A Novel Ion Trap That Enables High Duty Cycle and Wide m/z Range on an Orthogonal Injection TOF Mass Spectrometer

被引:43
作者
Loboda, Alexander V. [1 ]
Chernushevich, Igor V. [1 ]
机构
[1] MDS Analyt Technol, Concord, ON L4K 4V8, Canada
关键词
IMPROVEMENT;
D O I
10.1016/j.jasms.2009.03.018
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Although TOF analyzers with orthogonal ion injection provide the whole spectrum without scanning, their duty cycle is low compared with scanning analyzers in single ion monitoring mode. Typical duty cycle is in the range of 5% to 30% depending on the instrument geometry and ion m/z value. We present here a novel trapping/releasing setup, which offers the duty cycle near 100% over a wide range. Operation in the mass range from m/z 120 to almost 2000 is demonstrated. Ions are trapped in a short linear ion trap at the end of the collision cell in an axial pseudopotential well created by additional rf ("AC") voltage applied to all four rods of the trap with the same amplitude and phase. The pseudopotential created by AC field is mass dependent, and by ramping down the AC voltage, ions can be released from the trap sequentially from high m/z to low, while all ions are gaining the same kinetic energy. Upon entering the TOF accelerator, ions with lower m/z catch up with heavier ions, and the AC ramp parameters can be selected to make all ions meet in the center of the TOF extraction region, resulting in sensitivity gains from 3 to 14 without loss of mass accuracy or resolution. (J Am Soc Mass Spectrom 2009, 20,1342-1348) (C) 2009 American Society for Mass Spectrometry
引用
收藏
页码:1342 / 1348
页数:7
相关论文
共 15 条
[1]   Improvement of the duty cycle of an orthogonal acceleration time-of-flight mass spectrometer using ion gates [J].
Brenton, A. G. ;
Krastev, T. ;
Rousell, D. J. ;
Kennedy, M. A. ;
Craze, A. S. ;
Williams, C. M. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2007, 21 (18) :3093-3102
[2]   Duty cycle improvement for a quadrupole-time-of-flight mass spectrometer and its use for precursor ion scans [J].
Chernushevich, IV .
EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2000, 6 (06) :471-479
[3]   An introduction to quadrupole-time-of-flight mass spectrometry [J].
Chernushevich, IV ;
Loboda, AV ;
Thomson, BA .
JOURNAL OF MASS SPECTROMETRY, 2001, 36 (08) :849-865
[4]   Orthogonal-injection TOFMS for analyzing biomolecules [J].
Chernushevich, IV ;
Ens, W ;
Standing, KG .
ANALYTICAL CHEMISTRY, 1999, 71 (13) :452A-461A
[5]  
DODONOV AF, 1994, ACS SYM SER, V549, P108
[6]  
Gerlich D., 1992, ADV CHEM PHYS, V82
[7]  
Guilhaus M, 2000, MASS SPECTROM REV, V19, P65, DOI 10.1002/(SICI)1098-2787(2000)19:2<65::AID-MAS1>3.0.CO
[8]  
2-E
[9]   Duty cycle enhancement of an orthogonal acceleration TOF mass spectrometer using an axially-resonant excitation linear ion trap [J].
Hashimoto, Yuichiro ;
Hasegawa, Hideki ;
Satake, Hiroyuki ;
Baba, Takashi ;
Waki, Izumi .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2006, 17 (12) :1669-1674
[10]   Novel Linac II electrode geometry for creating an axial field in a multipole ion guide [J].
Loboda, A ;
Krutchinsky, A ;
Loboda, O ;
McNabb, J ;
Spicer, V ;
Ens, W ;
Standing, K .
EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2000, 6 (06) :531-536