In vitro comparison of the molecular adsorbent recirculation system (MARS) and single-pass albumin dialysis (SPAD)

被引:100
作者
Sauer, IM
Goetz, M
Steffen, I
Walter, G
Kehr, DC
Schwartlander, R
Hwang, YJ
Pascher, A
Gerlach, JC
Neuhaus, P
机构
[1] Freie Univ & Humboldt Univ, Univ Med Berlin, D-13353 Berlin, Germany
[2] Univ Pittsburgh, Med Ctr, Dept Surg, Pittsburgh, PA USA
[3] Kyungpook Natl Univ Hosp, Dept Surg, Taegu, South Korea
关键词
D O I
10.1002/hep.20195
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
The detoxification capacities of single-pass albumin dialysis (SPAD), the molecular adsorbents recirculation system, (MARS) and continuous veno-venous hemodiafiltration (CVVHDF) were compared in vitro. In each experiment 4,100 mL of toxin-loaded human plasma was processed for 6.5 hours. MARS treatment (n = 6) was undertaken in combination with CVVHDF. For SPAD (n = 6) and CVVHDF (n = 6) a high-flux hollow fiber hemodiafilter (identical to the MARS filter) was used. Levels of ammonia, urea, creatinine, bilirubin, and bile acids were determined. Concentrations before and after application of detoxification procedures were expressed as differences and were compared using the Kruskal-Wallis test. Post hoc comparisons for pairs of groups were adjusted according to Bonferroni-Holm. Time, group, and interaction effects were tested using the nonparametric ANOVA model for repeated measurements. SPAD and CVVHDF induced a significantly greater reduction of ammonia levels than MARS. No significant differences were found among SPAD, MARS, and CVVHDF with respect to other water-soluble substances. SPAD induced a significantly greater reduction in bilirubin levels than MARS. Reductions in bile acid levels were similar for SPAD and MARS. When operating MARS in continuous veno-venous hemodialysis mode, as recommended by the manufacturer, no significant differences in the removal of bilirubin, bile acids, urea, and creatinine were found. However, MARS in continuous veno-venous hemodialysis mode was significantly less efficient in removing ammonia than MARS in CVVHDF mode. In conclusion, the detoxification capacity of SPAD is similar to or even greater than that of MARS.
引用
收藏
页码:1408 / 1414
页数:7
相关论文
共 20 条
[11]  
LARSEN FS, 2000, CAN J GASTROENTER SD, V14, P105
[12]   Improvement in central nervous system functions during treatment of liver failure with albumin dialysis MARS - A review of clinical, biochemical, and electrophysiological data [J].
Mitzner, S ;
Loock, J ;
Peszynski, P ;
Klammt, S ;
Majcher-Peszynska, J ;
Gramowski, A ;
Stange, J ;
Schmidt, R .
METABOLIC BRAIN DISEASE, 2002, 17 (04) :463-475
[13]   Improvement of hepatorenal syndrome with extracorporeal albumin dialysis MARS:: Results of a prospective, randomized, controlled clinical trial [J].
Mitzner, SR ;
Stange, J ;
Klammt, S ;
Risler, T ;
Erley, CM ;
Bader, BD ;
Berger, ED ;
Lauchart, W ;
Peszynski, P ;
Freytag, J ;
Hickstein, H ;
Loock, J ;
Löhr, JM ;
Liebe, S ;
Emmrich, J ;
Korten, G ;
Schmidt, R .
LIVER TRANSPLANTATION, 2000, 6 (03) :277-286
[14]  
OGRADY JG, 1991, ACUTE LIVER FAILURE, P77
[15]   Albumin dialysis: single pass vs. recirculation (MARS) [J].
Peszynski, P ;
Peters, E ;
Schmidt, R .
LIVER, 2002, 22 :40-42
[16]   Acute liver failure: targeted artificial and hepatocyte-based support of liver regeneration and reversal of multiorgan failure [J].
Riordan, SM ;
Williams, R .
JOURNAL OF HEPATOLOGY, 2000, 32 :63-76
[17]   Hemodynamic changes during a single treatment with the molecular adsorbents recirculating system in patients with acute-on-chronic liver failure [J].
Schmidt, LE ;
Sorensen, VR ;
Svendsen, LB ;
Hansen, BA ;
Larsen, FS .
LIVER TRANSPLANTATION, 2001, 7 (12) :1034-1039
[18]   Long term treatment of patients with acute exacerbation of chronic liver failure by albumin dialysis [J].
Seige, M ;
Kreymann, B ;
Jeschke, B ;
Schweigart, U ;
Kopp, KF ;
Classen, M .
TRANSPLANTATION PROCEEDINGS, 1999, 31 (1-2) :1371-1375
[19]   Review article: the Molecular Adsorbents Recirculating System (MARS) in liver failure [J].
Sen, S ;
Mookerjee, RP ;
Davies, NA ;
Williams, R ;
Jalan, R .
ALIMENTARY PHARMACOLOGY & THERAPEUTICS, 2002, 16 :32-38
[20]   The molecular adsorbents recycling system as a liver support system based on albumin dialysis: A summary of preclinical investigations, prospective, randomized, controlled clinical trial, and clinical experience from 19 centers [J].
Stange, J ;
Hassanein, TI ;
Mehta, R ;
Mitzner, SR ;
Bartlett, RH .
ARTIFICIAL ORGANS, 2002, 26 (02) :103-110