Optical Response of Expired EBT3 Film for Absorbed Dose Measurement in X-ray and Electron Beam Range

被引:0
作者
Musa, Maisarah [1 ]
Abu Mhanna, Hamad Yahia [1 ]
Omar, Ahmad Fairuz [1 ]
Radzi, Yasmin Md [1 ]
Akhdar, Hanan Fawaz [2 ]
Ishak, Nor Hafizah [3 ]
Al Ewaidat, Haytham [4 ]
机构
[1] Univ Sains Malaysia, Sch Phys, George Town 11800, Malaysia
[2] Imam Mohammad Ibn Saud Islamic Univ IMSIU, Fac Sci, Dept Phys, Riyadh 13318, Saudi Arabia
[3] Univ Sains Malaysia USM, Adv Med & Dent Inst AMDI, George Town 11800, Malaysia
[4] Jordan Univ Sci & Technol, Fac Appl Med Sci, Dept Allied Med Sci Radiol Technol, Irbid 22110, Jordan
来源
APPLIED SCIENCES-BASEL | 2022年 / 12卷 / 08期
关键词
colorimetry; expired EBT3; expiry date; visible spectroscopy; RADIOCHROMIC FILM; DOSIMETRY; RADIATION; PROTOCOL; DEPENDENCE; PHOTON;
D O I
10.3390/app12083974
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The purpose of this study was to investigate the optical response of an expired External Beam Therapy (EBT3) film, which expired in 2018, using X-rays and electron beam doses. The film's optical responses were evaluated for its usability in measuring different radiation sources, energy, and absorbed doses ranging up to 5 Gy. Pieces of the expired EBT3 film were irradiated with 90 kVp, 6 MV X-ray photons, and 6 MeV electron beam. The analysis was performed using the Jaz visible spectrometer and EPSON Perfection V370 Photo scanner to obtain the absorbance and the net relative optical density (ROD) of the film samples respectively. The results showed that spectroscopic measurements of the exposed expired EBT3 films under these radiation sources were able to produce primary secondary peaks at lambda = 633.52 nm and lambda = 582.3 nm respectively. The best wavelength subsets that presented the best MLR regression fitting for all experiments were 541.48 nm, 561.11 nm, and 600.28 nm. While, for the 6 MV photon and the 6 MeV electron beam they were 600.28 nm, 650.79 nm and 654.10 nm. In case of the irradiation with the 6 MV photon and the 6 MeV electron beam, expired EBT3 film showed no significant differences, which made it suitable for dosimetry in various sources of radiation. The individual calibration of radiation dose produces very high measurement accuracy with coefficient of determination, R-2 above 0.99 and root mean square of error, RMSE of 0.038 Gy, 0.113 Gy, and 0.115 Gy for films irradiated with 90 kVp X-rays, 6 MV photon beam, and 6 MeV electron beam respectively. Hence, from the results, the expired EBT3 film used in this study showed promising usability of expired EBT3 films beyond their prescribed expiry dates.
引用
收藏
页数:12
相关论文
共 26 条
[1]   Spectroscopy and Light Emitting Diodes Based System in Characterizing External Beam Therapy 3 Films for Solar Ultraviolet Measurement [J].
Ahmad Shah, Wan Insaniah Saleha ;
Omar, Ahmad Fairuz .
PHOTONIC SENSORS, 2020, 10 (01) :34-44
[2]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[3]   Characterization of Gafchromic EBT3 films for ultraviolet radiation dosimetry [J].
Aydarous, Abdulkadir ;
Al-Omary, Enaam Ali ;
El Ghazaly, Mahmoud .
RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2014, 169 (03) :249-255
[4]   Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification [J].
Borca, Valeria Casanova ;
Pasquino, Massimo ;
Russo, Giuliana ;
Grosso, Pierangelo ;
Cante, Domenico ;
Sciacero, Piera ;
Girelli, Giuseppe ;
La Porta, Maria Rosa ;
Tofani, Santi .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02) :158-171
[5]   Measuring solar UV radiation with EBT radiochromic film [J].
Butson, Ethan T. ;
Cheung, Tsang ;
Yu, Peter K. N. ;
Butson, Martin J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (20) :N487-N493
[6]   Effects of read-out light sources and ambient light on radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Metcalfe, PE .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (08) :2407-2412
[7]   Modeling the dose dependence of the vis-absorption spectrum of EBT3 GafChromic™ films [J].
Callens, Maarten B. ;
Crijns, Wouter ;
Depuydt, Tom ;
Haustermans, Karin ;
Maes, Frederik ;
D'Agostino, Emiliano ;
Wevers, Martine ;
Pfeiffer, Helge ;
Van den Abeele, Koen .
MEDICAL PHYSICS, 2017, 44 (06) :2532-2543
[8]   Auto-development issue in quality assurance of biological X-ray irradiator using Gafchromic EBT3 film [J].
Chan, P. M. ;
Ng, C. Y. P. ;
Beni, Mehrdad Shahmohammadi ;
Law, S. S. ;
Yu, K. N. .
RESULTS IN PHYSICS, 2017, 7 :444-445
[9]   Energy Dependence of the New GafChromic-EBT3 Film's Dose Response-Curve [J].
Chiu-Tsao, S. ;
Massillon-JL, G. ;
Domingo-Munoz, I. Domingo-MuA ;
Chan, M. .
MEDICAL PHYSICS, 2012, 39 (06) :3724-3724
[10]   Spectroscopic analysis of irradiated radiochromic EBT-XD films in proton and photon beams [J].
Darafsheh, Arash ;
Zhao, Tianyu ;
Khan, Rao .
PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (20)