Quantitative Assessment of Limb Position Sense Following Stroke

被引:275
作者
Dukelow, Sean P. [2 ]
Herter, Troy M.
Moore, Kimberly D.
Demers, Mary Jo [3 ]
Glasgow, Janice I.
Bagg, Stephen D. [3 ]
Norman, Kathleen E.
Scott, Stephen H. [1 ]
机构
[1] Queens Univ, Ctr Neurosci Studies, Kingston, ON K7L 3N6, Canada
[2] Univ Calgary, Calgary, AB, Canada
[3] St Marys Lake Hosp, Kingston, ON, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
position sense; stroke; rehabilitation; robotics; proprioception; SENSORIMOTOR SYSTEM; PROPRIOCEPTION; HEMINEGLECT; HANDEDNESS;
D O I
10.1177/1545968309345267
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background. Impairment of position sense of the upper extremity (UE) may impede activities of daily living and limit motor gains after stroke. Most clinical assessments of position sense rely on categorical or ordinal ratings by clinicians that lack sensitivity to change or the ability to discriminate subtle deficits. Objective. Use robotic technology to develop a reliable, quantitative technique with a continuous scale to assess UE position sense following stroke. Methods. Forty-five patients recruited from an inpatient stroke rehabilitation service and 65 age-matched healthy controls performed an arm position matching task. Each UE was fitted in the exoskeleton of a KINARM device. One UE was passively placed in one of 9 positions, and the subject was told to match his or her position with the other UE. Patients were compared with statistical distributions of control data to identify those with deficits in UE position sense. Test-retest sessions using 2 raters established interrater reliability. Results. Two thirds of left hemiparetic and one third of right hemiparetic patients had deficits in limb position sense. Left-affected stroke subjects demonstrated significantly more trial-to-trial variability than right-affected or control subjects. The robotic assessment technique demonstrated good interrater reliability but limited agreement with the clinical thumb localizing test. Conclusions. Robotic technology can provide a reliable quantitative means to assess deficits in limb position sense following stroke.
引用
收藏
页码:178 / 187
页数:10
相关论文
共 48 条
[41]   Apparatus for measuring and perturbing shoulder and elbow joint positions and torques during reaching [J].
Scott, SH .
JOURNAL OF NEUROSCIENCE METHODS, 1999, 89 (02) :119-127
[42]  
SCOTT SH, 1994, J NEUROSCI, V14, P7529
[43]   THE HEMIPLEGIC ARM AFTER STROKE - MEASUREMENT AND RECOVERY [J].
WADE, DT ;
LANGTONHEWER, R ;
WOOD, VA ;
SKILBECK, CE ;
ISMAIL, HM .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1983, 46 (06) :521-524
[44]  
WADELL I, 1987, International Journal of Rehabilitation Research, V10, P103, DOI 10.1097/00004356-198712005-00021
[45]  
WILSON B, 1987, ARCH PHYS MED REHAB, V68, P98
[46]   A novel device for the measurement of proprioception in the hand [J].
Wycherley, AS ;
Helliwell, PS ;
Bird, HA .
RHEUMATOLOGY, 2005, 44 (05) :638-641
[47]   Neuromagnetic investigation of somatosensory cortical reorganization in hemiplegic patients after thalamic hemorrhage [J].
Yoshida, Hideki ;
Kondo, Takeo ;
Nakasato, Nobukazu .
JOURNAL OF PHYSICAL THERAPY SCIENCE, 2008, 20 (02) :123-127
[48]  
2000, BRAIN J