Corrosion and Adverse Local Tissue Reaction in One Type of Modular Neck Stem

被引:32
作者
Ghanem, Elie [1 ]
Ward, Daniel M. [2 ]
Robbins, Claire E. [2 ]
Nandi, Sumon [2 ]
Bono, James V. [2 ]
Talmo, Carl T. [2 ]
机构
[1] Geisinger Hlth Syst, Dept hoped Surg, Danville, PA USA
[2] New England Baptist Hosp, Dept Orthopaed Surg, Boston, MA 02120 USA
关键词
dual modular stem; Rejuvenate stem; corrosion; pseudotumor; TOTAL HIP-ARTHROPLASTY; METAL-ION LEVELS; FEMORAL-NECK; REPLACEMENT; PROSTHESES; JUNCTION; PSEUDOTUMORS; INTERFACE; CORRELATE; CHROMIUM;
D O I
10.1016/j.arth.2015.04.039
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Modular neck stems allow for optimization of joint biomechanics by restoring anteversion, offset, and limb length. A potential disadvantage is the generation of metal ions from fretting and crevice corrosion. We identified 118 total hip arthroplasty implanted with one type of dual-modular femoral component. Thirty-six required revision due to adverse local tissue reaction. Multivariate analysis isolated females and low offset necks as risk factors for failure. Kaplan-Meir analysis revealed small stem sizes failed at a higher rate during early follow-up period. Although the cobalt/chrome levels were higher in the failed group, these tests had low diagnostic accuracy for ALTR, while MRI scan was more sensitive. We conclude that the complications related to the use of dual modular stems of this design outweigh the potential benefits. (c) 2015 Elsevier Inc. All rights reserved.
引用
收藏
页码:1787 / 1793
页数:7
相关论文
共 35 条
[11]  
Grammatopoulos George, 2013, J Bone Joint Surg Am, V95, pe81, DOI 10.2106/JBJS.L.00775
[12]   Are Metal Ion Levels a Useful Trigger for Surgical Intervention? [J].
Griffin, William L. ;
Fehring, Thomas K. ;
Kudrna, James C. ;
Schmidt, Robert H. ;
Christie, Michael J. ;
Odum, Susan M. ;
Dennos, Anne C. .
JOURNAL OF ARTHROPLASTY, 2012, 27 (08) :32-36
[13]   Modular titanium alloy neck adapter failures in hip replacement - failure mode analysis and influence of implant material [J].
Grupp, Thomas M. ;
Weik, Thomas ;
Bloemer, Wilhelm ;
Knaebel, Hanns-Peter .
BMC MUSCULOSKELETAL DISORDERS, 2010, 11
[14]   Immune responses correlate with serum-metal in metal-on-metal hip arthroplasty [J].
Hallab, NJ ;
Anderson, S ;
Caicedo, M ;
Skipor, A ;
Campbell, P ;
Jacobs, JJ .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :88-93
[15]  
Hart AJ, 2012, J BONE JOINT SURG AM, V94A, P317, DOI [10.2106/JBJS.J.01508, 10.2106/JBJS.K.01266]
[16]   Do Survival Rate and Serum Ion Concentrations 10 Years After Metal-on-Metal Hip Resurfacing Provide Evidence for Continued Use? [J].
Hartmann, Albrecht ;
Luetzner, Joerg ;
Kirschner, Stephan ;
Witzleb, Wolf-Christoph ;
Guenther, Klaus-Peter .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (11) :3118-3126
[17]   A MRI classification of periprosthetic soft tissue masses (pseudotumours) associated with metal-on-metal resurfacing hip arthroplasty [J].
Hauptfleisch, Jennifer ;
Pandit, Hemant ;
Grammatopoulos, George ;
Gill, Harinderjit S. ;
Murray, David W. ;
Ostlere, Simon .
SKELETAL RADIOLOGY, 2012, 41 (02) :149-155
[18]   Influence of material coupling and assembly condition on the magnitude of micromotion at the stem-neck interface of a modular hip endoprosthesis [J].
Jauch, S. Y. ;
Huber, G. ;
Hoenig, E. ;
Baxmann, M. ;
Grupp, T. M. ;
Morlock, M. M. .
JOURNAL OF BIOMECHANICS, 2011, 44 (09) :1747-1751
[19]   Corrosion of a Hip Stem With a Modular Neck Taper Junction A Retrieval Study of 16 Cases [J].
Kop, Alan M. ;
Swarts, Eric .
JOURNAL OF ARTHROPLASTY, 2009, 24 (07) :1019-1023
[20]  
LIEBERMAN JR, 1994, CLIN ORTHOP RELAT R, P162