Endoscope reprocessing: Comparison of drying effectiveness and microbial levels with an automated drying and storage cabinet with forced filtered air and a standard storage cabinet

被引:37
作者
Perumpail, Ryan B. [1 ]
Marya, Neil B. [1 ]
McGinty, Betty L. [2 ]
Muthusamy, V. Raman [1 ]
机构
[1] UCLA, David Geffen Sch Med, Vatche & Tamar Manoukian Div Digest Dis, 200 UCLA Med Plaza,Rm 330-37, Los Angeles, CA 90095 USA
[2] Northside Hosp, Dept Gastroenterol Serv, Atlanta, GA USA
关键词
Duodenoscope; Drying time; High-level disinfection; Waterborne infection; MULTISOCIETY GUIDELINE; INFECTIONS; RESISTANT; DUODENOSCOPES; DISINFECTION;
D O I
10.1016/j.ajic.2019.02.016
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Background: Automated drying may help prevent endoscopically transmitted infections. We aimed to assess the efficacy of an automated drying and storage cabinet compared to a standard storage cabinet in achieving endoscope dryness postreprocessing and in reducing the risk of microbial growth. Methods: Drying times of bronchoscopes, colonoscopes, and duodenoscopes using 2 drying platforms (an automated drying and storage cabinet vs a standard storage cabinet) were measured using cobalt chloride paper. Drying assessments occurred at: 30 minutes, 1 hour, 2 hours, 3 hours, and 24 hours. A simple linear regression analysis compared rates of microbial growth after inoculation with Pseudomonas aeruginosa following high-level disinfection at: 0, 3 hours, 12 hours, 24 hours, and 48 hours. Results: Using the automated drying and storage cabinet, internal channels were dry at 1 hour and external surfaces at 3 hours in all endoscopes. With the standard storage cabinet, there was residual internal fluid at 24 hours, whereas external surfaces were dry at 24 hours. For bronchoscopes, colonoscopes, and duodenoscopes, the standard cabinet allowed for an average rate of colony forming unit growth of 8.1 x 10(6) per hour, 8.3 x 10(6) per hour, and 7.0 x 10 7 per hour, respectively; the automated cabinet resulted in colony forming unit growth at an average rate of -28.4 per hour (P = .02), -38.5 per hour (P = .01), and -200.2 per hour (P = .02), respectively. Conclusions: An automated cabinet is advantageous for rapid drying of endoscope surfaces and in reducing the risk of microbial growth postreprocessing. (C) 2019 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:1083 / 1089
页数:7
相关论文
共 30 条
  • [1] A novel polytetrafluoroethylene-channel model, which simulates low levels of culturable bacteria in buildup biofilm after repeated endoscope reprocessing
    Alfa, Michelle J.
    Ribeiro, Maira M.
    Luciano, Cristiana da Costa
    Franca, Rodrigo
    Olson, Nancy
    DeGagne, Pat
    Singh, Harminder
    [J]. GASTROINTESTINAL ENDOSCOPY, 2017, 86 (03) : 442 - U350
  • [2] IN-HOSPITAL EVALUATION OF CONTAMINATION OF DUODENOSCOPES - A QUANTITATIVE ASSESSMENT OF THE EFFECT OF DRYING
    ALFA, MJ
    SITTER, DL
    [J]. JOURNAL OF HOSPITAL INFECTION, 1991, 19 (02) : 89 - 98
  • [3] PSEUDOMONAS INFECTION OF THE BILIARY SYSTEM RESULTING FROM USE OF A CONTAMINATED ENDOSCOPE
    ALLEN, JI
    ALLEN, MO
    OLSON, MM
    GERDING, DN
    SHANHOLTZER, CJ
    MEIER, PB
    VENNES, JA
    SILVIS, SE
    [J]. GASTROENTEROLOGY, 1987, 92 (03) : 759 - 763
  • [4] [Anonymous], 2015, 164422015 BS EN BRIT
  • [5] Gastroscopy-associated transmission of extended-spectrum beta-lactamase-producing Pseudomonas aeruginosa
    Bajolet, O.
    Ciocan, D.
    Vallet, C.
    de Champs, C.
    Vernet-Garnier, V.
    Guillard, T.
    Brasme, L.
    Thiefin, G.
    Cadiot, G.
    Bureau-Chalot, F.
    [J]. JOURNAL OF HOSPITAL INFECTION, 2013, 83 (04) : 341 - 343
  • [6] Comparison of automated and manual drying in the elimination of residual endoscope working channel fluid after reprocessing
    Barakat, Monique T.
    Huang, Robert J.
    Banerjee, Subhas
    [J]. GASTROINTESTINAL ENDOSCOPY, 2019, 89 (01) : 124 - +
  • [7] Scoping the scope: endoscopic evaluation of endoscope working channels with a new high-resolution inspection endoscope (with video)
    Barakat, Monique T.
    Girotra, Mohit
    Huang, Robert J.
    Banerjee, Subhas
    [J]. GASTROINTESTINAL ENDOSCOPY, 2018, 88 (04) : 601 - +
  • [8] A randomized trial of single versus double high-level disinfection of duodenoscopes and linear echoendoscopes using standard automated reprocessing
    Bartles, Rebecca L.
    Leggett, James E.
    Hove, Shannan
    Kashork, Catherine D.
    Wang, Lian
    Oethinger, Margret
    Baxter, Lynda
    Brandabur, Jack J.
    [J]. GASTROINTESTINAL ENDOSCOPY, 2018, 88 (02) : 306 - +
  • [9] Guideline Implementation: Processing Flexible Endoscopes
    Bashaw, Marie A.
    [J]. AORN JOURNAL, 2016, 104 (03) : 226 - 233
  • [10] Efficacy of low-temperature plasma-activated gas disinfection against biofilm on contaminated GI endoscope channels
    Bhatt, Sudhir
    Mehta, Poonam
    Chen, Chen
    Schneider, Christina L.
    White, Lauren N.
    Chen, Hai-Lan
    Kong, Michael G.
    [J]. GASTROINTESTINAL ENDOSCOPY, 2019, 89 (01) : 105 - 114