摘要
The need to quantify blood flow through the heart has led to the development of different techniques for its measurement. The 3 main approaches are the Fick method, dye dilution, and thermodilution techniques. The latter 2 are based on the use of indicators that indirectly quantify blood flow. These have slowly been developed over centuries, from the concept of measuring blood flow, to a technique, and its clinical utility. Thermodilution is the most popular dilution method used for measuring cardiac output ( CO) in the clinical setting. The information obtained during this procedure is relevant in the process of clinical decision making in patients with critical illness, valvular heart disease, and congestive heart failure. The technique increased in popularity in the early 1970' s after Swan and Ganz invented the pulmonary artery catheter that simplified thermodilution enough to utilize it as a bedside procedure. This was only possible with simple yet clever engineering methods that are not commonly known in the medical community. Despite these advancements, the concept of measuring CO by dilution techniques is one where its optimal use in the clinical setting continues to be an area of investigation. The thermodilution concept and the mechanism of measuring CO will be discussed in the following review.
引用
收藏
页码:138 / 144
页数:7
相关论文
共 50 条
[1]   A THERMODILUTION FLOWMETER [J].
AFONSO, S .
JOURNAL OF APPLIED PHYSIOLOGY, 1966, 21 (06) :1883-&
[2]  
[Anonymous], BIOMEDICAL ENG
[3]  
Berne RobertM., 2004, Physiology
[4]   VALIDITY OF THERMODILUTION CARDIAC-OUTPUT DETERMINATION - EXPERIMENTAL STUDIES WITH AND WITHOUT PULMONARY INSUFFICIENCY [J].
BEYER, J ;
LAMBERTI, JJ ;
REPLOGLE, RL .
JOURNAL OF SURGICAL RESEARCH, 1976, 21 (05) :313-317
[5]   INVITRO DETERMINATION OF ACCURACY OF CARDIAC-OUTPUT MEASUREMENTS BY THERMAL DILUTION [J].
BILFINGER, TV ;
LIN, CY ;
ANAGNOSTOPOULOS, CE .
JOURNAL OF SURGICAL RESEARCH, 1982, 33 (05) :409-414
[6]  
BOLDT J, 1994, CRIT CARE MED, V22, P1913
[7]   MEASUREMENT OF CARDIAC OUTPUT BY THERMAL DILUTION IN MAN [J].
BRANTHWAITE, MA ;
BRADLEY, RD .
JOURNAL OF APPLIED PHYSIOLOGY, 1968, 24 (03) :434-+
[8]  
ELLIS RJ, 1972, J AMER MED ASSOC, V220, P507
[9]   MEASUREMENT OF CARDIAC OUTPUT IN ANAESTHETIZED ANIMALS BY A THERMO-DILUTION METHOD [J].
FEGLER, G .
QUARTERLY JOURNAL OF EXPERIMENTAL PHYSIOLOGY AND COGNATE MEDICAL SCIENCES, 1954, 39 (03) :153-164
[10]   MEDICAL THERAPY OF ACUTE MYOCARDIAL-INFARCTION BY APPLICATION OF HEMODYNAMIC SUBSETS .1. [J].
FORRESTER, JS ;
DIAMOND, G ;
CHATTERJEE, K ;
SWAN, HJC .
NEW ENGLAND JOURNAL OF MEDICINE, 1976, 295 (24) :1356-1362