Air versus bone conduction:: an equal loudness investigation

被引:38
作者
Stenfelt, S [1 ]
Håkansson, B [1 ]
机构
[1] Chalmers Univ Technol, Dept Signals & Syst, SE-41296 Gothenburg, Sweden
关键词
bone conduction; loudness estimation; acoustic reflex;
D O I
10.1016/S0378-5955(01)00407-5
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Air conduction (AC) versus bone conduction (BC) loudness balance testing was conducted at frequencies of 0.25, 0.5, 0.75, 1, 2, and 4 kHz for two groups: 23 normal hearing subjects and eight subjects with a mild to moderate pure sensorineural hearing loss. Narrow-band noise was presented interchangeably between earphones and a bone transducer fitted to the subjects. Loudness matching was carried out at each frequency and at the levels 30-80 dB hearing level (HL) (10 dB steps) in the following manner: the sound pressure from the earphones was fixed and the subject adjusted the output level of the bone transducer for equal loudness by bracketing the standard. The results revealed somewhat different loudness functions for AC and BC sound with a 6-10 dB difference in the AC and BC loudness functions for the normal hearing group over the dynamic range 30-80 dB HL at the frequencies 250-750 Hz. At the higher frequencies, 1-4 kHz, the difference was only 4-5 dB over the same dynamic range. Similar results were obtained for the sensorineural hearing-impaired group. The difference between the AC and the BC loudness functions may originate from changes with level of the AC sound path, e.g. contraction of the stapedius muscle, but also distortion from the bone transducer and tactile stimulation could have contributed to the results seen. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:1 / 12
页数:12
相关论文
共 48 条
[21]  
MAULDIN L, 1979, ARCH OTOLARYNGOL, V105, P656
[22]  
Merchant SN, 1996, HEARING RES, V97, P30, DOI 10.1016/0378-5955(96)00023-8
[23]  
NEUMAN J, 1996, AUDIOL NEURO-OTOL, V1, P354
[24]   TRANSCRANIAL ATTENUATION IN BONE CONDUCTION AUDIOMETRY [J].
NOLAN, M ;
LYON, DJ .
JOURNAL OF LARYNGOLOGY AND OTOLOGY, 1981, 95 (06) :597-608
[25]  
NORTHERN JL, 1994, HDB CLIN AUDIOLOGY, P300
[26]   Objective calibration of bone conductors using otoacoustic emissions [J].
Purcell, D ;
Kunov, H ;
Cleghorn, W .
EAR AND HEARING, 1999, 20 (05) :375-392
[27]   Distortion product otoacoustic emissions stimulated through bone conduction [J].
Purcell, D ;
Kunov, H ;
Madsen, P ;
Cleghorn, W .
EAR AND HEARING, 1998, 19 (05) :362-370
[28]  
Ringertz N., 1938, ACTA OTO-LARYNGOL, V26, P386
[29]  
ROSSI G, 1988, SCAND AUDIOL, P1
[30]   DICHOTIC SUMMATION OF LOUDNESS [J].
SCHARF, B .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1969, 45 (05) :1193-&