A Miniaturized Colorimeter with a Novel Design and High Precision for Photometric Detection

被引:12
作者
Yan, Jun-Chao [1 ,2 ,3 ,4 ]
Chen, Yan [1 ,2 ]
Pang, Yu [1 ,2 ]
Slavik, Jan [1 ,2 ,5 ]
Zhao, Yun-Fei [1 ,2 ]
Wu, Xiao-Ming [1 ,2 ]
Yang, Yi [1 ,2 ]
Yang, Si-Fan [1 ,2 ,4 ]
Ren, Tian-Ling [1 ,2 ]
机构
[1] Tsinghua Univ, Inst Microelect, Beijing 100084, Peoples R China
[2] Tsinghua Univ, Tsinghua Natl Lab Informat Sci & Technol TNList, Beijing 100084, Peoples R China
[3] Tsinghua Univ, Grad Sch Shenzhen, Shenzhen 518055, Peoples R China
[4] Tsinghua Univ, Dept Elect Engn, Beijing 100084, Peoples R China
[5] Brno Univ Technol, Cent Inst Technol CEITEC, Purkynova 123, Brno 61200, Czech Republic
关键词
colorimeter; phosphorus detection; Beer-Lambert law; microflow cell; photodetection sensor; PLASMA-MASS SPECTROMETRY; FLOW-INJECTION ANALYSIS; SOLENOID MICRO-PUMPS; LAMBERT-BEER LAW; ANALYSIS SYSTEM; BATCH ANALYZER; TABLE SALT; PHOSPHATE; FLUORESCENCE; QUALITY;
D O I
10.3390/s18030818
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Water quality detection plays an increasingly important role in environmental protection. In this work, a novel colorimeter based on the Beer-Lambert law was designed for chemical element detection in water with high precision and miniaturized structure. As an example, the colorimeter can detect phosphorus, which was accomplished in this article to evaluate the performance. Simultaneously, a modified algorithm was applied to extend the linear measurable range. The colorimeter encompassed a near infrared laser source, a microflow cell based on microfluidic technology and a light-sensitive detector, then Micro-Electro-Mechanical System (MEMS) processing technology was used to form a stable integrated structure. Experiments were performed based on the ammonium molybdate spectrophotometric method, including the preparation of phosphorus standard solution, reducing agent, chromogenic agent and color reaction. The device can obtain a wide linear response range (0.05 mg/L up to 7.60 mg/L), a wide reliable measuring range up to 10.16 mg/L after using a novel algorithm, and a low limit of detection (0.02 mg/L). The size of flow cell in this design is 18 mm x 2.0 mm x 800 mu m, obtaining a low reagent consumption of 0.004 mg ascorbic acid and 0.011 mg ammonium molybdate per determination. Achieving these advantages of miniaturized volume, high precision and low cost, the design can also be used in automated in situ detection.
引用
收藏
页数:14
相关论文
共 56 条
[1]  
Abdallah O., 2012, INT C APPL MATH PHAR, P601
[2]   PDMS Microfluidic Device with an Integrated Optical Sensor for Determination of Zinc (II) in Pharmaceuticals: Toward the Production of Miniaturized Green Chemistry Analytical Systems [J].
AlSuhaimi, Awadh ;
Greenway, Gullian M. ;
McCreedy, Tom ;
Youngvises, Napaporn .
SPECTROSCOPY LETTERS, 2009, 42 (6-7) :370-375
[3]  
[Anonymous], 2006, 57503 GBT
[4]  
[Anonymous], 1989, 1189389 GBT
[5]  
Babin SM, 2008, J HOPKINS APL TECH D, V27, P403
[6]   Downscaling a multicommuted flow injection analysis system for the photometric determination of iodate in table salt [J].
Borges, Sivanildo S. ;
Peixoto, Jailson de Souza ;
Feres, Mario A. ;
Reis, Boaventura F. .
ANALYTICA CHIMICA ACTA, 2010, 668 (01) :3-7
[7]   NEW CONCEPTION OF BOUGUER-LAMBERT-BEER LAW [J].
BOTEV, I .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1979, 297 (05) :419-419
[8]  
Boumans P.W.J.M., 1966, FRESEN Z ANAL CHEM, V220, P241
[9]   Recent Applications of Fourier Transform Infrared Spectrophotometry in Herbal Medicine Analysis [J].
Bunaciu, Andrei A. ;
Aboul-Enein, Hassan Y. ;
Fleschin, Serban .
APPLIED SPECTROSCOPY REVIEWS, 2011, 46 (04) :251-260
[10]   Micro-analyzer with optical detection and wireless communications [J].
Camarillo-Escobedo, R. M. ;
Valdes-Perezgasga, F. ;
Rodriguez-Rivera, R. .
SENSORS AND ACTUATORS A-PHYSICAL, 2013, 199 :181-186