Repeated Gunther Tulip inferior vena cava filter repositioning to prolong implantation time

被引:25
作者
Tay, KH
Martin, ML
Fry, PD
Webb, JG
Machan, LS
机构
[1] Univ British Columbia Hosp, Dept Radiol, Vancouver, BC V6T 2B5, Canada
[2] Univ British Columbia Hosp, Dept Vasc Surg, Vancouver, BC V6T 2B5, Canada
[3] St Pauls Hosp, Dept Intervent Cardiol, Vancouver, BC V6Z 1Y6, Canada
关键词
embolism; pulmonary; thrombosis; venous; venae cavae; filters;
D O I
10.1016/S1051-0443(07)61531-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A patient presented with iliofemoral deep vein thrombosis, a small pulmonary embolism, and a paradoxic embolus to the axillary artery resulting from a patent foramen ovale (PFO). As prophylaxis against further paradoxic emboli while awaiting percutaneous PFO closure, a Gunther Tulip inferior vena cava (IVC) filter was implanted. To prevent incorporation of the IVC filter into the caval wall, it was repositioned twice with use of a filter retrieval set from a transjugular approach. In this way, the implantation time of the filter was extended beyond the recommended period of 10 days. The filter was successfully retrieved 19 days later during percutaneous closure of the PFO.
引用
收藏
页码:509 / 512
页数:4
相关论文
共 13 条
[1]   COMPLICATIONS OF VENA-CAVA FILTERS [J].
ATHANASOULIS, CA .
RADIOLOGY, 1993, 188 (03) :614-615
[2]   The Tempofilter((R)): A multicenter study of a new temporary caval filter implantable for up to six weeks [J].
Bovyn, G ;
Gory, P ;
Reynaud, P ;
Ricco, JB .
ANNALS OF VASCULAR SURGERY, 1997, 11 (05) :520-528
[3]   Incorporation of the Gunther temporary inferior vena cava filter into the caval wall [J].
Burbridge, BE ;
Walker, DR ;
Millward, SF .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1996, 7 (02) :289-290
[4]   PERCUTANEOUS INFERIOR VENA-CAVAL FILTERS - FOLLOW-UP OF 7 DESIGNS IN 320 PATIENTS [J].
FERRIS, EJ ;
MCCOWAN, TC ;
CARVER, DK ;
MCFARLAND, DR .
RADIOLOGY, 1993, 188 (03) :851-856
[5]  
GIMENO MJ, 2001, J VASC INTERV RADI S, V12, pS78
[6]   Gunther tulip filter: Preliminary clinical experience with retrieval [J].
Millward, SF ;
Bhargava, A ;
Aquino, J ;
Peterson, RA ;
Veinot, JP ;
Bormanis, J ;
Wells, PS .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (01) :75-82
[7]   Temporary and retrievable inferior vena cava filters: Current status [J].
Millward, SF .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1998, 9 (03) :381-387
[8]   PRELIMINARY CLINICAL-EXPERIENCE WITH THE GUNTHER TEMPORARY INFERIOR VENA-CAVA FILTER [J].
MILLWARD, SF ;
BORMANIS, J ;
BURBRIDGE, BE ;
MARKMAN, SJ ;
PETERSON, RA .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1994, 5 (06) :863-868
[9]   Gunther tulip retrievable vena cava filter: Results from the registry of the Canadian Interventional Radiology Association [J].
Millward, SF ;
Oliva, VL ;
Bell, SD ;
Valenti, DA ;
Rasuli, P ;
Asch, M ;
Hadziomerovic, A ;
Kachura, JR .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (09) :1053-1058
[10]   NEW RETRIEVABLE PERCUTANEOUS VENA-CAVA FILTER - EXPERIMENTAL IN-VITRO AND IN-VIVO EVALUATION [J].
NEUERBURG, J ;
GUNTHER, RW ;
RASSMUSSEN, E ;
VORWERK, D ;
TONN, K ;
HANDT, S ;
KUPPER, W ;
HANSEN, JV .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1993, 16 (04) :224-229