Point-of-care hemolysis detection in blood gas specimens directly at the emergency department

被引:27
作者
Duhalde, Henrik [1 ]
Skogo, Johan [2 ]
Karlsson, Mathias [3 ]
机构
[1] Hemcheck, Univ Gatan 2, S-65188 Karlstad, Sweden
[2] Vastmanland Hosp, Dept Clin Chem, Vasteras, Sweden
[3] Cent Hosp Karlstad, Dept Clin Chem, Karlstad, Sweden
关键词
Blood gas analysis; hemolysis; emergency service; hospital; pre-analytical phase; point-of-care systems; quality improvement; ICTERUS; LIPEMIA;
D O I
10.1080/00365513.2019.1612089
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Blood gas analysis represents an important and frequently used clinical decision support. Spurious haemolysis is common at the emergency department (ED) and causes clinically relevant bias to many analyses. Blood gas analysers inability to identify hemolysis can therefore lead to clinical misinterpretations. The objective of this study was to assess the proportion of haemolytic blood gas samples performed at the ED and calculate the predictive values of a novel point-of-care solution for haemolysis detection. Approximately 100 mu l of blood from anonymized syringes including venous or arterial blood was, directly after performed blood gas analysis, tested with the POC-Method (Helge H10 s-system, Hemcheck, Karlstad, Sweden) at the ED and result recorded. Thereafter the blood was centrifuged (2000 g) for 10 minutes, 100 mu l of plasma put into vials and run in batches of 40 daily on an AU680 analyzer (Beckman Coulter, Brea, California, USA) used as reference method. The threshold for clinically relevant hemolysis was defined as >= 50 mg/dL free haemoglobin. Out of 1270 samples, haemolysis was detected in 7.9% (n = 100). The POC-method identified haemolytic samples with a sensitivity of 80%, specificity of 99%, positive and negative predictive values was 89% and 98% respectively. Spurious hemolysis is frequently present in blood gas samples at the ED and should be taken into consideration when assessed in patients. Errors could be detected directly at the point-of-care with a novel technology for hemolysis detection.
引用
收藏
页码:283 / 287
页数:5
相关论文
共 17 条
[1]   Hemolysis in serum samples drawn by emergency department personnel versus laboratory phlebotomists [J].
Burns, ER ;
Yoshikawa, N .
LABORATORY MEDICINE, 2002, 33 (05) :378-380
[2]   Point-of-care testing in critical care: the clinician's point of view [J].
Casagranda, Ivo .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (07) :931-934
[3]  
CLSI, 2009, CLSI DOCUMENT C46 A2
[4]   Impact of Endogenous and Exogenous Interferences on Clinical Chemistry Parameters Measured on Blood Gas Analyzers [J].
Grieme, Caleb V. ;
Voss, Dena R. ;
Davis, Scott R. ;
Krasowski, Matthew D. .
CLINICAL LABORATORY, 2017, 63 (03) :561-568
[5]   Measurement of whole-blood potassium - Is it clinically safe? [J].
Hawkins, R .
CLINICAL CHEMISTRY, 2003, 49 (12) :2105-2106
[6]   Effectiveness of practices to reduce blood sample hemolysis in EDs: A laboratory medicine best practices systematic review and meta-analysis [J].
Heyer, Nicholas J. ;
Derzon, James H. ;
Winges, Linda ;
Shaw, Colleen ;
Mass, Diana ;
Snyder, Susan R. ;
Epner, Paul ;
Nichols, James H. ;
Gayken, Julie A. ;
Ernst, Dennis ;
Liebow, Edward B. .
CLINICAL BIOCHEMISTRY, 2012, 45 (13-14) :1012-1032
[7]   Practical recommendations for managing hemolyzed samples in clinical chemistry testing [J].
Lippi, Giuseppe ;
Cadamuro, Janne ;
von Meyer, Alexander ;
Simundic, Ana-Maria .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (05) :718-727
[8]   Prevalence of hemolytic specimens referred for arterial blood gas analysis [J].
Lippi, Giuseppe ;
Ippolito, Luigi ;
Fontana, Rossana .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 (05) :931-932
[9]   Influence of spurious hemolysis on blood gas analysis [J].
Lippi, Giuseppe ;
Fontana, Rossana ;
Avanzini, Paola ;
Sandei, Franca ;
Ippolito, Luigi .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (08) :1651-1654
[10]   Hemolyzed specimens: a major challenge for emergency departments and clinical laboratories [J].
Lippi, Giuseppe ;
Plebani, Mario ;
Di Somma, Salvatore ;
Cervellin, Gianfranco .
CRITICAL REVIEWS IN CLINICAL LABORATORY SCIENCES, 2011, 48 (03) :143-153