Transit and non-transit 3D EPID dosimetry versus detector arrays for patient specific QA

被引:33
作者
Olaciregui-Ruiz, Igor [1 ]
Vivas-Maiques, Begona [1 ]
Kaas, Jochem [1 ]
Perik, Thijs [1 ]
Wittkamper, Frits [1 ]
Mijnheer, Ben [1 ]
Mans, Anton [1 ]
机构
[1] Netherlands Canc Inst Antoni van Leeuwenhoek, Dept Radiat Oncol, Amsterdam, Netherlands
关键词
detector arrays; EPID dosimetry; IMRT; patient specific QA; VMAT; POLYMER GEL DOSIMETRY; IN-VIVO DOSIMETRY; DOSE RECONSTRUCTION; CLINICAL-EXPERIENCE; DIODE-ARRAY; IMRT; VERIFICATION; VALIDATION; ALGORITHM; SOFTWARE;
D O I
10.1002/acm2.12610
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Despite their availability and simplicity of use, Electronic Portal Imaging Devices (EPIDs) have not yet replaced detector arrays for patient specific QA in 3D. The purpose of this study is to perform a large scale dosimetric evaluation of transit and non-transit EPID dosimetry against absolute dose measurements in 3D. Methods After evaluating basic dosimetric characteristics of the EPID and two detector arrays (Octavius 1500 and Octavius 1000(SRS)), 3D dose distributions for 68 VMAT arcs, and 10 IMRT plans were reconstructed within the same phantom geometry using transit EPID dosimetry, non-transit EPID dosimetry, and the Octavius 4D system. The reconstructed 3D dose distributions were directly compared by gamma-analysis (2L2 = 2% local/2 mm and 3G2 = 3% global/2 mm, 50% isodose) and by the percentage difference in median dose to the high dose volume (% increment HDVD50). Results Regarding dose rate dependency, dose linearity, and field size dependence, the agreement between EPID dosimetry and the two detector arrays was found to be within 1.0%. In the 2L2 gamma-comparison with Octavius 4D dose distributions, the average gamma-pass rate value was 92.2 +/- 5.2%(1SD) and 94.1 +/- 4.3%(1SD) for transit and non-transit EPID dosimetry, respectively. 3G2 gamma-pass rate values were higher than 95% in 150/156 cases. % increment HDVD50 values were within 2% in 134/156 cases and within 3% in 155/156 cases. With regard to the clinical classification of alerts, 97.5% of the treatments were equally classified by EPID dosimetry and Octavius 4D. Conclusion Transit and non-transit EPID dosimetry are equivalent in dosimetric terms to conventional detector arrays for patient specific QA. Non-transit 3D EPID dosimetry can be readily used for pre-treatment patient specific QA of IMRT and VMAT, eliminating the need of phantom positioning.
引用
收藏
页码:79 / 90
页数:12
相关论文
共 39 条
[31]   The Octavius1500 2D ion chamber array and its associated phantoms: Dosimetric characterization of a new prototype [J].
Van Esch, Ann ;
Basta, Katarzyna ;
Evrard, Marie ;
Ghislain, Michel ;
Sergent, Francois ;
Huyskens, Dominique P. .
MEDICAL PHYSICS, 2014, 41 (09)
[32]   A literature review of electronic portal imaging for radiotherapy dosimetry [J].
van Etmpt, Wouter ;
McDermott, Leah ;
Nijsten, Sebastiaan ;
Wendling, Markus ;
Lambin, Philippe ;
Mijnheer, Ben .
RADIOTHERAPY AND ONCOLOGY, 2008, 88 (03) :289-309
[33]   Validation of a method for in vivo 3D dose reconstruction for IMRT and VMAT treatments using on-treatment EPID images and a model-based forward-calculation algorithm [J].
Van Uytven, Eric ;
Van Beek, Timothy ;
McCowan, Peter M. ;
Chytyk-Praznik, Krista ;
Greer, Peter B. ;
McCurdy, Boyd M. C. .
MEDICAL PHYSICS, 2015, 42 (12) :6945-6954
[34]   Evaluation of radiochromic gel dosimetry and polymer gel dosimetry in a clinical dose verification [J].
Vandecasteele, Jan ;
De Deene, Yves .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (18) :6241-6262
[35]  
Varatharaj C, 2011, J APPL CLIN MED PHYS, V12, P338
[36]  
Watanabe Y, 2017, WORLD J RADIOL, V9, P112, DOI 10.4329/wjr.v9.i3.112
[37]   Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method [J].
Wendling, M ;
Louwe, RJW ;
McDermott, LN ;
Sonke, JJ ;
van Herk, M ;
Mijnheer, BJ .
MEDICAL PHYSICS, 2006, 33 (02) :259-273
[38]   In aqua vivo EPID dosimetry [J].
Wendling, Markus ;
McDermott, Leah N. ;
Mans, Anton ;
Olaciregui-Ruiz, Igor ;
Pecharroman-Gallego, Raul ;
Sonke, Jan-Jakob ;
Stroom, Joep ;
van Herk, Marcel ;
Mijnheer, Ben J. .
MEDICAL PHYSICS, 2012, 39 (01) :367-377
[39]   A simple backprojection algorithm for 3D in vivo EPID dosimetry of IMRT treatments [J].
Wendling, Markus ;
McDermott, Leah N. ;
Mans, Anton ;
Sonke, Jan-Jakob ;
van Herk, Marcel ;
Mijnheer, Ben J. .
MEDICAL PHYSICS, 2009, 36 (07) :3310-3321