Chronic recording and electrochemical performance of amorphous silicon carbide-coated Utah electrode arrays implanted in rat motor cortex

被引:26
作者
Joshi-Imre, Alexandra [1 ]
Black, Bryan J. [1 ]
Abbott, Justin [1 ]
Kanneganti, Aswini [1 ]
Rihani, Rashed [1 ]
Chakraborty, Bitan [1 ]
Danda, Vindhya R. [1 ]
Maeng, Jimin [1 ]
Sharma, Rohit [2 ]
Rieth, Loren [2 ]
Negi, Sandeep [3 ]
Pancrazio, Joseph J. [1 ]
Cogan, Stuart F. [1 ]
机构
[1] Univ Texas Dallas, 800 W Campbell Rd, Richardson, TX 75080 USA
[2] Univ Utah, 201 Presidents Cir, Salt Lake City, UT 84112 USA
[3] Blackrock Microsyst LLC, 630 Komas Dr,Suite 200, Salt Lake City, UT 84108 USA
关键词
single unit recordings; active electrode yield; electrochemical impedance spectroscopy; cyclic voltammetry; Utah electrode array; amorphous silicon carbide; sputtered iridium oxide; STIMULATION; CAPABILITIES; TETRAPLEGIA;
D O I
10.1088/1741-2552/ab1bc8
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Clinical applications of implantable microelectrode arrays are currently limited by device failure due to, in part, mechanical and electrochemical failure modes. To overcome this challenge, there is significant research interest in the exploration of novel array architectures and encapsulation materials. Amorphous silicon carbide (a-SiC) is biocompatible and corrosion resistant, and has recently been employed as a coating on biomedical devices including planar microelectrode arrays. However, to date, the three-dimensional Utah electrode array (UEA) is the only array architecture which has been approved by the food and drug administration (FDA) for long-term human trials. Approach. Here, we demonstrate, for the first time, that UEAs can be fabricated with a-SiC encapsulation and sputtered iridium oxide film (SIROF) electrode coatings, and that such arrays are capable of single-unit recordings over a 30 week implantation period in rat motor cortex. Over the same period, we carried out electrochemical measurements, including voltage transients, cyclic voltammetry, and electrochemical impedance spectroscopy (EIS), to evaluate potential failure modes. Furthermore, we evaluated chronic foreign body response via fluorescence immunohistochemistry following device explantation. Main results. During the indwelling period, we observed a reduction in active electrode yield percentage from 94.6 +/- 5.4 (week 1) to 16.4 +/- 11.5% (week 30). While the average active electrode yield showed a steady reduction, it is noteworthy that 3 out of 8 UEAs recorded greater than 60% active electrode yield at all times through 24 weeks and 1 out of 8 UEAs recorded greater than 60% active electrode yield at all times through the whole implantation period. Significance. In total, these findings further suggest that a-SiC may serve as a mechanically and electrochemically stable device encapsulation alternative to polymeric coatings such as Parylene-C.
引用
收藏
页数:11
相关论文
共 39 条
  • [1] Restoration of reaching and grasping movements through brain-controlled muscle stimulation in a person with tetraplegia: a proof-of-concept demonstration
    Ajiboye, A. Bolu
    Willett, Francis R.
    Young, Daniel R.
    Memberg, William D.
    Murphy, Brian A.
    Miller, Jonathan P.
    Walter, Benjamin L.
    Sweet, Jennifer A.
    Hoyen, Harry A.
    Keith, Michael W.
    Peckham, P. Hunter
    Simeral, John D.
    Donoghue, John P.
    Hochberg, Leigh R.
    Kirsch, Robert F.
    [J]. LANCET, 2017, 389 (10081) : 1821 - 1830
  • [2] Improvement of stenting therapy with a silicon carbide coated tantalum stent
    Amon, M
    Bolz, A
    Schaldach, M
    [J]. JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1996, 7 (05) : 273 - 278
  • [3] [Anonymous], 2002, MED SCI MONITOR
  • [4] Scanning electron microscopy of chronically implanted intracortical microelectrode arrays in non-human primates
    Barrese, James C.
    Aceros, Juan
    Donoghue, John P.
    [J]. JOURNAL OF NEURAL ENGINEERING, 2016, 13 (02)
  • [5] A wafer-scale etching technique for high aspect ratio implantable MEMS structures
    Bhandari, R.
    Negi, S.
    Rieth, L.
    Solzbacher, F.
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 2010, 162 (01) : 130 - 136
  • [6] Chronic recording and electrochemical performance of Utah microelectrode arrays implanted in rat motor cortex
    Black, Bryan J.
    Kanneganti, Aswini
    Joshi-Imre, Alexandra
    Rihani, Rashed
    Chakraborty, Bitan
    Abbott, Justin
    Pancrazio, Joseph J.
    Cogan, Stuart F.
    [J]. JOURNAL OF NEUROPHYSIOLOGY, 2018, 120 (04) : 2083 - 2090
  • [7] Restoring cortical control of functional movement in a human with quadriplegia
    Bouton, Chad E.
    Shaikhouni, Ammar
    Annetta, Nicholas V.
    Bockbrader, Marcia A.
    Friedenberg, David A.
    Nielson, Dylan M.
    Sharma, Gaurav
    Sederberg, Per B.
    Glenn, Bradley C.
    Mysiw, W. Jerry
    Morgan, Austin G.
    Deogaonkar, Milind
    Rezai, Ali R.
    [J]. NATURE, 2016, 533 (7602) : 247 - +
  • [8] Review: Human Intracortical Recording and Neural Decoding for Brain-Computer Interfaces
    Brandman, David M.
    Cash, Sydney S.
    Hochberg, Leigh R.
    [J]. IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2017, 25 (10) : 1687 - 1696
  • [9] Unique electrophysiological and impedance signatures between encapsulation types: An analysis of biological Utah array failure and benefit of a biomimetic coating in a rat model
    Cody, Patrick A.
    Eles, James R.
    Lagenaur, Carl F.
    Kozai, Takashi D. Y.
    Cui, X. Tracy
    [J]. BIOMATERIALS, 2018, 161 : 117 - 128
  • [10] Neural stimulation and recording electrodes
    Cogan, Stuart F.
    [J]. ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2008, 10 : 275 - 309