Technical Reliability and Clinical Safety of a Remote Monitoring System for Antiarrhythmic Cardiac Devices

被引:16
作者
Osca, Joaquin [1 ]
Sancho Tello, Maria Jose [1 ]
Navarro, Josep [1 ]
Cano, Oscar [1 ]
Raso, Rafael [1 ]
Enrique Castro, Jose [1 ]
Olaguee, Jose [1 ]
Salvador, Antonio [1 ]
机构
[1] Hosp Univ La Fe, Serv Cardiol, Unidad Arritmias, Valencia, Spain
来源
REVISTA ESPANOLA DE CARDIOLOGIA | 2009年 / 62卷 / 08期
关键词
Pacemaker; Defibrillator; Remote monitoring; IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS; RESYNCHRONIZATION THERAPY SYSTEMS; FOLLOW-UP; PACEMAKER; FREQUENCY;
D O I
10.1016/S0300-8932(09)72071-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction and objectives. In recent years, physicians have experienced a huge increase in workload associated with cardiac devices. Remote monitoring enables these devices to be monitored at a distance and could reduce the number of hospital visits. The aim of this study was to assess the technical reliability and clinical safety of the Home Monitoring (TM) remote monitoring system (Biotronik GmbH, Germany). Methods. The study included 162 patients: 115 with pacemakers, 44 with implantable cardioverter-defibrillators (ICDs) and three with ICDs incorporating cardiac resynchronization therapy (CRT). Patients were followed up for a mean of 303 +/- 244 days. We compared the survival time free of event alerts in patients with pacemakers and in those with ICDs. The remote monitoring system's reliability was evaluated by comparing the data it transmitted with clinical observations made immediately afterwards. Results. The mean percentage of days monitored was 88.2%. The incidence of event alerts was higher in pacemaker than ICD patients (45% vs. 34%; P=.01). Event alerts preceded the following hospital visit by 76 +/- 47 days. In the ICD group, the data sent by the monitoring system were judged as true-positives in 42% of cases, as true-negatives in 57% and as false-positives in 1%. No false-negative was observed. In the ICD group, the remote monitoring system had a sensitivity of 100% and a specificity of 97% (positive predictive value 96%, negative predictive value 100%). Conclusions. Remote monitoring was a reliable and safe method for following up patients with cardiac devices. Its routine use could enable the early detection of device malfunctions or arrhythmic events.
引用
收藏
页码:886 / 895
页数:10
相关论文
共 16 条
  • [1] What evidence do we have to replace in-hospital implantable cardioverter clefibrillator follow-up?
    Brugada, P
    [J]. CLINICAL RESEARCH IN CARDIOLOGY, 2006, 95 (Suppl 3) : 3 - 9
  • [2] Elsner C.H., 2006, Comput Cardiol, V33, P241
  • [3] Pacemaker follow-up with prolonged intervals in the stable period 1 to 5 years postimplant
    Grendahl, H
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (08): : 1219 - 1224
  • [4] Potential role of remote monitoring for scheduled and unscheduled evaluations of patients with an implantable defibrillator
    Heidbuchel, Hein
    Lioen, Pieter
    Foulon, Stefaan
    Huybrechts, Wim
    Ector, Joris
    Wiliems, Rik
    Ector, Hugo
    [J]. EUROPACE, 2008, 10 (03): : 349 - 355
  • [5] Remote interrogation and monitoring of implantable cardioverter defibrillators
    Joseph, GK
    Wilkoff, BL
    Dresing, T
    Burkhardt, J
    Khaykin, Y
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2004, 11 (02) : 161 - 166
  • [6] Advances in remote monitoring of implantable pacemakers, cardioverter defibrillators and cardiac resynchronization therapy systems
    Jung, Werner
    Rillig, Andreas
    Birkemeyer, Ralf
    Miljak, Tomislav
    Meyerfeldt, Udo
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2008, 23 (01) : 73 - 85
  • [7] Lazarus A, 2007, PACE, V30, pS2
  • [8] Early automatic remote detection of combined lead insulation defect and ICD damage
    Neuzil, Petr
    Taborsky, Milos
    Holy, Frantisek
    Wallbrueck, Karsten
    [J]. EUROPACE, 2008, 10 (05): : 556 - 557
  • [9] Automatic home monitoring of implantable cardioverter defibrillators
    Nielsen, Jens Cosedis
    Kottkamp, Hans
    Zabel, Markus
    Aliot, Etienne
    Kreutzer, Ulrich
    Bauer, Alexander
    Schuchert, Andreas
    Neuser, Hans
    Schumacher, Burghard
    Schmidinger, Herwig
    Stix, Guenter
    Clementy, Jacques
    Danilovic, Dejan
    Hindricks, Gerhard
    [J]. EUROPACE, 2008, 10 (06): : 729 - 735
  • [10] The role of remote monitoring in the reduction of inappropriate implantable cardioverter defibrillator therapies
    Res, JCJ
    Theuns, DAMJ
    Jordaens, L
    [J]. CLINICAL RESEARCH IN CARDIOLOGY, 2006, 95 (Suppl 3) : 17 - 21