Dosimetric verification for intensity-modulated arc therapy plans by use of 2D diode array, radiochromic film and radiosensitive polymer gel

被引:10
作者
Hayashi, Naoki [1 ]
Malmin, Ryan L. [2 ]
Watanabe, Yoichi [3 ]
机构
[1] Fujita Hlth Univ, Sch Hlth Sci, Fac Radiol Technol, Toyoake, Aichi 4701192, Japan
[2] Ridgeview Reg Radiat Oncol, Waconia, MN 55387 USA
[3] Dept Radiat Oncol, Minneapolis, MN 55455 USA
关键词
IMAT; dosimetric verification; 3D dose measurement; radiosensitive polymer gel; radiochromic films; 2D array detector; RADIATION-THERAPY; QUALITY-ASSURANCE; TOMOTHERAPY; IMRT;
D O I
10.1093/jrr/rrt139
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Several tools are used for the dosimetric verification of intensity-modulated arc therapy (IMAT) treatment delivery. However, limited information is available for composite on-line evaluation of these tools. The purpose of this study was to evaluate the dosimetric verification of IMAT treatment plans using a 2D diode array detector (2D array), radiochromic film (RCF) and radiosensitive polymer gel dosimeter (RPGD). The specific verification plans were created for IMAT for two prostate cancer patients by use of the clinical treatment plans. Accordingly, the IMAT deliveries were performed with the 2D array on a gantry-mounting device, RCF in a cylindrical acrylic phantom, and the RPGD in two cylindrical phantoms. After the irradiation, the planar dose distributions from the 2D array and the RCFs, and the 3D dose distributions from the RPGD measurements were compared with the calculated dose distributions using the gamma analysis method (3% dose difference and 3-mm distance-to-agreement criterion), dose-dependent dose difference diagrams, dose difference histograms, and isodose distributions. The gamma passing rates of 2D array, RCFs and RPGD for one patient were 99.5%, 96.5% and 93.7%, respectively; the corresponding values for the second patient were 97.5%, 92.6% and 92.9%. Mean percentage differences between the RPGD measured and calculated doses in 3D volumes containing PTVs were -0.29 +/- 7.1% and 0.97 +/- 7.6% for the two patients, respectively. In conclusion, IMAT prostate plans can be delivered with high accuracy, although the 3D measurements indicated less satisfactory agreement with the treatment plans, mainly due to the dosimetric inaccuracy in low-dose regions of the RPGD measurements.
引用
收藏
页码:541 / 552
页数:12
相关论文
共 17 条
[1]   Using an EPID for patient-specific VMAT quality assurance [J].
Bakhtiari, M. ;
Kumaraswamy, L. ;
Bailey, D. W. ;
de Boer, S. ;
Malhotra, H. K. ;
Podgorsak, M. B. .
MEDICAL PHYSICS, 2011, 38 (03) :1366-1373
[2]   Polymer gel dosimetry [J].
Baldock, C. ;
De Deene, Y. ;
Doran, S. ;
Ibbott, G. ;
Jirasek, A. ;
Lepage, M. ;
McAuley, K. B. ;
Oldham, M. ;
Schreiner, L. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05) :R1-R63
[3]   RapidArc treatment verification in 3D using polymer gel dosimetry and Monte Carlo simulation [J].
Ceberg, Sofie ;
Gagne, Isabelle ;
Gustafsson, Helen ;
Scherman, Jonas Bengtsson ;
Korreman, Stine S. ;
Kjaer-Kristoffersen, Flemming ;
Hilts, Michelle ;
Back, Sven A. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (17) :4885-4898
[4]   American College of Radiology (ACR) and American Society for Radiation Oncology (ASTRO) Practice Guideline for Intensity-modulated Radiation Therapy (IMRT) [J].
Hartford, Alan C. ;
Galvin, James M. ;
Beyer, David C. ;
Eichler, Thomas J. ;
Ibbott, Geoffrey S. ;
Kavanagh, Brian ;
Schultz, Christopher J. ;
Rosenthal, Seth A. .
AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 2012, 35 (06) :612-617
[5]   Evaluation of triple channel correction acquisition method for radiochromic film dosimetry [J].
Hayashi, Naoki ;
Watanabe, Yoichi ;
Malmin, Ryan ;
Kato, Hideki .
JOURNAL OF RADIATION RESEARCH, 2012, 53 (06) :930-935
[6]   Dosimetric verification of RapidArc treatment delivery [J].
Korreman, Stine ;
Medin, Joakim ;
Kjaer-Kristoffersen, Flemming .
ACTA ONCOLOGICA, 2009, 48 (02) :185-191
[7]   Dosimetry tools and techniques for IMRT [J].
Low, Daniel A. ;
Moran, Jean M. ;
Dempsey, James F. ;
Dong, Lei ;
Oldham, Mark .
MEDICAL PHYSICS, 2011, 38 (03) :1313-1338
[8]   Effects of phantom volume and shape on the accuracy of MRI BANG gel closimetry using BANG3™ [J].
MacDougall, N. D. ;
Miquel, M. E. ;
Keevil, S. F. .
BRITISH JOURNAL OF RADIOLOGY, 2008, 81 (961) :46-50
[9]   3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry [J].
Mans, Anton ;
Remeijer, Peter ;
Olaciregui-Ruiz, Igor ;
Wendling, Markus ;
Sonke, Jan-Jakob ;
Mijnheer, Ben ;
van Herk, Marcel ;
Stroom, Joep C. .
RADIOTHERAPY AND ONCOLOGY, 2010, 94 (02) :181-187
[10]   Evaluation of the basic properties of the BANGkit™ gel dosimeter [J].
Murakami, Y. ;
Nakashima, T. ;
Watanabe, Y. ;
Akimitsu, T. ;
Matsuura, K. ;
Kenjo, M. ;
Kaneyasu, Y. ;
Wadasaki, K. ;
Hirokawa, Y. ;
Ito, K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (08) :2301-2311