Modulation transfer function: Rigid versus foldable phakic intraocular lenses

被引:25
作者
Artigas, Jose M. [1 ,2 ]
Peris, Cristina [2 ]
Felipe, Adelina [1 ,2 ]
Menezo, Jose L. [2 ]
Sanchez-Cortina, Isaias [1 ,2 ]
Lopez-Gil, Norberto [3 ]
机构
[1] Univ Valencia, Fac Fis, Dept Opt, E-46100 Valencia, Spain
[2] Fdn Oftalmol Mediterraneo, Valencia, Spain
[3] Univ Murcia, Dept Fis, E-30071 Murcia, Spain
关键词
INCISION CATARACT-SURGERY; MODERATELY HIGH MYOPIA; PUPIL SIZE; OPTICAL-QUALITY; HUMAN EYE; ARTISAN; IMPLANTATION; PERFORMANCE; ABERRATIONS; ASTIGMATISM;
D O I
10.1016/j.jcrs.2008.12.020
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To study in a standard eye model the changes in modulation transfer function (MTF) of a monofocal intraocular lens (IOL) when a phakic IOL (pIOL) is placed in the anterior chamber, compare the MTFs of the rigid Artisan pIOL and foldable Artiflex pIOL, and evaluate the temporal evolution of the MTF of the foldable pIOL after the mechanical stress the pIOL undergoes when injected. SETTING: Fundacion Oftalmologica del Mediterraneo, Valencia, Spain. METHODS: The MTF values of the IOLs were calculated from the cross-line spread function recorded with the Opal Vector System. The measurements were taken using an eye model following the British and EN-ISO standards with 2.0 mm, 3.0 mm, 4.0 mm, and 5.0 mm pupils. A 28.00 diopter (D) Ophtec monofocal IOL was used as the crystalline lens. The 2 pIOLs were -9.00 D. RESULTS: The MTF of the rigid pIOL was slightly better than the MTF of the foldable pIOL with all pupil sizes. Both pIOLs provided good optics quality when compared with the monofocal IOL. The injection effect of the foldable IOL disappeared after 2 hours. CONCLUSIONS: The MTF of the monofocal IOL was slightly reduced with implantation of a negative pIOL in the anterior chamber. The rigid pIOL provided better optical performance than the foldable pIOL with all pupil sizes, as shown by the MTF values. The decrease in MTF caused by the mechanical stress on the foldable pIOL was nullified after 2 hours with no effect on optical quality.
引用
收藏
页码:747 / 752
页数:6
相关论文
共 27 条
[1]   Optical performance of 3 intraocular lens designs in the presence of decentration [J].
Altmann, GE ;
Nichamin, LD ;
Lane, SS ;
Pepose, JS .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (03) :574-585
[2]  
[Anonymous], 9335 ISO
[3]   Image quality with multifocal intraocular lenses and the effect of pupil size -: Comparison of refractive and hybrid refractive-diffractive designs [J].
Artigas, Jose M. ;
Menezo, Jose L. ;
Peris, Cristina ;
Felipe, Adelina ;
Diaz-Llopis, Manuel .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (12) :2111-2117
[4]   Astigmatism outcomes of horizontal temporal versus nasal dear corneal incision cataract surgery [J].
Barequet, IS ;
Yu, E ;
Vitale, S ;
Cassard, S ;
Azar, DT ;
Stark, WJ .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2004, 30 (02) :418-423
[5]   Corneal topographic changes induced by different oblique cataract incisions [J].
Beltrame, G ;
Salvetat, ML ;
Chizzolini, M ;
Driussi, G .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (05) :720-727
[6]   OPTICAL QUALITY OF HUMAN EYE [J].
CAMPBELL, FW ;
GUBISCH, RW .
JOURNAL OF PHYSIOLOGY-LONDON, 1966, 186 (03) :558-+
[7]   Glistenings in the Artiflex phakic intraocular lens [J].
Cisneros-Lanuza, Angel ;
Hurtado-Sarrio, Mercedes ;
Duch-Samper, Antonio ;
Gallego-Pinazo, Roberto ;
Menezo-Rozalen, Jose L. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (08) :1405-1408
[8]   Iris-supported phakic lenses (rigid vs foldable version) for treating moderately high myopia:: Randomized paired eye comparison [J].
Coullet, Julien ;
Guell, Jose-Luis ;
Fournie, Pierre ;
Grandjean, Helene ;
Gaytan, Javier ;
Arne, Jean-Louis ;
Malecaze, Franicois .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 142 (06) :909-916
[9]   Change in pupil size after implantation of an iris-fixated toric phakic intraocular lens [J].
Dick, HB ;
Aliyeva, S ;
Tehrani, M .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (02) :302-307
[10]   Comparison of iris-fixed Artisan lens implantation with excimer laser in situ keratomileusis in correcting myopia between-9.00 and-19.50 diopters - A randomized study [J].
El Danasoury, MA ;
El Maghraby, A ;
Gamali, TO .
OPHTHALMOLOGY, 2002, 109 (05) :955-964