Quality assurance procedures for the neutron beam monitors at the FiR 1 BNCT facility

被引:7
作者
Auterinen, I
Serén, T
Kotiluoto, P
Uusi-Simola, J
Savolainen, S
机构
[1] VTT Proc, VTT Tech Res Ctr Finland, FIN-02044 Espoo, Finland
[2] Univ Helsinki, Dept Phys Sci, FIN-00014 Helsinki, Finland
[3] Univ Helsinki, HUSLAB, Dept Clin Physiol & Nucl Med, FIN-00029 Helsinki, Finland
[4] Univ Helsinki, HUS Helsinki Med Imaging Ctr, FIN-00029 Helsinki, Finland
关键词
BNCT; beam monitoring; quality assurance; non-linearity;
D O I
10.1016/j.apradiso.2004.05.034
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In order to assure the stability of the beam, the reliability of the beam monitoring system and the quality of the patient dose delivered, several procedures are followed at the FiR 1 epithermal beam in Finland. Routine procedures include in-phantom activation measurements before each patient treatment and a long-term follow-up of the results. The sensitivity of the beam monitors to external objects in the beam and to variations in the control rod positions in the reactor has been checked and found insignificant. The linearity of the beam monitor channels has been checked with activation measurements. It was found that due to saturation effects a correction of 11% has to be applied when extrapolating results from experiments at low power to full power using the reference monitor channel. The correction is even larger for other channels with higher count rates. (C) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1015 / 1019
页数:5
相关论文
共 6 条
[1]  
AARNIO P, 1999, SAMPO ADV GAMMA SPEC
[2]   Design and construction of shoulder recesses into the beam aperture shields for improved patient positioning at the FiR 1 BNCT facility [J].
Auterinen, I ;
Kotiluoto, P ;
Hippeläinen, E ;
Kortesniemi, M ;
Seppälä, T ;
Serén, T ;
Mannila, V ;
Pöyry, P ;
Kankaanranta, L ;
Collan, J ;
Kouri, M ;
Joensuu, H ;
Savolainen, S .
APPLIED RADIATION AND ISOTOPES, 2004, 61 (05) :799-803
[3]  
Auterinen I, 2001, FRONTIERS IN NEUTRON CAPTURE THERAPY, VOLS 1 AND 2, P267
[4]   Boron neutron capture therapy of brain tumors:: clinical trials at the Finnish facility using boronophenylalanine [J].
Joensuu, H ;
Kankaanranta, L ;
Seppälä, T ;
Auterinen, I ;
Kallio, M ;
Kulvik, M ;
Laakso, J ;
Vähätalo, J ;
Kortesniemi, M ;
Kotiluoto, P ;
Serén, T ;
Karila, J ;
Brander, A ;
Järviluoma, E ;
Ryynänen, P ;
Paetau, A ;
Ruokonen, I ;
Minn, H ;
Tenhunen, M ;
Jääkeläinen, J ;
Färkkilä, M ;
Savolainen, S .
JOURNAL OF NEURO-ONCOLOGY, 2003, 62 (01) :123-134
[5]   On-line neutron beam monitoring of the Finnish BNCT facility [J].
Tanner, V ;
Auterinen, I ;
Helin, J ;
Kosunen, A ;
Savolainen, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 422 (1-3) :101-105
[6]  
Voorbraak W.P., 2003, RECOMMENDATIONS DOSI