Quadratic regularization design for 3d cylindrical PET

被引:0
作者
Shi, Hugo [1 ]
Fessler, Jeffrey A. [1 ]
机构
[1] Univ Michigan, Dept Elect Engn & Comp Sci, Ann Arbor, MI 48109 USA
来源
2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5 | 2005年
关键词
RECONSTRUCTION; RESOLUTION;
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Statistical methods for tomographic image reconstruction lead to improved spatial resolution and noise properties in PET. Penalized-likelihood (PL) image reconstruction methods involve maximizing an objective function that is based on the log-likelihood of the sinogram measurements and on a roughness penalty function to control noise. In emission tomography, PL methods (and MAP methods) based on conventional quadratic regularization functions lead to nonuniform and anisotropic spatial resolution, even for idealized shift-invariant imaging systems. We have previously addressed this problem for parallel-beam 2D emission tomography [1], and for fan-beam 2D transmission tomography [2] by designing data-dependent, shift-variant regularizers that improve resolution uniformity and isotropy. even for idealized shift-invariant imaging systems. This paper extends those methods to 3D cylindrical PET, using an analytical design approach that is numerically efficient.
引用
收藏
页码:2301 / 2305
页数:5
相关论文
共 5 条
[1]   Spatial resolution properties of penalized-likelihood image reconstruction: Space-invariant tomographs [J].
Fessler, JA ;
Rogers, WL .
IEEE TRANSACTIONS ON IMAGE PROCESSING, 1996, 5 (09) :1346-1358
[2]  
Fessler JA, 2004, IEEE NUCL SCI CONF R, P2022
[3]   Resolution and noise properties of MAP reconstruction for fully 3-D PET [J].
Qi, JY ;
Leahy, RM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (05) :493-506
[4]   Quadratic regularization design for fan beam transmission tomography [J].
Shi, H ;
Fessler, JA .
MEDICAL IMAGING 2005: IMAGE PROCESSING, PT 1-3, 2005, 5747 :2023-2033
[5]  
STAYMAN JW, 1999, BIOMEDICAL IMAGING D, pA4