Accuracy of Continuous Glucose Monitoring Measurements in Normo-Glycemic Individuals

被引:43
作者
Akintola, Abimbola A. [1 ]
Noordam, Raymond [1 ]
Jansen, Steffy W. [1 ]
de Craen, Anton J. [1 ]
Ballieux, Bart E. [2 ]
Cobbaert, Christa M. [2 ]
Mooijaart, Simon P. [1 ]
Pijl, Hanno [3 ]
Westendorp, Rudi G. [1 ,4 ]
van Heemst, Diana [1 ]
机构
[1] Leiden Univ, Med Ctr, Dept Gerontol & Geriatr, Leiden, Netherlands
[2] Leiden Univ, Med Ctr, Dept Clin Chem, Leiden, Netherlands
[3] Leiden Univ, Med Ctr, Dept Endocrinol, Leiden, Netherlands
[4] Univ Copenhagen, Dept Publ Hlth, Copenhagen, Denmark
关键词
VARIABILITY; SYSTEM; PERFORMANCE; VALIDATION; SENSOR; RISK;
D O I
10.1371/journal.pone.0139973
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background The validity of continuous glucose monitoring (CGM) is well established in diabetic patients. CGM is also increasingly used for research purposes in normo-glycemic individuals, but the CGM validity in such individuals is unknown. We studied the accuracy of CGM measurements in normo-glycemic individuals by comparing CGM-derived versus venous blood-derived glucose levels and measures of glycemia and glycemic variability. Methods In 34 healthy participants (mean age 65.7 years), glucose was simultaneously measured every 10 minutes, via both an Enlite (R) CGM sensor, and in venous blood sampled over a 24-hour period. Validity of CGM-derived individual glucose measurements, calculated measures of glycemia over daytime (09:00h-23:00h) and nighttime (23:00h-09:00h), and calculated measures of glycemic variability (e.g. 24h standard deviation [SD]) were assessed by Pearson correlation coefficients, mean absolute relative difference (MARD) and paired t-tests. Results The median correlation coefficient between CGM and venous glucose measurements per participant was 0.68 (interquartile range: 0.40-0.78), and the MARD was 17.6% (SD = 17%). Compared with venous sampling, the calculated measure of glycemia during daytime was 0.22 mmol/L higher when derived from CGM, but no difference was observed during nighttime. Most measures of glycemic variability were lower with CGM than with venous blood sampling (e.g., 24h SD:1.07 with CGM and 1.26 with venous blood; p-value = 0.004). Conclusion In normo-glycemic individuals, CGM-derived glucose measurements had good agreement with venous glucose levels. However, the measure of glycemia was higher during the day and most measures of glycemic variability were lower when derived from CGM.
引用
收藏
页数:13
相关论文
共 31 条
[1]   A simple and versatile method for frequent 24 h blood sample collection in healthy older adults [J].
Akintola, A. A. ;
Jansen, S. W. ;
Wilde, R. B. P. ;
Hultzer, G. ;
Rodenburg, R. ;
van Heemst, D. .
METHODSX, 2015, 2 :33-38
[2]   Standards of Medical Care in Diabetes-2014 [J].
不详 .
DIABETES CARE, 2014, 37 :S14-S80
[3]  
[Anonymous], 2014, J DIABETES SCI TECHN
[4]   Accuracy and Acceptability of the 6-Day Enlite Continuous Subcutaneous Glucose Sensor [J].
Bailey, Timothy S. ;
Ahmann, Andrew ;
Brazg, Ronald ;
Christiansen, Mark ;
Garg, Satish ;
Watkins, Elaine ;
Welsh, John B. ;
Lee, Scott W. .
DIABETES TECHNOLOGY & THERAPEUTICS, 2014, 16 (05) :277-283
[5]   Nocturnal Continuous Glucose Monitoring: Accuracy and Reliability of Hypoglycemia Detection in Patients with Type 1 Diabetes at High Risk of Severe Hypoglycemia [J].
Bay, Christiane ;
Kristensen, Peter Lommer ;
Pedersen-Bjergaard, Ulrik ;
Tarnow, Lise ;
Thorsteinsson, Birger .
DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 (05) :371-377
[6]   Validation of the continuous glucose monitoring sensor in preterm infants [J].
Beardsall, K. ;
Vanhaesebrouck, S. ;
Ogilvy-Stuart, A. L. ;
Vanhole, C. ;
vanWeissenbruch, M. ;
Midgley, P. ;
Thio, M. ;
Cornette, L. ;
Ossuetta, I. ;
Palmer, C. R. ;
Iglesias, I. ;
de Jong, M. ;
Gill, B. ;
de Zegher, F. ;
Dunger, D. B. .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2013, 98 (02) :F136-F140
[7]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[8]   A Comparative Effectiveness Analysis of Three Continuous Glucose Monitors [J].
Damiano, Edward R. ;
El-Khatib, Firas H. ;
Zheng, Hui ;
Nathan, David M. ;
Russell, Steven J. .
DIABETES CARE, 2013, 36 (02) :251-259
[9]   Evaluation of a continuous glucose monitoring system in diabetic dogs [J].
Davison, LJ ;
Slater, LA ;
Herrtage, ME ;
Church, DB ;
Judge, S ;
Ristic, JME ;
Catchpole, B .
JOURNAL OF SMALL ANIMAL PRACTICE, 2003, 44 (10) :435-442
[10]   Validation of interstitial fluid continuous glucose monitoring in cystic fibrosis [J].
Dobson, L ;
Sheldon, CD ;
Hattersley, AT .
DIABETES CARE, 2003, 26 (06) :1940-1941