamoto, S ;
Okamura, K ;
Yamashita, T ;
Ikada, Y ;
Matsuda, S .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2004, 71B (02) :252-259
[5]   Commercially available extracellular matrix materials for rotator cuff repairs: State of the art and future trends [J].
Aurora, Amit ;
McCarron, Jesse ;
Iannotti, Joseph P. ;
Derwin, Kathleen .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2007, 16 (05) :171S-178S
[6]   An assessment of porcine dermal xenograft as an augmentation graft in the treatment of extensive rotator cuff tears [J].
Badhe, Sachin P. ;
Lawrence, Tom M. ;
Smith, F. D. ;
Lunn, P. G. .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2008, 17 (01) :35S-39S
[7]   Homotypic variation of canine flexor tendons: implications for the design of experimental studies in animal models [J].
Baker, AR ;
Abreu, EL ;
Mascha, E ;
Derwin, KA .
JOURNAL OF BIOMECHANICS, 2004, 37 (07) :959-968
[8]   Bilok interference screws for anterior cruciate ligament reconstruction: Clinical and radiographic outcomes [J].
Barber, F. Alan ;
Boothby, Michael H. .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2007, 23 (05) :476-481
[9]   Tendon augmentation grafts: Biomechanical failure loads and failure patterns [J].
Barber, F. Alan ;
Herbert, Morley A. ;
Coons, David A. .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2006, 22 (05) :534-538
[10]  
BARTOLOZZI A, 1994, CLIN ORTHOP RELAT R, P90