摘要
The "dedicated" MRI units have characteristics of high diagnostic accuracy and lower installation management costs as compared with whole-body systems. The dedicated PL IRI units are easy to install. The low weight allows their installation also under unfavorable circumstances. In a dedicated system cost-effectiveness rind ease of installation must be accompanied by the capability of providing high-quality images. Tn our experience, the: high number of examinations performed, the most part of which provided with the surgical controls, allowed an accurate evaluation of the diagnostic potentialities of thr dedicated magnet. We were not able to perform the examinations in only 3% of cases due to the physical shape of the patient and thr clinical condition of the patient which may hinder the correct positioning of en the limb. The overlapping of the diagnostic accuracy of the E-scan and Artoscan units in thr study of the lower limbs, compared with whole-body units and surgery, prompted us to exploit the potentialities or the E-Scan in the study of the shoulder. We had a, good correlation between E-Scan, whole-body units, and surgical findings, which confirmed the high diagnostic accuracy of the dedicated system. In conclusion, in our experience carried out in the musculoskeletal system, the dedicated magnets showed promising results. Their diagnostic ri liability anti utility was comparable to that obtained from conventional units operating at higher magnetic fields.
引用
收藏
页码:250 / 255
页数:6
相关论文
共 12 条
  • [11] Magnetic resonance imaging on a low-field system (0.2 T): comparison with a 1.0-T system
    Trattnig, S
    Kontaxis, G
    Breitenseher, M
    Czerny, C
    Rand, T
    Turetschek, K
    Barth, M
    Imhof, H
    [J]. RADIOLOGE, 1997, 37 (10): : 773 - 777
  • [12] MRI of the foot and ankle: Diagnostic performance and patient acceptance of a dedicated low field MR scanner
    Verhoek, G
    Zanetti, M
    Duewell, S
    Zollinger, H
    Hodler, J
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (03): : 711 - 716