Comparison of the marginal fit of milled yttrium stabilized zirconium dioxide crowns obtained by scanning silicone impressions and by scanning stone replicas

被引:8
作者
Aranda Yus, Estefania [1 ]
Anglada Cantarell, Josep Maria [2 ]
Minarro Alonso, Antonio [3 ]
机构
[1] Univ Barcelona, Dept Prosthodont, Fac Dent Med Doctorate & Translat Res, Barcelona 08907, Spain
[2] Univ Barcelona, Dept Prosthodont, Fac Dent, Barcelona, Spain
[3] Univ Barcelona, Dept Genet Microbiol & Stadist, Fac Biol, Barcelona, Spain
关键词
Computer-aided design and computer-aided manufacturing (CAD/CAM); Marginal discrepancy; Marginal fit; Scanning silicone impressions; Scanning stone replicas; DENTAL CAD/CAM TECHNOLOGY; COMPUTER-AIDED ANALYSIS; DIGITAL IMPRESSIONS; CLINICAL-EVALUATION; IN-VITRO; ACCURACY; DIGITIZATION;
D O I
10.4047/jap.2018.10.3.236
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
PURPOSE. To determine the discrepancy in monolithic zirconium dioxide crowns made with computer-aided design and computer-aided manufacturing (CAD/CAM) systems by comparing scans of silicone impressions and of master casts. MATERIALS AND METHODS. From a Cr-Co master die of a first upper left molar, 30 silicone impressions were taken. The 30 silicone impressions were scanned with the laboratory scanner, thus obtaining 30 milled monolithic yttrium stabilized zirconium dioxide (YSZD) crowns (the silicone group). They were poured and the working models were scanned, obtaining 30 milled monolithic yttrium stabilized zirconium dioxide (YSZD) crowns (the plaster group). Three predetermined points were analyzed in each side of the crown (Mesial, Distal ,Vestibular and Palatal), and the marginal fit was evaluated with SEM (x600). The response variable is the discrepancy from the master model. A repeated measures ANOVA with two within subject factors was performed to study significance of main factors and interaction. RESULTS. Mean marginal discrepancy was 22.42 +/- 35.65 mu m in the silicone group and 8.94 +/- 14.69 mu m in the plaster group. The statistical analysis showed significant differences between the two groups and also among the four aspects. Interaction was also significant (P=.02). CONCLUSION. The mean marginal fit values of the two groups were within the clinically acceptable values. Significant differences were found between the groups according to the aspects studied. Various factors influenced the accuracy of digitizing such as the design, the geometry, and the preparation guidance, as well as the texture, roughness and the color of the scanned material.
引用
收藏
页码:236 / 244
页数:9
相关论文
共 29 条
[1]   Effect of imaging powder and CAD/CAM stone types on the marginal gap of zirconia crowns [J].
Alghazzawi, Tariq F. ;
Al-Samadani, Khalid H. ;
Lemons, Jack ;
Liu, Perng-Ru ;
Essig, Milton E. ;
Bartolucci, Alfred A. ;
Janowski, Gregg M. .
JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 2015, 146 (02) :111-120
[2]   EVALUATING THE MARGINAL FIT OF ZIRCONIA COPINGS WITH DIGITAL IMPRESSIONS WITH AN INTRAORAL DIGITAL SCANNER [J].
An, Shinyoung ;
Kim, Sungtae ;
Choi, Hyunmin ;
Lee, Jae-Hoon ;
Moon, Hong-Seok .
JOURNAL OF PROSTHETIC DENTISTRY, 2014, 112 (05) :1171-1175
[3]   Finite Element Analysis of the Influence of Geometry and Design of Zirconia Crowns on Stress Distribution [J].
Anami, Lilian Costa ;
da Costa Lima, Julia Magalhaes ;
Corazza, Pedro Henrique ;
Colauto Yamamoto, Eron Toshio ;
Bottino, Marco Antonio ;
Souto Borges, Alexandre Luiz .
JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2015, 24 (02) :146-151
[4]  
Caparroso Pérez Carlos Bernardo, 2011, Rev Fac Odontol Univ Antioq, V22, P186
[5]   Accuracy of contacts calculated from 3D images of occlusal surfaces [J].
DeLong, R. ;
Knorr, S. ;
Anderson, G. C. ;
Hodges, J. ;
Pintado, M. R. .
JOURNAL OF DENTISTRY, 2007, 35 (06) :528-534
[6]  
DeLong R, 2001, J Prosthodont, V10, P78, DOI 10.1111/j.1532-849X.2001.00078.x
[7]   Precision of intraoral digital dental impressions with iTero and extraoral digitization with the iTero and a model scanner [J].
Fluegge, Tabea V. ;
Schlager, Stefan ;
Nelson, Katja ;
Nahles, Susanne ;
Metzger, Marc C. .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2013, 144 (03) :471-478
[8]   CONSIDERATIONS IN MEASUREMENT OF MARGINAL FIT [J].
HOLMES, JR ;
BAYNE, SC ;
HOLLAND, GA ;
SULIK, WD .
JOURNAL OF PROSTHETIC DENTISTRY, 1989, 62 (04) :405-408
[9]   Fit of 4-unit FDPs made of zirconia and CoCr-alloy after chairside and labside digitalization - A laboratory study [J].
Keul, Christine ;
Stawarczyk, Bogna ;
Erdelt, Kurt-Juergen ;
Beuer, Florian ;
Edelhoff, Daniel ;
Gueth, Jan-Frederik .
DENTAL MATERIALS, 2014, 30 (04) :400-407
[10]   THREE-DIMENSIONAL EVALUATION OF GAPS ASSOCIATED WITH FIXED DENTAL PROSTHESES FABRICATED WITH NEW TECHNOLOGIES [J].
Kim, Ki-Baek ;
Kim, Jae-Hong ;
Kim, Woong-Chul ;
Kim, Ji-Hwan .
JOURNAL OF PROSTHETIC DENTISTRY, 2014, 112 (06) :1432-1436