Successful Retrieval of Entrapped Gaia Guidewire in Calcified Chronic Total Occlusion Using Rotational Atherectomy Device

被引:7
作者
Cho, Jae Young [1 ]
Hong, Soon Jun [1 ]
机构
[1] Korea Univ, Cardiovasc Ctr, Dept Cardiol, Anam Hosp, Seoul, South Korea
关键词
Guidewire entrapment; Coronary intervention; LUMINAL CORONARY ANGIOPLASTY; BROKEN GUIDEWIRE; WIRE FRAGMENT; MANAGEMENT; ENTRAPMENT; LESION;
D O I
10.1536/ihj.17-188
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report a novel technique for retrieving an entrapped Gaia guidewires in severely calcified coronary chronic total occlusion (CTO) using a rotational atherectomy device. By successfully cutting the entrapped and untangled guidewire within the CTO with the use of a rotational atherectomy device, we could effectively remove the Gaia guidewire from the right coronary artery.