Role of beat noise in limiting the sensitivity of optical coherence tomography

被引:10
作者
Haskell, Richard C. [1 ]
Liao, David [1 ]
Pivonka, Adam E. [1 ]
Bell, Tera L. [1 ]
Haberle, Brendan R. [1 ]
Hoeling, Barbara M. [1 ]
Petersen, Daniel C. [1 ]
机构
[1] Harvey Mudd Coll, Dept Phys, Claremont, CA 91711 USA
关键词
D O I
10.1364/JOSAA.23.002747
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The sensitivity and dynamic range of optical coherence tomography (OCT) are calculated for instruments utilizing two common interferometer configurations and detection schemes. Previous researchers recognized that the performance of dual-balanced OCT instruments is severely limited by beat noise, which is generated by incoherent light backscattered from the sample. However, beat noise has been ignored in previous calculations of Michelson OCT performance. Our measurements of instrument noise confirm the presence of beat noise even in a simple Michelson interferometer configuration with a single photodetector. Including this noise, we calculate the dynamic range as a function of OCT light source power, and find that instruments employing balanced interferometers and balanced detectors can achieve a sensitivity up to six times greater than those based on a simple Michelson interferometer, thereby boosting image acquisition speed by the same factor for equal image quality. However, this advantage of balanced systems is degraded for source powers greater than a few milliwatts. We trace the concept of beat noise back to an earlier paper.
引用
收藏
页码:2747 / 2755
页数:9
相关论文
共 22 条
  • [1] HIGH-RESOLUTION OPTICAL COHERENCE TOMOGRAPHIC IMAGING USING A MODE-LOCKED TI-AL2O3 LASER SOURCE
    BOUMA, B
    TEARNEY, GJ
    BOPPART, SA
    HEE, MR
    BREZINSKI, ME
    FUJIMOTO, JG
    [J]. OPTICS LETTERS, 1995, 20 (13) : 1486 - 1488
  • [2] Power-efficient nonreciprocal interferometer and linear-scanning fiber-optic catheter for optical coherence tomography
    Bouma, BE
    Tearney, GJ
    [J]. OPTICS LETTERS, 1999, 24 (08) : 531 - 533
  • [3] Sensitivity advantage of swept source and Fourier domain optical coherence tomography
    Choma, MA
    Sarunic, MV
    Yang, CH
    Izatt, JA
    [J]. OPTICS EXPRESS, 2003, 11 (18): : 2183 - 2189
  • [4] Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography
    de Boer, JF
    Cense, B
    Park, BH
    Pierce, MC
    Tearney, GJ
    Bouma, BE
    [J]. OPTICS LETTERS, 2003, 28 (21) : 2067 - 2069
  • [5] Comparative signal-to-noise analysis of fibre-optic based optical coherence tomography systems
    Ford, HD
    Beddows, R
    Casaubieilh, P
    Tatam, RP
    [J]. JOURNAL OF MODERN OPTICS, 2005, 52 (14) : 1965 - 1979
  • [6] STATISTICS OF THERMAL AND LASER RADIATION
    HODARA, H
    [J]. PROCEEDINGS OF THE INSTITUTE OF ELECTRICAL AND ELECTRONICS ENGINEERS, 1965, 53 (07): : 696 - &
  • [7] Improved phase modulation for an en-face scanning three-dimensional optical coherence microscope
    Hoeling, BM
    Peter, ME
    Petersen, DC
    Haskell, RC
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (10) : 3348 - 3350
  • [8] An optical coherence microscope for 3-dimensional imaging in developmental biology
    Hoeling, BM
    Fernandez, AD
    Haskell, RC
    Huang, E
    Myers, WR
    Petersen, DC
    Ungersma, SE
    Wang, RY
    Williams, ME
    Fraser, SE
    [J]. OPTICS EXPRESS, 2000, 6 (07): : 136 - 146
  • [9] Phase modulation at 125 kHz in a Michelson interferometer using an inexpensive piezoelectric stack driven at resonance
    Hoeling, BM
    Fernandez, AD
    Haskell, RC
    Petersen, DC
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (03) : 1630 - 1633
  • [10] Performance of fourier domain vs. time domain optical coherence tomography
    Leitgeb, R
    Hitzenberger, CK
    Fercher, AF
    [J]. OPTICS EXPRESS, 2003, 11 (08): : 889 - 894