Stability and Trunnion Wear Potential in Large-diameter Metal-on-Metal Total Hips: A Finite Element Analysis

被引:54
作者
Elkins, Jacob M. [1 ,2 ]
Callaghan, John J. [1 ,2 ,3 ]
Brown, Thomas D. [1 ,2 ,4 ]
机构
[1] Univ Iowa, Dept Orthopaed & Rehabil, Iowa City, IA 52242 USA
[2] Univ Iowa, Dept Biomed Engn, Iowa City, IA 52242 USA
[3] Iowa City Vet Adm Med Ctr, Iowa City, IA USA
[4] Univ Iowa, Orthopaed Biomech Lab, Iowa City, IA 52242 USA
关键词
LARGE FEMORAL HEADS; ACETABULAR COMPONENT; EARLY FAILURE; CUP ANGLE; IN-VITRO; ARTHROPLASTY; REPLACEMENT; CORROSION; BEARINGS; DISLOCATION;
D O I
10.1007/s11999-013-3244-8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Large-diameter femoral heads for metal-on-metal THA hold theoretical advantages of joint stability and low bearing surface wear. However, recent reports have indicated an unacceptably high rate of wear-associated failure with large-diameter bearings, possibly due in part to increased wear at the trunnion interface. Thus, the deleterious consequences of using large heads may outweigh their theoretical advantages. We investigated (1) to what extent femoral head size influenced stability in THA for several dislocation-prone motions; and the biomechanics of wear at the trunnion interface by considering the relationship between (2) wear potential and head size and (3) wear potential and other factors, including cup orientation, type of hip motion, and assembly/impaction load. Computational simulations were executed using a previously validated nonlinear contact finite element model. Stability was determined at 36 cup orientations for five distinct dislocation challenges. Wear at the trunnion interface was calculated for three separate cup orientations subjected to gait, stooping, and sit-to-stand motions. Seven head diameters were investigated: 32 to 56 mm, in 4-mm increments. Stability improved with increased diameter, although diminishing benefit was seen for sizes of greater than 40 mm. By contrast, contact stress and computed wear at the trunnion interface all increased unabatedly with increasing head size. Increased impaction forces resulted in only small decreases in trunnion wear generation. These data suggest that the theoretical advantages of large-diameter femoral heads have a limit. Diameters of greater than 40 mm demonstrated only modest improvement in terms of joint stability yet incurred substantial increase in wear potential at the trunnion. Our model has potential to help investigators and designers of hip implants to better understand the optimization of trunnion design for long-term durability.
引用
收藏
页码:529 / 542
页数:14
相关论文
共 58 条
  • [1] [Anonymous], J ARTHROPLASTY
  • [2] CONTACT AND RUBBING OF FLAT SURFACES
    ARCHARD, JF
    [J]. JOURNAL OF APPLIED PHYSICS, 1953, 24 (08) : 981 - 988
  • [3] Australian Orthopaedic Association, ANN REP AUSTR ORTH A
  • [4] The effect of femoral component head size on posterior dislocation of the artificial hip joint
    Bartz, RL
    Noble, PC
    Kadakia, NR
    Tullos, HS
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (09) : 1300 - 1307
  • [5] The Durom Large Diameter Head acetabular component EARLY RESULTS WITH A LARGE-DIAMETER METAL-ON-METAL BEARING
    Berton, C.
    Girard, J.
    Krantz, N.
    Migaud, H.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (02): : 202 - 208
  • [6] High failure rates with a large-diameter hybrid metal-on-metal total hip replacement CLINICAL, RADIOLOGICAL AND RETRIEVAL ANALYSIS
    Bolland, B. J. R. F.
    Culliford, D. J.
    Langton, D. J.
    Millington, J. P. S.
    Arden, N. K.
    Latham, J. M.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05): : 608 - 615
  • [7] The Epidemiology of Revision Total Hip Arthroplasty in the United States
    Bozic, Kevin J.
    Kurtz, Steven M.
    Lau, Edmund
    Ong, Kevin
    Vail, Thomas P.
    Berry, Daniel J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (01) : 128 - 133
  • [8] Friction of total hip replacements with different bearings and loading conditions
    Brockett, Claire
    Williams, Sophie
    Jin, Zhongmin
    Isaac, Graham
    Fisher, John
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2007, 81B (02) : 508 - 515
  • [9] FRETTING CORROSION ACCELERATES CREVICE CORROSION OF MODULAR HIP TAPERS
    BROWN, SA
    FLEMMING, CAC
    KAWALEC, JS
    PLACKO, HE
    VASSAUX, C
    MERRITT, K
    PAYER, JH
    KRAAY, MJ
    [J]. JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) : 19 - 26
  • [10] Failed Metal-on-Metal Hip Arthroplasties: A Spectrum of Clinical Presentations and Operative Findings
    Browne, James A.
    Bechtold, C. Dustin
    Berry, Daniel J.
    Hanssen, Arlen D.
    Lewallen, David G.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (09) : 2313 - 2320