Computer optimization of noncoplanar beam setups improves stereotactic treatment of liver tumors

被引:23
作者
de Pooter, Jacco A. [1 ]
Romero, Alejandra Mendez [1 ]
Jansen, Wim P. A. [1 ]
Storchi, Pascal R. M. [1 ]
Levendag, Peter C. [1 ]
Heijmen, Ben J. M. [1 ]
机构
[1] Dr Daniel Den Hoed Canc Ctr, Dept Radiat Oncol, Erasmus MC, NL-3075 EA Rotterdam, Netherlands
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2006年 / 66卷 / 03期
关键词
inverse planning; liver; noncoplanar beams; computer optimization; stereotactic treatment;
D O I
10.1016/j.ijrobp.2006.06.018
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To investigate whether computer-optimized fully noncoplanar beam setups may improve treatment plans for the stereotactic treatment of liver tumors. Methods: An algorithm for automated beam orientation and weight selection (Cycle) was extended for noncoplanar stereotactic treatments. For 8 liver patients previously treated in our clinic using a prescription isodose of 65%, Cycle was used to generate noncoplanar and coplanar plans with the highest achievable minimum planning target volume (PTV) dose for the clinically delivered isocenter and mean liver doses, while not violating the clinically applied hard planning constraints. The clinical and the optimized coplanar and noncoplanar plans were compared, with respect to D-PTV,D-99%, the dose received by 99% of the PTV, the PTV generalized equivalent uniform dose (gEUD), and the compliance with the clinical constraints. Results: For each patient, the ratio between D-PTV,D-99% and D-isoc, and the gEUD-, and gEUD(-5), values of the optimized noncoplanar plan were higher than for the clinical plan with an average increase of respectively 18.8% (range, 7.8-24.0%), 6.4 Gy (range, 3.4-11.8 Gy), and 10.3 Gy (range, 6.7-12.5). D-PTV,D-99%/D-isoc, gEUD(-5), and gEUD-(20) of the optimized noncoplanar plan was always higher than for the optimized coplanar plan with an average increase of, respectively, 4.5% (range, 0.2-9.7%), 2.7 Gy (range, 0.6-9.7 Gy), and 3.4 Gy (range, 0.6-9.9 Gy). All plans were within the imposed hard constraints. On average, the organs at risk were better spared with the optimized noncoplanar plan than with the optimized coplanar plan and the clinical plan. Conclusions: The use of automatically generated, fully noncoplanar beam setups results in plans that are favorable compared with coplanar techniques. Because of the automation, we found that the planning workload can be decreased from 1 to 2 days to 1 to 2 h. (c) 2006 Elsevier Inc.
引用
收藏
页码:913 / 922
页数:10
相关论文
共 20 条
[1]   Simultaneous optimization of beam orientations, wedge filters and field weights for inverse planning with anatomy-based MLC fields [J].
Beaulieu, F ;
Beaulieu, L ;
Tremblay, D ;
Roy, R .
MEDICAL PHYSICS, 2004, 31 (06) :1546-1557
[2]   Automatic generation of anatomy-based MLC fields in aperture-based IMRT [J].
Beaulieu, F ;
Beaulieu, L ;
Tremblay, D ;
Lachance, B ;
Roy, R .
MEDICAL PHYSICS, 2004, 31 (06) :1539-1545
[3]   Biologic susceptibility of hepatocellular carcinoma patients treated with radiotherapy to radiation-induced liver disease [J].
Cheng, JCH ;
Wu, JK ;
Lee, PCT ;
Liu, HS ;
Jian, JJM ;
Lin, YM ;
Sung, JL ;
Jan, GJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (05) :1502-1509
[4]  
Dawson LA, 2002, INT J RADIAT ONCOL, V53, P810, DOI 10.1016/S0360-3016(02)02846-8
[5]  
Hadinger Ulrich, 2002, Z Med Phys, V12, P221
[6]   TARGET DOSE VERSUS EXTRATARGET DOSE IN STEREOTAXIC RADIOSURGERY [J].
LAX, I .
ACTA ONCOLOGICA, 1993, 32 (04) :453-457
[7]  
Niemierko A., 1999, MED PHYS, V26, P1100, DOI DOI 10.1118/1.598063
[8]   Local radiotherapy for patients with unresectable hepatocellular carcinoma [J].
Park, W ;
Lim, DH ;
Paik, SW ;
Koh, KC ;
Choi, MS ;
Park, CK ;
Yoo, BC ;
Lee, JE ;
Kang, MK ;
Park, YJ ;
Nam, HR ;
Ahn, YC ;
Huh, SJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 61 (04) :1143-1150
[9]   A Phase I trial of stereotactic body radiation therapy (SBRT) for liver metastases [J].
Schefter, TE ;
Kavanagh, BD ;
Timmerman, RD ;
Cardenes, HR ;
Baron, A ;
Gaspar, LE .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (05) :1371-1378
[10]   Potential benefits of eliminating planning target volume expansions for patient breathing in the treatment of liver tumors [J].
TenHaken, RK ;
Balter, JM ;
Marsh, LH ;
Robertson, JM ;
Lawrence, TS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (03) :613-617