Stability and Trunnion Wear Potential in Large-diameter Metal-on-Metal Total Hips: A Finite Element Analysis

被引:54
作者
Elkins, Jacob M. [1 ,2 ]
Callaghan, John J. [1 ,2 ,3 ]
Brown, Thomas D. [1 ,2 ,4 ]
机构
[1] Univ Iowa, Dept Orthopaed & Rehabil, Iowa City, IA 52242 USA
[2] Univ Iowa, Dept Biomed Engn, Iowa City, IA 52242 USA
[3] Iowa City Vet Adm Med Ctr, Iowa City, IA USA
[4] Univ Iowa, Orthopaed Biomech Lab, Iowa City, IA 52242 USA
关键词
LARGE FEMORAL HEADS; ACETABULAR COMPONENT; EARLY FAILURE; CUP ANGLE; IN-VITRO; ARTHROPLASTY; REPLACEMENT; CORROSION; BEARINGS; DISLOCATION;
D O I
10.1007/s11999-013-3244-8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Large-diameter femoral heads for metal-on-metal THA hold theoretical advantages of joint stability and low bearing surface wear. However, recent reports have indicated an unacceptably high rate of wear-associated failure with large-diameter bearings, possibly due in part to increased wear at the trunnion interface. Thus, the deleterious consequences of using large heads may outweigh their theoretical advantages. We investigated (1) to what extent femoral head size influenced stability in THA for several dislocation-prone motions; and the biomechanics of wear at the trunnion interface by considering the relationship between (2) wear potential and head size and (3) wear potential and other factors, including cup orientation, type of hip motion, and assembly/impaction load. Computational simulations were executed using a previously validated nonlinear contact finite element model. Stability was determined at 36 cup orientations for five distinct dislocation challenges. Wear at the trunnion interface was calculated for three separate cup orientations subjected to gait, stooping, and sit-to-stand motions. Seven head diameters were investigated: 32 to 56 mm, in 4-mm increments. Stability improved with increased diameter, although diminishing benefit was seen for sizes of greater than 40 mm. By contrast, contact stress and computed wear at the trunnion interface all increased unabatedly with increasing head size. Increased impaction forces resulted in only small decreases in trunnion wear generation. These data suggest that the theoretical advantages of large-diameter femoral heads have a limit. Diameters of greater than 40 mm demonstrated only modest improvement in terms of joint stability yet incurred substantial increase in wear potential at the trunnion. Our model has potential to help investigators and designers of hip implants to better understand the optimization of trunnion design for long-term durability.
引用
收藏
页码:529 / 542
页数:14
相关论文
共 58 条
  • [31] Metal-on-Metal Hip Arthroplasty in Patients Thirty Years of Age or Younger
    Girard, Julien
    Bocquet, Donatien
    Autissier, Guillaume
    Fouilleron, Nicolas
    Fron, Damien
    Migaud, Henri
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (14) : 2419 - 2426
  • [32] In vitro corrosion testing of modular hip tapers
    Goldberg, JR
    Gilbert, JL
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2003, 64B (02) : 78 - 93
  • [33] Goldberg JR, 2002, CLIN ORTHOP RELAT R, P149
  • [34] Impact magnitudes applied by surgeons and their importance when applying the femoral head onto the Morse taper for total hip arthroplasty
    Heiney, Jake P.
    Battula, Suneel
    Vrabec, Gregory A.
    Parikh, Anand
    Blice, Rebecca
    Schoenfeld, Andrew J.
    Njus, Glenn O.
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2009, 129 (06) : 793 - 796
  • [35] INSALL JN, 1988, CLIN ORTHOP RELAT R, P43
  • [36] Corrosion of metal orthopaedic implants
    Jacobs, JJ
    Gilbert, JL
    Urban, RM
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (02) : 268 - 282
  • [37] The exercise-related rise in plasma cobalt levels after metal-on-metal hip resurfacing arthroplasty
    Khan, M.
    Kuiper, J. -H.
    Richardson, J. B.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (09): : 1152 - 1157
  • [38] Corrosion of a Hip Stem With a Modular Neck Taper Junction A Retrieval Study of 16 Cases
    Kop, Alan M.
    Swarts, Eric
    [J]. JOURNAL OF ARTHROPLASTY, 2009, 24 (07) : 1019 - 1023
  • [39] Accelerating failure rate of the ASR total hip replacement
    Langton, D. J.
    Jameson, S. S.
    Joyce, T. J.
    Gandhi, J. N.
    Sidaginamale, R.
    Mereddy, P.
    Lord, J.
    Nargol, A. V. F.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (08): : 1011 - 1016
  • [40] Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement A CONSEQUENCE OF EXCESS WEAR
    Langton, D. J.
    Jameson, S. S.
    Joyce, T. J.
    Hallab, N. J.
    Natu, S.
    Nargol, A. V. F.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (01): : 38 - 46