Impact of implementing smart infusion pumps in a pediatric intensive care unit

被引:37
作者
Manrique-Rodriguez, Silvia [1 ]
Sanchez-Galindo, Amelia C. [1 ]
Lopez-Herce, Jesus [1 ]
Angel Calleja-Hernandez, Miguel [2 ]
Martinez-Martinez, Fernando [3 ]
Iglesias-Peinado, Irene [4 ]
Carrillo-Alvarez, Angel [1 ]
Sanjurjo Saez, Maria [1 ]
Fernandez-Llamazares, Cecilia M. [1 ]
机构
[1] Hosp Gen Univ Gregorio Maranon, Pediat Intens Care Unit, Madrid, Spain
[2] Hosp Univ Virgen Nieves, Serv Pharm, Granada, Spain
[3] Campus Cartuja, Fac Pharm, Granada, Spain
[4] Univ Complutense Madrid, Fac Pharm, Madrid, Spain
关键词
ADVERSE DRUG EVENTS; MEDICATION ERRORS; TECHNOLOGY; INTERVENTIONS; PHARMACISTS; SEVERITY;
D O I
10.2146/ajhp120767
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Purpose. The impact of smart infusion pumps on the interception of errors in the programming of i.v. drug administrations on a pediatric intensive care unit (PICU) is investigated. Methods. A prospective observational intervention study was conducted in the PICU of a hospital in Madrid, Spain; to estimate the patient safety benefits resulting from the implementation of smart pump technology (Alaris System, CareFusion, San Diego, CA). A systematic analysis of data stored by the devices during the designated study period (January 2010-June 2011) was conducted using the system software (Guardrails CQI Event Reporter, CareFusion). The severity of intercepted errors was independently classified by a group of four clinical pharmacists and a group of four intensive care pediatricians; analyses of intragroup and intergroup agreement in perceptions of severity were performed. Results. During the 17-month study period, the overall rate of user compliance with the safety software was 78%. The use of smart pump technology resulted in the interception of 92 programming errors, 84% of which involved analgesics, antiinfectives, inotropes, and sedatives. About 97% of the errors resulted from user programming of doses or infusion rates above the hard limits defined in the smart pump drug library. The potential consequences of the intercepted errors were considered to be of moderate, serious, or catastrophic severity in 49% of cases. Conclusion. The use of smart pumps in a PICU improved patient safety by enabling the interception of infusion programming errors that posed the potential for severe injury to pediatric patients.
引用
收藏
页码:1897 / 1906
页数:10
相关论文
共 36 条
[1]   Continuous quality improvement using intelligent infusion pump data analysis [J].
Breland, Burnis D. .
AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2010, 67 (17) :1446-1455
[2]   A validated, reliable method of scoring the severity of medication errors [J].
Dean, BS ;
Barber, ND .
AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 1999, 56 (01) :57-62
[3]  
Echarri L., 2011, EVALUACION ERRORES A, P66
[4]   Impact of smart infusion technology on administration of anticoagulants (unfractionated heparin, argatroban, lepirudin, and bivalirudin) [J].
Fanikos, John ;
Fiumara, Karen ;
Baroletti, Steve ;
Luppi, Carol ;
Saniuk, Catherine ;
Mehta, Amar ;
Silverman, Jon ;
Goldhaber, Samuel Z. .
AMERICAN JOURNAL OF CARDIOLOGY, 2007, 99 (07) :1002-1005
[5]   Impact of clinical pharmacist interventions in reducing paediatric prescribing errors [J].
Fernandez-Llamazares, Cecilia M. ;
Calleja-Hernandez, Miguel A. ;
Manrique-Rodriguez, Silvia ;
Perez-Sanz, Cristina ;
Duran-Garcia, Esther ;
Sanjurjo-Saez, Maria .
ARCHIVES OF DISEASE IN CHILDHOOD, 2012, 97 (06) :564-568
[6]  
FOLLI HL, 1987, PEDIATRICS, V79, P718
[7]   Long-term Reduction in Adverse Drug Events: An Evidence-Based Improvement Model [J].
Gazarian, Madlen ;
Graudins, Linda Velta .
PEDIATRICS, 2012, 129 (05) :E1334-E1342
[8]  
Giannopoulos HT, 2008, PHARM INFORM PRIMER, P119
[9]  
Hadaway L C, 2000, Nurs Manage, V31, P38
[10]   AN 11-YEAR REVIEW OF THE PHARMACY LITERATURE - DOCUMENTATION OF THE VALUE AND ACCEPTANCE OF CLINICAL PHARMACY [J].
HATOUM, HT ;
CATIZONE, C ;
HUTCHINSON, RA ;
PUROHIT, A .
DRUG INTELLIGENCE & CLINICAL PHARMACY, 1986, 20 (01) :33-48