Evaluation of the magnitude of EBT Gafchromic film polarization effects

被引:37
作者
Butson, M. J. [1 ,2 ]
Cheung, T. [2 ]
Yu, P. K. N. [2 ]
机构
[1] Illawarra Canc Care Ctr, Dept Med Phys, Wollongong, NSW 2500, Australia
[2] City Univ Hong Kong, Dept Phys & Mat Sci, Kowloon Tong, Hong Kong, Peoples R China
关键词
Gafchromic; EBT; Radiochromic film; densitometry; orientation; RADIOCHROMIC FILM; QUALITY-ASSURANCE; DOSE-RESPONSE; DOSIMETRY; VERIFICATION; LIGHT;
D O I
10.1007/BF03178624
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Gafchromic EBT film, has become a main dosimetric tools for quantitative evaluation of radiation doses in radiation therapy application. One aspect of variability using EBT Gafchromic film is the magnitude of the orientation effect when analysing the film in landscape or portrait mode. This work has utilized a >99% plane polarized light source and a non-polarized diffuse light source to investigate the absolute magnitude of EBT Gafchromic films polarization or orientation effects. Results have shown that using a non-polarized light source produces a negligible orientation effect for EBT Gafchromic film and thus the angle of orientation is not important. However, the film exhibits a significant variation in transmitted optical density with angle of orientation to polarized light producing more than 100% increase, or over a doubling of measured OD for films irradiated with x-rays up to dose levels of 5 Gy. The maximum optical density was found to be in a plane at an angle of 14 degrees +/- 7 degrees (2SD) when the polarizing sheet is turned clockwise with respect to the film. As the magnitude of the orientation effect follows a sinusoidal shape it becomes more critical for alignment accuracy of the film with respect to the polarizing direction in the anticlockwise direction as this will place the alignment of the polarizing axes on the steeper gradient section of the sinusoidal pattern. An average change of 4.5% per 5 degrees is seen for an anticlockwise polarizer rotation where as the effect is 1.2% per 5 degrees for an clockwise polarizer rotation. This may have consequences to the positional accuracy of placement of the EBT Gafchromic film on a scanner as even a 1 degrees alignment error can cause an approximate 1% error in analysis. The magnitude of the orientation effect is therefore dependant on the degree of polarization of the scanning light source and can range from negligible (diffuse LED light source) through to more than 100% or doubling of OD variation with a fully linear polarized light source.
引用
收藏
页码:21 / 25
页数:5
相关论文
共 30 条
[1]   Measuring energy response for RTQA Radiochromic film to improve quality assurance procedures [J].
Butson, M. J. ;
Cheung, T. ;
Yu, P. K. N. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2008, 31 (03) :203-206
[2]   Radiochromic film for verification of superficial x-ray backscatter factors [J].
Butson, M. J. ;
Cheung, T. ;
Yu, P. K. N. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2007, 30 (04) :269-273
[3]   Scanning orientation effects on Gafchromic EBT film dosimetry [J].
Butson M.J. ;
Cheung T. ;
Yu P.K.N. .
Australasian Physical and Engineering Sciences in Medicine, 2006, 29 (03) :281-284
[4]   Megavoltage x-ray skin dose variation with an angle using grid carbon fibre couch tops [J].
Butson, Martin J. ;
Cheung, Tsang ;
Yu, Peter K. N. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (20) :N485-N492
[5]  
Butson Martin J, 2004, Australas Phys Eng Sci Med, V27, P230
[6]   Extrapolated surface dose measurements with radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Metcalfe, PE .
MEDICAL PHYSICS, 1999, 26 (03) :485-488
[7]   Effects of read-out light sources and ambient light on radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Metcalfe, PE .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (08) :2407-2412
[8]   Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
APPLIED RADIATION AND ISOTOPES, 2006, 64 (01) :60-62
[9]   Absorption spectra variations of EBT radiochromic film from radiation exposure [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (13) :N135-N140
[10]   Polarization effects on a high-sensitivity radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Cheung, T ;
Inwood, D .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (15) :N207-N211