Disruption of the vascular prosthesis caused by aortic calcification after replacement of the thoracoabdominal aortic aneurysm

被引:10
作者
Tanaka, Hiroshi [1 ]
Okada, Kenji [1 ]
Yamashita, Teruo [1 ]
Kawanishi, Yujiro [1 ]
Matsumori, Masamiti [1 ]
Okita, Yutaka [1 ]
机构
[1] Kobe Univ Hosp, Dept Cardiovasc Thorac & Pediat Surg, Chuo Ku, Kobe, Hyogo 6500017, Japan
关键词
D O I
10.1016/j.athoracsur.2006.01.040
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A 52-year-old man underwent a replacement of the thoracoabdominal aorta. The aorta was severely calcified, and was replaced by a 24-mm woven Dacron (Vascutek, Renfrewshire, Scotland) graft wrapped with the calcificated aneurysmal wall. His postoperative course was uneventful; however, he collapsed on the 18th postoperative day. He underwent an emergent thoracotomy and the wrapped aneurysmal wall was taken down. The prosthesis graft had a 1-mm disruption in the middle portion, which did not relate to the anastomoses. Experimental study ex vivo showed that disruption of the prosthesis could have occurred after a 3-week pulsatile force caused by a seashell simulating aortic calcification.
引用
收藏
页码:1097 / 1099
页数:4
相关论文
共 2 条
[1]  
*CTR DEV RAD HLTH, 1993, FOOD DRUG ADM GUID P
[2]   Late disruption of Dacron aortic grafts [J].
Wilson, SE ;
Krug, R ;
Mueller, G ;
Wilson, L .
ANNALS OF VASCULAR SURGERY, 1997, 11 (04) :383-386