End-to-end dosimetric audit: A novel procedure developed for Irish HDR brachytherapy centres

被引:13
作者
Bassi, Sarah [1 ]
Berrigan, Leanne [2 ]
Zuchora, Anysja [3 ]
Fahy, Louise [2 ]
Moore, Margaret [2 ]
机构
[1] Azienda Sanitaria Univ, Friuli Cent, Udine, Italy
[2] Galway Univ Hosp, Med Phys & Bioengn, Galway, Ireland
[3] Hlth Serv Execut, Dept Med Phys & Bioeengn, Galway, Ireland
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2020年 / 80卷
关键词
Dosimetric audit; HDR brachytherapy; 3D printable materials; Ir-192; dosimetry; TLD; GafChromic films; Farmer chamber; PDR IR-192 BRACHYTHERAPY; CALIBRATION; ENERGY; CODE;
D O I
10.1016/j.ejmp.2020.10.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A dosimetric audit of Ir-192 high dose rate (HDR) brachytherapy remote after-loading units was carried out in 2019. All six brachytherapy departments on the island of Ireland participated in an end-to-end test and in a review of local HDR dosimetry procedures. Materials and methods: A 3D-printed customised phantom was created to position the following detectors at known distances from the HDR source: a Farmer ionization chamber, GafChromic film and thermoluminescent dosimeters (TLDs). Dedicated HDR applicator needles were used to position an Ir-192 source at 2 cm distance from these detectors. The end-to-end dosimetry audit pathway was performed at each host site and included the stages of imaging, applicator reconstruction, treatment planning and delivery. Deviations between planned and measured dose distributions were quantified using gamma analysis methods. Local procedures were also discussed between auditors and hosts. Results: The mean difference between Reference Air Kerma Rate (RAKR) measured during the audit and RAKR specified by the vendor source certificate was 1.3%. The results of end-to-end tests showed a mean difference between calculated and measured dose of 2.5% with TLDs and less than 0.5% with Farmer chamber measurements. GafChromic films showed a mean gamma passing rates of >95% for plastic and metal applicators with 2%/1 mm global tolerance criteria. Conclusions: The results of this audit indicate dosimetric consistency between centres. The 'end to end' dosimetry audit methodology for HDR brachytherapy has been successfully implemented in a multicentre environment, which included different models of Ir-192 sources and different treatment planning systems. The ability to create a 3D-printed water-equivalent phantom customised to accurately position all three detector types simultaneously at controlled distances from the Ir-192 source under evaluation gives good reproducibility for end-to-end methodology.
引用
收藏
页码:221 / 229
页数:9
相关论文
共 31 条
[1]  
[Anonymous], RADIOTHER ONCOL
[2]  
[Anonymous], 2003, CANC THER MISS TUM S
[3]  
[Anonymous], 2003, CRITICALLY ILL RECEI
[4]  
Aukett R J, 2005, Phys Med Biol, V50, P2739, DOI 10.1088/0031-9155/50/12/001
[5]   Energy and dose dependence of GafChromic EBT3-V3 film across a wide energy range [J].
Bassi, Sarah ;
Cummins, Donal ;
McCavana, Patrick .
REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2020, 25 (01) :60-63
[6]   Dosimetry assessment of patient-specific 3D printable materials for HDR surface brachytherapy [J].
Bassi, Sarah ;
Langan, Brian ;
Malone, Ciaran .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 67 :166-175
[7]  
Diez P, 2015, RADIOTHER ONCOL, V115, pS139
[8]  
Eckert Ziegler BEBIG GmbH, 2014, 2014LCT82 ECK ZIEG B
[9]  
German Standard, 1988, APPL XRAYS TUBE VOLT
[10]   Australasian brachytherapy audit: Results of the "end-to-end' dosimetry pilot study [J].
Haworth, Annette ;
Wilfert, Lisa ;
Butler, Duncan ;
Ebert, Martin A. ;
Todd, Stephen ;
Bucci, Joseph ;
Duchesne, Gillian M. ;
Joseph, David ;
Kron, Tomas .
JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2013, 57 (04) :490-498